  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Malvern",
      "address_1": "51 Valley Stream Pkwy",
      "reason_for_recall": "Siemens has become aware of the potential malfunction when using syngo Imaging XS, version VA70A or higher. In some cases a miscalculation may occur for the grey scale values in the functions \u001cRegion of Interest\u001d, \u001cPixel Lens\u001d, \u001cEdge enhancement\u001d and \u001cHistograms\u001d. This potential malfunction may occur for images with a pixel depth greater than 12 bit. All values calculated for images with a pixel d",
      "address_2": "",
      "product_quantity": "141",
      "code_info": "model number: 10496279",
      "center_classification_date": "20140723",
      "distribution_pattern": "Nationwide Distribution",
      "state": "PA",
      "product_description": "Siemens syngo Imaging XS, version VA70A or higher    Syngo Imaging XS is a Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital medical images. It supports the physician in diagnosis and treatment planning. Syngo Imaging XS also supports storage and archiving of Structured DICOM reports. Syngo Imaging XS optionally uses a variety of advanced post processing applications.",
      "report_date": "20140730",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "recall_number": "Z-2111-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "68769",
      "termination_date": "20150224",
      "more_code_info": "",
      "recall_initiation_date": "20140630",
      "postal_code": "19355",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}