      "status": "Ongoing",
      "city": "Orlando",
      "state": "FL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80057",
      "recalling_firm": "Invivo Corporation",
      "address_1": "12151 Research Pkwy Ste 200",
      "address_2": "N/A",
      "postal_code": "32826-3286",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) to states of: AK, AL, AR, AZ, CA, CT, DC, FL, GA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MO, NC, NH, NJ, NM, NY, NV, OH, PA, RI, SC, TN, TX, VA, WA, and WI ; and countries of: Canada, China, Great Britain and Korea.",
      "recall_number": "Z-2115-2018",
      "product_description": "Xper Flex Cardio Physiomonitoring System    The Xper Flex Cardio physiomonitoring system is used to facilitate invasive investigation of heart and vascular, disease when non-invasive indicators warrant such.  The Xper Flex Cardio physiomonitoring system may be used to display and analyze surface ECG (electrocardiogram), respiration, invasive pressure, Sp02 (Pulse Oximetry), End Tidal C02 (carbon dioxide) and non-invasive pressure waveforms; surface body temperature and thermal cardiac output curves.",
      "product_quantity": "1,040 units (USA) and 7 units (Foreign)",
      "reason_for_recall": "While in Full Disclosure playback, a user may inadvertently close the Full Disclosure Control Window using the ESC key function, instead of pressing the  X  icon in the upper right corner of this window.",
      "recall_initiation_date": "20171120",
      "center_classification_date": "20180607",
      "report_date": "20180613",
      "code_info": "MODEL #(s):   453564428541 (Service # 453564241911), 453564243601 (Service # 453564483331), 453564634211 (Service # 453564621801)  SERIAL NUMBERS:  US1203010251, US1239000027, US1203010279, US1203010280, US1203010308, US1203010334, US1203010346, US1203010352,  US1203010358, US1203010360, US1203010366, US1203010375,   US1203010398, US1203010407, US1304000043, US1304000046,   US1224000050, US1550001470, US1251000063, US1109010070,   US1109010072, US1109010075, US1201010130, US1331000013,  US1505000507, US1343000034, US1311000042, US1337000033,  US1201010112, US1201010119, US1201010120, US1201010122,  US1202010171, US1202010186, US1202010202, US1203010396,   US1231000007, US1231000032, US1231000050, US1231000074,  US1231000090, US1234000008, US1311000016, US1311000023,  US1248000071, US1708000101, US1720000441, US1550001478,   US1248000065, US1729000603, US1352000006, US1352000022, US1352000044, US1352000046, US1352000048, US1107010005,   US1231000060, US1537001268, US1537001272, US1537001276,  US1537001277, US1537001280, US1537001338, US1202010174,   US1417000023, US1419000042, US1412000038, US1419000004,   US1419000023, US1419000052, US1424000010, US1424000018,  US1431000027, US1239000034, US1550001491, US1338000001,   US1550001502, US1550001468, US1611001807, US1611001814,   US1611001826, US1611001831, US1611001850, US1248000004,   US1248000025, US1248000028, US1248000035, US1517000665,   US1517000671, US1517000680, US1443000225, US1724000551,   US1724000561, US1729000573, US1729000601, US1729000604,   US1311000065, US1407000058, US1334000037, US1224000027,   US1304000017, US1517000691, US1517000696, US1352000036,  US1652000003, US1652000011, US1652000016, US1652000042,  US1505000498, US1505000518, US1505000567, US1337000007,  US1337000008, US1343000057, US1352000052, US1203010408,   US1511000639, US1534001157, US1534001158, US1534001163,  US1534001165, US1534001167, US1534001171, US1534001220,   US1534001223, US1534001226, US1534001229, US1534001230,   US1534001232, US1534001233, US1534001237, US1537001259,   US1537001265, US1324000046, US1443000180, US1613001985, US1617002038, US1619002059, US1619002079, US1619002081, US1517000666, US1517000669, US1517000685, US1407000059, US1441000164, US1443000181, US1443000195, US1441000163,   US1502000477, US1529001080, US1224000020, US1231000023,   US1231000035, US1231000064, US1231000081, US1231000083,   US1234000001, US1234000004, US1234000006, US1234000055,   US1352000017, US1352000023, US1352000024, US1304000041,  US1352000059, US1107010009, US1107010013, US1107010014,  US1108010043, US1108010057, US1109010073, US1352000062,  