  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Firenze",
      "state": "N/A",
      "country": "Italy",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "74391",
      "recalling_firm": "Esaote S.p.A.",
      "address_1": "Via Di Caciolle,15",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Domestic: FL, TX, OH, MN, NV;   Foreign: No customers in Canada;  No VA/DOD.",
      "recall_number": "Z-2124-2016",
      "product_description": "Magnetic Resonance System G-scan Brio (Part Number /Model No.101002000)",
      "product_quantity": "11 USA;  29 OUS",
      "reason_for_recall": "The device is being recalled because dynamic loading forces from larger patients can fracture the extremity portion of the patient table.  Additionally, a weight limit is added to the patient seat for weight-bearing examinations.",
      "recall_initiation_date": "20160606",
      "center_classification_date": "20160705",
      "termination_date": "20170530",
      "report_date": "20160713",
      "code_info": "Item No., Unit Description and Serial No.s:  101002000, UNIT, G-SCAN BRIO, 05202;  101002000, UNIT, G-SCAN BRIO, 05203;  101002000, UNIT,  G-SCAN BRIO 2.1A; 05205;  101002000, UNIT, G-SCAN BRIO 2.1A, 05207;  101002000 UNIT, G-SCAN BRIO 2.1A, 05208;  101002000 UNIT, G-SCAN BRIO 2.1A, 05209;  101002000 UNIT, G-SCAN BRIO 2.1A, 05210;  101002000 UNIT, G-SCAN BRIO 2.1A, 05211;  101002000 UNIT,  G-SCAN BRIO 2.1A, 05305;  101002000 UNIT,  G-SCAN BRIO 2.1A, 05308"
    }
  ]
}