  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Minneapolis",
      "address_1": "7000 Central Ave Ne",
      "reason_for_recall": "Potential for misalignment of the Z-stage scale. Using one of these devices for a procedure could result in the microelectrode being inserted to an incorrect target depth.",
      "address_2": "",
      "product_quantity": "300  EXPANDED 10/02/2014 130 devices",
      "code_info": "Lot numbers:  082605614, 082601414, 082602014, 082634313, 082630913, 082629713, 082629613, 082626913, 082623113, 082621413, and 082618213.      EXPANDED 10/02/2014 Lot Numbers:  082616214, 082616314,  082618314, and 082618414.",
      "center_classification_date": "20140725",
      "distribution_pattern": "Worldwide Distribution - USA including CA, CO, FL, IL, KS, KY, LA, MN, MT, NC, NE, OR, TX, VA, and UT, and Internationally to Canada, Spain, and United Kingdom.",
      "state": "MN",
      "product_description": "Medtronic Nexdrive Micropositioning Drive. Models Ml-1000 and Ml-2000.      For use in conjunction with the Medtronic Nexframe Stereotactic System for the precise positioning of microelectrodes and implantable leads.  A stereotactic guidance system used in conjunction with Medtronic StealthStation Navigation Systems-image-guided surgery (IGS) systems-for Deep Brain Stimulation (DBS) procedures.",
      "report_date": "20140806",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Medtronic Neuromodulation",
      "recall_number": "Z-2118-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "68615",
      "termination_date": "20150122",
      "more_code_info": "",
      "recall_initiation_date": "20140611",
      "postal_code": "55432-3568",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}