US1407000038, US1407000041, US1408000002, US1716000343,   US1716000344, US1716000345, US1720000353, US1720000354,   US1720000363, US1709000268, US1525000986, US1525000991,   US1529000995, US1529001027, US1529001033, US1529001068,   US1107010034, US1109010069, US1201010136, US1505000478,   US1505000511, US1505000513, US1505000546, US1343000092,   US1417000006, US1443000219, US1638002490, US1611001828,   US1611001832, US1611001848, US1611001863, US1505000508,   US1505000528, US1505000529, US1505000540, US1534001217,   US1720000386, US1720000407, US1720000409, US1720000416,   US1720000419, US1720000422, US1720000427, US1720000428,  US1446000242, US1244000010, US1251000075, US1224000034,  US1244000048, US1608001665, US1609001701, US1609001702,  US1609001703, US1609001704, US1609001706, US1609001707,   US1609001708, US1609001709, US1609001712, US1609001714,   US1609001722, US1609001727, US1609001729, US1652000009,   US1652000025, US1652000037, US1652000045, US1652000052, US1652000062, US1652000064, US1652000066, US1248000057,  US1331000025, US1331000053, US1331000063, US1331000050,   US1331000069, US1331000070, US1334000001, US1331000001,  US1331000002, US1331000003, US1331000012, US1331000022,  US1238000001, US1202010150, US1304000045, US1325000027,  US1328000007, US1328000008, US1334000025, US1343000049,  US1331000007, US1446000243, US1446000287, US1450000325  US1431000040, US1433000064, US1433000067, US1537001311,  US1537001331, US1547001372, US1550001454, US1550001472,  US1550001476, US1550001484, US1407000012, US1244000068,  US1248000007, US1248000010, US1248000015, US1248000024,  US1248000030, US1251000023, US1239000014, US1203010353,  US1202010152, US1202010198, US1202010212, US1202010235,  US1203010374, US1203010395, US1202010180, US1202010181,   US1202010184, US1202010185, US1202010187, US1202010190,  US1202010192, US1202010195, US1107010004, US1248000031,   US1248000042, US1248000046, US1248000050, US1248000059,   US1502000401, US1502000409, US1502000427, US1502000431,   US1502000433, US1502000435, US1502000437, US1502000447,   US1502000449, US1502000452, US1438000148, US1505000480,   US1343000022, US1107010035, US1109010082, US1109010087,   US1109010091, US1109010100, US1505000499, US1505000541,   US1505000542, US1443000235, US1446000245, ,US1446000276  US1324000023, US1324000047, US1325000013, US1325000014, US1325000001, US1325000003, US1325000021, US1224000039,  US1331000023, US1343000089, US1611001820, US1611001876,  US1611001888, US1611001916, US1613001928, US1613001932,   US1613001947, US1613001950, US1613001957, US1613001967,  US1613001969, US1613001971, US1708000102, US1708000110,  US1708000126, US1709000263, US1716000304, US1537001302,   US1534001159, US1534001184, US1534001188, US1547001362,  US1547001387, US1547001399, US1547001400, US1547001403,  US1547001405, US1608001636, US1608001637, US1608001638,   US1608001641, US1608001642, US1608001643, US1608001644,   US1608001652, US1608001656, US1517000663, US1517000672,   US1517000677, US1517000682, US1517000684, US1304000035,   US1311000036, US1251000090, US1251000092, US1251000093,   US1251000095, US1251000096, US1304000008, US1304000010,   US1304000014, US1304000019, US1304000029, US1304000030,   US1304000039, US1251000080, US1251000081, US1251000083   US1251000091, US1304000021, US1304000023, US1304000024,   US1248000081, US1311000082, US1517000651, US1108010045,   US1610001780, US1610001783, US1610001786, US1610001787,   US1610001789, US1610001792, US1610001793, US1610001794, US1608001645, US1608001651, US1608001654, US1608001672,  US1608001673, US1608001677, US1608001679, US1608001682,   US1724000543, US1724000485, US1724000499, US1724000510,   US1724000521, US1724000523, US1724000526, US1724000530,   US1724000535, US1724000538, US1724000546, US1724000559,   US1251000086, US1248000037, US1251000072, US1304000004,   US1324000027, US1324000042, US1325000012, US1325000018,  US1328000003, US1343000016, US1324000019, US1244000041,  US1720000382, US1720000405, US1720000456, US1724000463,  US1724000464, US1724000473, US1724000480, US1724000482,   US1529001083, US1529001090, US1534001196, US1534001204,  