  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Northfield",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98642",
      "recalling_firm": "Medline Industries, LP",
      "address_1": "3 Lakes Dr",
      "address_2": "",
      "postal_code": "60093-2753",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide distribution.",
      "recall_number": "Z-2120-2026",
      "product_description": "Medline Convenience Kits:  1) BURN PACK, Model Number: DYNJ15668R;   2) BURN PACK-LF, Model Number: DYNJ42829B;   3) BURN CARE PACK-LF, Model Number: DYNJT3437",
      "product_quantity": "291 units",
      "reason_for_recall": "Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.",
      "recall_initiation_date": "20260107",
      "center_classification_date": "20260507",
      "report_date": "20260513",
      "code_info": "1) DYNJ15668R, UDI-DI: 10195327108502(each), 40195327108503(case), Lot Number: 23HMB037; 2) DYNJ42829B, UDI-DI: 10193489880151(each), 40193489880152(case), Lot Number: 22IBI084; 3) DYNJT3437, UDI-DI: 10198459325076(each), 40198459325077(case), Lot Number: 25FBK633",
      "more_code_info": ""
    }
  ]
}