{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Cleveland",
      "address_1": "595 Miner Rd",
      "reason_for_recall": "Socket Head Cap Screws (SHCS) used in the Z-Axis Adjustment Plate were found to be made of stainless steel rather than the specified alloy steel.",
      "address_2": "",
      "product_quantity": "47 units",
      "code_info": "Big Bore Oncology System Code 728243; Serial numbers:   75002, 75003, 75004, 75005, 75006, 75007, 75008, 75009, 75010, 75011, 75012, 75014, 75015, 75016, 75017, 75018, 75019, 75020, 75021, 75022, 75023, 75024, 75025, 75026, 75030, 75032, 75035, 75036, 75037, 75040, 75042, 75044, 75047, 75048, 75049, 75051, 75053, 75060, 75064, 75100, 75102, 75104, 75105, 750111, 750114",
      "center_classification_date": "20160630",
      "distribution_pattern": "Worldwide distribution to Belgium, China, Egypt, Germany, India, Korea, Republic of, Myanmar, Nigeria, Peru, Russian Federation, Rwanda, Singapore, Spain, Sweden, Taiwan, Thailand, and United Kingdom",
      "state": "OH",
      "product_description": "Big Bore Oncology tomography X-ray system scanner, model number 728243.   The Big Bore Oncology scanner is a whole body computed tomography X-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes",
      "report_date": "20160706",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Philips Medical Systems (Cleveland) Inc",
      "recall_number": "Z-2109-2016",
      "initial_firm_notification": "E-Mail",
      "product_type": "Devices",
      "event_id": "74507",
      "termination_date": "20180702",
      "more_code_info": "",
      "recall_initiation_date": "20160614",
      "postal_code": "44143-2131",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}