US1537001293, US1412000045, US1720000444, US1224000016,  US1408000011, US1408000041, US1412000032, US1412000041,  US1417000009, US1417000010, US1417000029, US1518000751,  US1523000816, US1617001988, US1619002078, US1619002080,  US1619002083, US1202010183, US1304000018, US1231000033,  US1231000057, US1231000084, US1234000014, US1234000062,  US1239000021, US1450000351, US1450000357, US1505000556,  US1550001463, US1550001473, US1550001479, US1550001500,   US1550001505, US1608001640, US1611001913, US1611001914,  US1611001917, US1613001943, US1613001961, US1613001968,  US1613001970, US1613001976, US1613001979, US1613001980,   US1613001981, US1613001982, US1617001997, US1617002005,   US1617001989, US1617001991, US1619002071, US1619002076,  US1619002077, US1619002103, US1619002105, US1334000038,  US1334000045, US1334000043, US1251000011, US1203010257,  US1203010262, US1441000165, US1443000166, US1443000171,  US1443000172, US1443000202, US1443000205, US1443000208,   US1203010260, US1203010287, US1203010325, US1203010336,   US1203010362, US1203010363, US1203010271, US1203010294,  US1203010322, US1203010383, US1203010298, US1203010315, US1203010319, US1203010364. US1505000482. US1505000502.   US1505000509, US1505000565, US1108010060, US1201010147,  US1407000016, US1224000004, US1224000021, US1224000037,  US1224000041, US1224000046, US1244000003, US1224000003,  US1224000024, US1231000012, US1231000031, US1231000052,   US1231000069, US1231000078, US1234000003, US1234000016,  US1234000020, US1239000046, US1407000024, US1319000011,   US1724000492, US1724000537, US1724000548, US1201010124,   US1201010134, US1201010139, US1201010140, US1201010142,   US1337000023, US1337000029, US1343000005, US1343000028,   US1343000033, US1343000039, US1343000061, US1343000084,   US1343000025, US1324000004, US1446000248, US1450000323,  US1502000426, US1502000441, US1505000569, US1547001441,  US1604001573, US1604001625, US1244000025, US1202010148,  US1202010154, US1202010178, US1202010204, US1202010237,   US1547001367, US1708000116, US1716000346, US1716000347,  US1720000355, US1720000358, US1720000377, US1720000378,  US1720000421, US1720000423, US1720000430, US1720000431,  US1720000435, US1720000436, US1720000437, US1720000439,  US1407000063, US1550001469, US1408000036, US1202010234,   US1625002311, US1625002324, US1625002333, US1625002338,  US1729000572, US1729000588, US1729000600, US1729000606,  US1331000047, US1352000015, US1352000045, US1352000054,  US1352000057, US1623002112, US1619002049, US1623002116,  US1623002118, US1623002141, US1623002144, US1623002147,  US1623002150, US1623002151, US1623002155, US1623002159,  US1623002176, US1624002198, US1624002206, US1624002209,  US1624002223, US1625002226, US1505000484, US1505000487,  US1505000488, US1505000494, US1505000497, US1505000523, US1505000524, US1505000531, US1505000564, US1505000566,   US1626002468, US1638002480, US1638002485, US1638002489,  US1638002491, US1638002501, US1638002515, US1638002518,  US1613001963, US1617001987, US1518000806, US1625002344,  US1626002375, US1626002380, US1626002388, US1626002389,   US1626002403, US1626002413, US1626002416, US1626002438,  US1626002460, US1626002464, US1311000012, US1311000027,  US1311000033, US1311000035, US1311000038, US1311000043,   US1311000045, US1311000047, US1311000048, US1224000017,  US1239000039, US1708000115, US1638002529, US1638002545,  US1638002556, US1638002567, US1638002572, US1638002574,  US1638002578, US1638002579, US1638002583, US1638002584,   US1638002586, US1638002588, US1638002595, US1638002493,   US1638002483, US1638002494, US1638002499, US1638002508,   US1638002486, US1248000073, US1251000006, US1202010206,  US1334000022, US1407000040, US1407000048, US1407000053,  US1407000056, US1407000060, US1433000079, US1433000081,  US1438000111, US1438000119, US1438000121, US1438000126,  US1438000128, US1438000132, US1438000140, US1438000141,  US1438000142, US1438000143, US1438000149, US1517000676,  US1625002315, US1626002374, US1626002379, US1626002405,  US1626002406, US1325000005, US1331000011, US1343000019,  US1324000016, US1248000075, US1608001626, US1202010161,   US1248000058, US1248000067, US1248000068, US1248000070,  US1248000072, US1251000003, US1251000012, US1251000015,  