  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Rochester",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "92440",
      "recalling_firm": "Ortho-Clinical Diagnostics",
      "address_1": "1000 Lee Rd",
      "address_2": "N/A",
      "postal_code": "14606-4250",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda, Brazil, Canada L3R 4G5, Chile, China, Colombia, Denmark, France, Germany,  India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore 609917, Spain, Sweden ,  The Netherlands, United Kingdom.",
      "recall_number": "Z-2129-2023",
      "product_description": "VITROS Chemistry Products Calibrator Kit 20-in vitro diagnostic use only.  VITROS Chemistry Products Calibrator Kit 20- Used to calibrate the VITROS 5,1 FS/4600 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems for the quantitative measurement of transferrin, C3, C4, IgA, IgG and IgM.  Product Code: 6801704",
      "product_quantity": "2400 units",
      "reason_for_recall": "Potential to cause biased results in the upper end of the reportable range for the quantitative measurement of Transferrin, C3, C4, IgA, IgG and IgM.  Within the reference interval, the highest bias observed in selected sample testing investigation was +35.3% for IgM; -4.3% for IgG; +12.3% for IgA; -6.9% for C3; -6.6% for C4; and -17.8% for transferrin. Falsely elevated or reduced results could be reported depending on the analyte and/or sample concentration.",
      "recall_initiation_date": "20230508",
      "center_classification_date": "20230711",
      "report_date": "20230719",
      "code_info": "UDI-DI: 10758750006687 Lot Code: 2022 Exp. Date :06-Sept-2023",