  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Newport News",
      "state": "VA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83144",
      "recalling_firm": "Burlington Medical, LLC",
      "address_1": "PO Box 3194",
      "address_2": "3 Elmhurst St",
      "postal_code": "23603-1137",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution.    The products were distributed to the following foreign countries:  Korea, Thailand.",
      "recall_number": "Z-2115-2019",
      "product_description": "Overlap Vest, Single Item, Product Number FVest, Part Numbers REFVEST*F, REFVEST*M, REFVESTE, REFVESTXLM, SEFVEST*F, SEFVEST*M",
      "product_quantity": "7",
      "reason_for_recall": "The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.",
      "recall_initiation_date": "20190607",
      "center_classification_date": "20190730",
      "termination_date": "20230428",
      "report_date": "20190807",
      "code_info": "Serial Numbers: 385688  387953  385694  385696  387781  384424  382874"
    }
  ]
}