  "results": [
    {
      "status": "Terminated",
      "city": "Molnlycke",
      "state": "N/A",
      "country": "Sweden",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "69760",
      "recalling_firm": "Breas Medical AB",
      "address_1": "Foretagsvagen 1",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide - United States Nationwide Distribution and the countries: Argentina, Belgium, Chile, Cyprus, Denmark, Ecuador, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Kuwait, Lebanon, Malaysia, Netherlands, Norway, Panama, Poland, Portugal, Russia, Saudi Arabia, Serbia, Singapore, Slovenia, Spain, Sweden, Switzerland, Thailand, Turkey, UAE and United Kingdom.",
      "recall_number": "Z-2094-2015",
      "product_description": "Breas Vivo 50 Home Care Ventilator     Product Usage:  Respiratory Care",
      "product_quantity": "846 US",
      "reason_for_recall": "Unintended treatment termination could result from a keypad malfunction in some situations. The device erroneously interprets this as a Stop Treatment Instruction. An alarm will not sound, or be registered. Accessories and monitoring equipment connected to the Vivo 50 will stop functioning as the device enters stand-by mode.",
      "recall_initiation_date": "20141107",
      "center_classification_date": "20150716",
      "termination_date": "20160407",
      "report_date": "20150722",
      "code_info": "All Vivo 50 Ventilators (215000, 215016) with firmware version 2.04",
      "more_code_info": ""
    }
  ]
}