  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Warsaw",
      "address_1": "1800 W Center St",
      "reason_for_recall": "A complaint was received in which it was reported that an 8-hole plate from lot 62276042 had an enlarged hole that is unable to lock the screw, which resulted in the screw head pulling through plate hole.",
      "address_2": "",
      "product_quantity": "28 units",
      "code_info": "Part Number 00-2358-014-04; lot 62276042",
      "center_classification_date": "20140730",
      "distribution_pattern": "Worldwide Distribution - US including the states of  CA and OH., and the countries of  BRAZIL, CHINA, KOREA  and TAIWAN.",
      "state": "IN",
      "product_description": "Zimmer Periarticular Locking Plate System Distal Radial Volar Plate Lateral Column (1st generation stainless steel only).  Orthopedic fixation plate.    For temporary internal fixation and stabilization of osteotomies and fractures.",
      "report_date": "20140806",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Zimmer, Inc.",
      "recall_number": "Z-2131-2014",
      "initial_firm_notification": "",
      "product_type": "Devices",
      "event_id": "68657",
      "termination_date": "20150109",
      "more_code_info": "",
      "recall_initiation_date": "20140610",
      "postal_code": "46580-2304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}