  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Canton",
      "address_1": "5449 Research Dr",
      "reason_for_recall": "The original mean assigned to the control too high, upon reassignment with additional data the new assigned mean was outside of the originally published range.",
      "address_2": "",
      "product_quantity": "132919 Vol vials",
      "code_info": "Model Number(s): C7590-50 (Chemistry Control Level I), C7591-50 (Chemistry Control Level II), and C7592-100 (Chemistry Control Level I and II).",
      "center_classification_date": "20170520",
      "distribution_pattern": "Nationwide",
      "state": "MI",
      "product_description": "Pointe Scientific Chemistry Controls Level I and II    LEVl: Amber glass vial w/PP white plastic cap.  LEV2: Amber glass vial w/PP black plastic cap.",
      "report_date": "20170531",
      "classification": "Class III",
      "openfda": {},
      "recalling_firm": "Medtest Holdings, Inc.",
      "recall_number": "Z-2121-2017",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "77083",
      "termination_date": "20170602",
      "more_code_info": "",
      "recall_initiation_date": "20140515",
      "postal_code": "48188-2261",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}