      "status": "Ongoing",
      "city": "Mounds View",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88175",
      "recalling_firm": "Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)",
      "address_1": "8200 Coral Sea St Ne",
      "address_2": "N/A",
      "postal_code": "55112-4391",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution",
      "recall_number": "Z-2102-2021",
      "product_description": "Medtronic Reveal LINQ LNQ11 / PA96000",
      "product_quantity": "666210 devices",
      "reason_for_recall": "Reveal LINQ with TruRhythm ICMs, that undergo a partial electrical reset appear to be programmed \"ON\", but are no longer able to detect and report Brady and Pause events to clinicians.",
      "recall_initiation_date": "20210601",
      "center_classification_date": "20210720",
      "report_date": "20210728",
      "code_info": "All serial numbers distributed on or after February 23, 2017       GTIN:  643169845763, 643169845770, 643169845787, 643169845794, 643169845800, 643169845817, 643169845824, 00643169845831, 00643169845848, 00643169845855, 00643169845862, 00643169845879, 00643169845893, 00763000188382, 00763000188399, 00763000188405, 00763000188412, 00763000188429, 00763000188436, 00763000188443, 00763000188450, 00763000188467, 643169845688, 643169845695, 643169845701, 643169845718, 643169845732, 643169845749, 643169845756",
      "more_code_info": ""
    }
  ]
}