  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Waltham",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "62512",
      "recalling_firm": "Perkinelmer",
      "address_1": "940 Winter Street",
      "address_2": "N/A",
      "postal_code": "02451",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US (nationwide) and the countries of  Australia, Belgium, Brazil,  Egypt, France,  Germany, Netherlands, Norway, Saudi Arabia,Switzerland,",
      "recall_number": "Z-2121-2012",
      "product_description": "Perkin Elmer Genetic Screening Processor (GSP)  Product Code: 2021-0010    Product Usage:  The GSP instrument is a fully automated, high throughput batch analyzer for time- resolved and prompt fluorescence analysis of samples in microtitration plates. It is intended for in vitro quantitative / qualitative determination of analytes in body fluids. The reagent kits running on the GSP instrument are used to test blood specimens dried on filter paper as an aid in screening newborns for congenital disorders. Currently assays for the determination of T4, TSH, 17OHP, IRT and GALT are cleared and available for use with the GSP instrument.",
      "product_quantity": "78 units",
      "reason_for_recall": "Contamination from GSP Shaker unit results in decreased signal activity of assay measurements for Genetic Screening for Neonatal IRT, TSH and IRT",
      "recall_initiation_date": "20120627",
      "center_classification_date": "20120731",
      "termination_date": "20140418",
      "report_date": "20120808",
      "code_info": "Serial Numbers: 20210001 through 20210078"
    }
  ]
}