  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Hertford",
      "state": "",
      "country": "United Kingdom",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "77019",
      "recalling_firm": "Del Mar Reynolds Medical, Ltd.",
      "address_1": "1 2 Harforde Court John Tate Road",
      "address_2": "Business Park",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US distribution in FL, LA, MS, NC, NC, SC and WY and Internationally to Panama and United Kingdom.",
      "recall_number": "Z-2093-2017",
      "product_description": "Spacelabs Healthcare Arkon Anesthesia Delivery System is intended for use in the hospital environment and operating room.     Absorber Adapter Assembly Service Kit, part number: 050-9045-00.  Absorber Adapter Assembly, part number: 650-1648-01.  Arkon Absorber Canister Carrier, part number: 380-0916-01.",
      "product_quantity": "62 total (US= 19 devices  and 32 kits; International = 11 devices)",
      "reason_for_recall": "Reports of the Canister Carrier breaking during absorbent exchange or during anesthesia machine transit, causing a leak in the breathing circuit, which could cause hypoventilation in the patient.",
      "recall_initiation_date": "20170414",
      "center_classification_date": "20170516",
      "termination_date": "20171016",
      "report_date": "20170524",
      "code_info": "*********the serial numbers and part numbers of units distributed in the US *********    Absorber Adapter Assembly Service Kit part number: 050-9045-00.    Devices serial numbers: ARKN-000395, ARKN-000396, ARKN-000397, ARKN-000398, ARKN-000399, ARKN-000400, ARKN-000401, ARKN-000402, ARKN-000403, ARKN-000404, ARKN-000405, ARKN-000406, ARKN-000407, ARKN-000408, ARKN-000409, ARKN-000410, ARKN-000411, ARKN-000412, and ARKN-000413.    *********the serial numbers of units distributed international **************************    ARKN-000375, ARKN-000414, ARKN-000415, ARKN-000416, ARKN-000417, ARKN-000418, ARKN-000419, ARKN-000420, ARKN-000421, ARKN-000422, and ARKN-000423.",
      "more_code_info": ""
    }
  ]
}