  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Gainesville",
      "state": "FL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "62566",
      "recalling_firm": "Exactech, Inc.",
      "address_1": "2320 NW 66th Ct",
      "address_2": "N/A",
      "postal_code": "32653-1630",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA including FL, MO, MA, AL, KY, IN, NC, LA,PA, VA, OH, NY, ME, MI, TX, KS, TN, OK, NJ, WA, CO, GA, CA, MN, IA, OR, and IL and Internationaly to Austria, Luxenbourg, Sweden, Germany, Spain, France, Greece, Italy, Nederland, United Kingdom, Canada, India, Australia, Switzerland, and Sao Palo.",
      "recall_number": "Z-2120-2012",
      "product_description": "Octane-C Cervical, 6¿ Lordotic  Intended for anterior cervical spine intervertebral body fusion procedures at one level (C3-C7).",
      "product_quantity": "54,877 units for all products in Recall Event",
      "reason_for_recall": "Exactech, Inc. initiated a recall of Equinoxe Replicator Plates, Equinoxe Locking Screws, Equinoxe Compression Screw Kits, Equinoxe Fracture Adapter Tray Screws, and Octane-C Intervertebral fusion devices due to the firm determining that the device labeling identifies the affected parts as having a ten year shelf life; however, the packaging configuration that was used it qualified to a five year",
      "recall_initiation_date": "20120615",
      "center_classification_date": "20120801",
      "termination_date": "20130703",
      "report_date": "20120808",
      "code_info": "Sizes:   5 mm Catalog # 05-010-02-0605 All serial numbers  6 mm Catalog # 05-010-02-0606 All serial numbers  7 mm Catalog # 05-010-02-0607 All serial numbers  8 mm Catalog # 05-010-02-0608 All serial numbers  9 mm Catalog # 05-010-02-0609 All serial numbers  10mm Catalog # 05-010-02-0610 All serial numbers  11mm Catalog # 05-010-02-0611 All serial numbers  12mm Catalog # 05-010-02-0612 All serial numbers"
    }
  ]
}