  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Elyria",
      "address_1": "1200 Taylor St",
      "reason_for_recall": "M41 Powered Wheelchairs received through Invacare Service Parts (ISP) have controllers that were programmed in a way this is not aligned to standard factory settings.  The controller during different settings such as max forward speed, max forward acceleration, max reverse speed and max turning speed can result in momentary loss of control.",
      "address_2": "",
      "product_quantity": "819",
      "code_info": "P/N 1143839",
      "center_classification_date": "20140722",
      "distribution_pattern": "Nationwide Distriution: CA, SC, IL, UT, TN, KS, HI, NJ, LA, TX,FL, IN, AL, AZ, MN, WI, MN, MI, AR,MA, GA, MS, NJ, NM, OK, ME, PA, NY, NC, OK, MO, WA, IA, OR,KY, MD, NE, ID, VA, WA, ND, MT, OH, NH, CT.  Also shipped to Canada",
      "state": "OH",
      "product_description": "MK5 NX Controller--Wheelchair controller",
      "report_date": "20140730",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Invacare Corporation",
      "recall_number": "Z-2091-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "68622",
      "termination_date": "20141015",
      "more_code_info": "",
      "recall_initiation_date": "20140624",
      "postal_code": "44035-6248",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}