US1251000016, US1248000005, US1547001384, ,US1311000076,  US1343000088, US1108010044, US1611001817, US1407000020,  US1407000027, US1408000015, US1251000085, US1626002469,  US1626002470, US1626002471, US1638002481, US1638002500,  US1638002507, US1638002514, US1638002523, US1638002534,   US1502000414, US1502000428, US1517000656, US1450000326,  US1324000048, US1231000017, US1502000444, ,US1107010018  US1107010000, US1108010048, US1109010079, US1625002279,  US1203010341, US1203010420, US1203010283, US1319000085,  US1534001193, US1534001195, US1534001209, US1534001215,  US1352000041, US1407000046, US1352000035, US1419000035,  US1248000002, US1248000014, US1248000026, US1251000014,  US1331000056, US1613001951, US1441000158, US1443000201,  US1443000221, US1443000224, US1443000226, US1443000227,  US1443000228, US1443000230, US1443000233, US1446000241,  US1446000260, US1446000261, US1446000279, US1408000018,  US1224000006, US1224000010, US1224000036, US1224000052,  US1108010052, US1238000007, US1613001931, US1613001933,  US1613001935, US1611001867, US1611001869, US1611001891,  US1611001895, US1611001896, US1611001900, US1611001920,  US1613001934, US1613001940, US1613001949, US1652000005,  US1652000008, US1652000034, US1652000035, ,US1652000063,  US1708000113, US1708000203, US1638002495, US1638002504,  US1638002524, US1638002533, US1337000025, US1343000075,  US1343000085, US1343000086, US1343000087, US1352000001,  US1352000028, US1352000032, US1352000042, US1352000060,  US1337000041, US1334000039, US1609001718, US1609001719,  US1609001738, US1609001739, US1609001743, US1610001749,  US1610001750, US1610001757, US1608001678, US1505000514,  US1611001906, US1611001924, US1334000019, US1331000062,  US1517000717, US1517000725, US1517000726, US1518000742,  US1518000743, US1518000746, US1517000712, US1517000713,  US1517000715, ,US1517000716 ,US1517000738, US1518000744,  US1518000773, US1518000774, US1518000775, US1438000110,  US1419000026, US1441000162, US1502000388, US1517000679,   US1202010218, US1202010221, US1202010222, US1202010223,  US1202010228, US1202010230, US1202010232, US1202010240,  US1202010244, US1202010246, US1202010249, US1203010293,  US1203010297, US1203010309, US1203010311, US1203010323,  US1203010327, US1203010328, US1203010339, US1239000044,  US1251000046, US1311000003, US1231000004, US1319000009,  US1638002566, US1638002599, US1244000027, US1109010080,  US1231000075, US1234000012, US1234000037, US1234000064,  US1238000008, US1238000010, US1239000020, US1239000031,  US1239000033, US1239000038, US1244000006, US1244000050,  US1202010194, US1202010245, US1331000044, US1328000013,  US1331000021, US1251000084, US1248000052, US1331000049,  US1331000068, US1334000004, US1224000047, US1234000080,  US1251000039, US1251000043, US1251000045, US1251000048,  US1251000049, US1251000060, US1251000062, US1251000068,  US1224000023, US1107010028, US1107010029, US1107010039,  US1107010032, US1550001453, US1638002623, US1644002626,  US1644002628, US1644002630, US1511000649, US1625002268,  US1625002273, US1625002278, US1625002286, US1234000070,  US1708000120, US1709000273, US1239000030, US1239000049,  US1324000031, US1324000044, US1502000439, US1443000183,  US1550001480, US1202010215, US1231000019, US1251000089,  US1534001245, US1251000022, US132000007, US132000019,  US1203010397, US1238000002, US1238000003, US1523000881,   US1523000854, US1523000855, US1523000866, US1523000893,  US1431000005, US1431000011, US1431000014, US1431000026,  US1617002002, US1617002003, US1617002010, US1617002025,  US1617002028, US1617002045, US1231000071, US1239000028,  US1724000489, US1248000063, US1525000907, US1505000517,  US1724000506, US1724000512, US1724000517, US1724000539,  US1724000552, US1311000081, US1319000014, US1319000018,  US1319000095, US1324000040, US1502000470, US1525000915,  US1525000924, US1525000927, US1525000933, US1525000935,  US1525000945, US1525000952, US1251000025, US1505000490,  US1720000458, US1724000466, US1724000474, US1724000487,  US1724000495, US1724000496, US1724000501, US1724000507,  US1619002099, US1623002117, US1624002199, US1224000035,  US1201010128, US1720000448, and US1224000013"
    }
  ]
}