  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Tarrytown",
      "address_1": "511 Benedict Ave",
      "reason_for_recall": "The ADVIA 120 optics cover or hood is located on the very top of the ADVIA 120 analyzer.  When the optics cover is raised, it is supported by two gas struts (SMN 10309266) attached to the cover on the left and right sides.  Over time, it is possible that these struts may lose their effectiveness and fail to support the optics cover in the open position.  This could cause the cover to fall during maintenance procedures.",
      "address_2": "",
      "product_quantity": "Domestic: 950 units; Foreign: 3369 units",
      "code_info": "All serial numbers",
      "center_classification_date": "20130828",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and the countries of  Austria, Australia, Belgium, Bosnia - Herzegovina, China, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hungary, Iceland, India, Indonesia, Ireland, Iraq, Israel, Italy, Japan, Kenya, Republic of Korea, Kuwait, Lebanon, Lithuania, Luxembourg, Martinique, Malaysia, Netherlands, Norway, New Zealand, Philippines, Poland, Portugal, Romania, Russian Republic, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom and Vietnam.",
      "state": "NY",
      "product_description": "ADVIA 120 Hematology Analyzer, Catalog Numbers SMN 10360958 and 10360959 .    The ADVIA 120 is a hematology system that utilizes the principles of flow cytometry in order to provide complete blood counts.",
      "report_date": "20130904",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Siemens Healthcare Diagnostics",
      "recall_number": "Z-2114-2013",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "65852",
      "termination_date": "20170228",
      "more_code_info": "",
      "recall_initiation_date": "20130626",
      "postal_code": "10591-5005",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}