{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Gainesville",
      "state": "FL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "62566",
      "recalling_firm": "Exactech, Inc.",
      "address_1": "2320 NW 66th Ct",
      "address_2": "N/A",
      "postal_code": "32653-1630",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA including FL, MO, MA, AL, KY, IN, NC, LA,PA, VA, OH, NY, ME, MI, TX, KS, TN, OK, NJ, WA, CO, GA, CA, MN, IA, OR, and IL and Internationaly to Austria, Luxenbourg, Sweden, Germany, Spain, France, Greece, Italy, Nederland, United Kingdom, Canada, India, Australia, Switzerland, and Sao Palo.",
      "recall_number": "Z-2116-2012",
      "product_description": "Equinoxe Fixed Angle Replicator Plate    Intended for Primary and Reverse Total Shoulder Arthroplasty.",
      "product_quantity": "54,877 units for all products in Recall Event",
      "reason_for_recall": "Exactech, Inc. initiated a recall of Equinoxe Replicator Plates, Equinoxe Locking Screws, Equinoxe Compression Screw Kits, Equinoxe Fracture Adapter Tray Screws, and Octane-C Intervertebral fusion devices due to the firm determining that the device labeling identifies the affected parts as having a ten year shelf life; however, the packaging configuration that was used it qualified to a five year",
      "recall_initiation_date": "20120615",
      "center_classification_date": "20120801",
      "termination_date": "20130703",
      "report_date": "20120808",
      "code_info": "Size: 0mm  Catalog Number: 300-25-00  Serial Range: 1999574-2381151"
    }
  ]
}