  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Yokne'am Ilit",
      "state": "N/A",
      "country": "Israel",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "74464",
      "recalling_firm": "Medtronic Navigation",
      "address_1": "Kochav Yokneam",
      "address_2": "1 Hamelacha St.",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide.",
      "recall_number": "Z-2129-2016",
      "product_description": "PoleStar system (N20 / N30) Warning Lamps Control Box, Warning Lamps Control Box Cables, Thermoflex Chiller    Product The PoleStar Surgical MRI System is intended for use as an intraoperative imaging device to produce MRI images of sections of the head selected by the physician.",
      "product_quantity": "52",
      "reason_for_recall": "Medtronic Navigation is recalling certain components of the PoleStar system (N20 / N30) because of misplaced ground pins which may cause electrical shock to the user.",
      "recall_initiation_date": "20160620",
      "center_classification_date": "20160706",
      "termination_date": "20170621",
      "report_date": "20160713",
      "code_info": "Model No. 506A0107-01 & 506A0107-02 - Warning Lamps Control Box; 504A0276-01, 504A0277-01& 504A0278-01 - Warning Lamps Control Box Cables; 506A0297-01 & 506A0298-01 - Thermoflex Chiller"
    }
  ]
}