  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Grove City",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79813",
      "recalling_firm": "Tosoh Bioscience Inc",
      "address_1": "3600 Gantz Rd",
      "address_2": "N/A",
      "postal_code": "43123-1895",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: US (Nationwide) and countries of; Canada, Mexico, Columbia, Chile, Uruguay, Nicaragua, Argentina, Peru, Chile, Honduras. Equador, Caymen Islands, Panama, and Costa Rica.",
      "recall_number": "Z-2099-2018",
      "product_description": "ST-AIA PACK Cystatin C; Part Number: 025217     Assay, Kidney Marker",
      "product_quantity": "1047",
      "reason_for_recall": "Asfotase Alfa (Strensiq) interferes with certain Tosoh assays which potentially causes falsely increased or falsely decreased immunoassay test results, depending upon the assay.",
      "recall_initiation_date": "20180305",
      "center_classification_date": "20180605",
      "termination_date": "20200831",
      "report_date": "20180613",
      "code_info": "All Lot Codes"
    }
  ]
}