{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Horten",
      "state": "N/A",
      "country": "Norway",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "92505",
      "recalling_firm": "GE Vingmed Ultrasound As",
      "address_1": "Strandpromenaden 45",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution.",
      "recall_number": "Z-2126-2023",
      "product_description": "GE Healthcare Vivid S60N, ultrasound device",
      "product_quantity": "N/A",
      "reason_for_recall": "GE HealthCare has become aware that certain Vivid ultrasound systems cannot boot up in a timely fashion. If this occurs, it can delay availability of the device in time-critical emergency situations.",
      "recall_initiation_date": "20230530",
      "center_classification_date": "20230710",
      "report_date": "20230719",
      "code_info": "Software versions:  v203, v204, v205, v206",
      "more_code_info": ""
    }
  ]
}