  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "71574",
      "recalling_firm": "GE Healthcare",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide Distribution including GU and DC., and to the countries of : Australia, Bulgaria, Canada, Chile, Dominican Republic, Germany, Ireland, Italy, Kuwait, Lebanon, Portugal, Romania, Russia, Saudi Arabia,  Slovakia, Spain, Taiwan,Turkey, United Kingdom, Venezuela and Yemen.",
      "recall_number": "Z-2123-2015",
      "product_description": "GE Healthcare, Precision MPi fluoroscopic imaging systems.",
      "product_quantity": "788 (701 US; 87 OUS)",
      "reason_for_recall": "A reported incident of a monitor that fell from the suspension.  A fall of a monitor could result in an injury to a patient or operator.  Recall affects the single and dual LCD monitor suspensions used on Precision 500D Classical R & F system, Precision RXi 23A/32A Analog System, Precision RXi Digital system and Precision MPi fluoroscopic imaging systems.",
      "recall_initiation_date": "20150612",
      "center_classification_date": "20150720",
      "termination_date": "20170803",
      "report_date": "20150729",
      "code_info": "Mfg Lot or Serial #  00000000687MC4  00000001286MC4  00000004134MC3  00000003843MC0  00000003890MC1  00000004159MC0  00000004689MC6  00000001505MC7  00000003889MC3  00000003937MC0  00000002207MC9  00000003931MC3  00000001452MC2  00000003847MC1  00000004427MC1  00000004649MC0  00000003642MC6  00000001343MC3  00000004650MC8  00000004161MC6  00000001459MC7  00000001658MC4  00000001333MC4  00000003511MC3  00000001430MC8  00000001057MC9  00000001055MC3  00000001275MC7  00000001420MC9  00000001406MC8  00000001889MC5  00000001408MC4  00000003936MC2  00000001101MC5  00000001415MC9  00000001746MC7  00000001419MC1  00000004701MC9  0000000470FMC9  00000004700MC1  00000000952MC2  00000001302MC9  00000000921MC7"
    }
  ]
}