  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85572",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution to states of:  AR, AZ, CA, FL, IL, IN, MA, MD, MI, MN, NC, NE, NH, NJ, NY, OH, OK, SC, TN, TX, UT, VA, WA, and Puerto Rico.",
      "recall_number": "Z-2127-2020",
      "product_description": "Cios Alpha Image-Intensified Fluoroscopic X-Ray System, Model Numbers VA20-10308191 and VA30-10308194    The Cios Alpha and Cios Spin are a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications.",
      "product_quantity": "58",
      "reason_for_recall": "The unplugging of the main cable on the monitor trolley from the X10 connector can cause an unsafe electrical voltage (up to 67,6 V DC) at the X10 connector. This can result in an electrical shock if the user or other persons touch the contact pins of the X10 connector.",
      "recall_initiation_date": "20200417",
      "center_classification_date": "20200526",
      "termination_date": "20200825",
      "report_date": "20200603",
      "code_info": "Serial Numbers:  11200  13288  13259  12339  13241  10019  12377  13219  12946  13232  13045  12618  12972  40109  40110  40073  40057  40013  40082  40093  40078  40107  40108  40111  40120  40126  40134  40135  40104  40074  40095  40076  40080  40058  40077  40081  40055  40054  40075  40125  40113  40102  40071  40061  40094  40124  40070  40103  40118  40119  40121  40084  40122  40123  40083  40062  40130  40116"
    }
  ]
}