  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Irvine",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "92493",
      "recalling_firm": "Pro-Dex Inc",
      "address_1": "2361 McGaw Ave",
      "address_2": "N/A",
      "postal_code": "92614-5831",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US: FL",
      "recall_number": "Z-2098-2023",
      "product_description": "KLS Martin Battery Pack, REF: KLS BP 001, and KLS BP 040 (bulk case of 40), used in conjunction with the KLS Martin MaxDriver",
      "product_quantity": "2000",
      "reason_for_recall": "Battery packs used in surgical procedures where a surgeon uses a powered screwdriver and a right-angle attachment have switched manufacturing and expiration dates on sterile pouch labels, so batteries will appear to be prematurely expired.",
      "recall_initiation_date": "20230215",
      "center_classification_date": "20230705",
      "report_date": "20230712",
      "code_info": "REF: KLS-BP-001/KLS-BP-040, UDI-DI: 00862377000140, Lot: K0M6U, Expiration: 8/31/2024"
    }
  ]
}