  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Mahwah",
      "address_1": "325 Corporate Drive",
      "reason_for_recall": "Stryker has updated the instruction for use (IFU) (product correction)for the ABGII Modular and Rejuvenate Modular Hip Systems.  This is based on a reported rate of less than one percent for revisions potentially associated with fretting and or corrosion at or about the modular neck junction. On June 29, 2012, Stryker has decided to remove the affected product from the field since there may be an",
      "address_2": "",
      "product_quantity": "(NLS/NLV) - 23193 units and (SPT) - 20,743 units",
      "code_info": "510 K092561  all lots  Rejuvenate  (NLS/NLV)  Rejuvenate (SPT)  (4845-4XXX and NLS/NLV)",
      "center_classification_date": "20120726",
      "distribution_pattern": "Nationwide Distribution",
      "state": "NJ",
      "product_description": "Rejuvenate Modular Stems   Stryker Ireland Carrigtwohill Industrial Estate  Carrigtwohill County Cork, Ireland    Rejuvenate Modular Necks Stryker Howmedica Osteonics Corp. Mahwah, NJ 07430.    Stem Bodies Rejuvenate Modular consists of 6 modular stems bodies in sizes 7-12. The Modular Stem bodies are made of TMZF alloy, developed specifically for orthopaedic applications. Rejuvenate Modular Necks The Rejuvenate Modular Necks are made of Cobalt Chrome. The necks in this system consist of 4 lengths in 4 mm increments. They are 30 mm, 34 mm, 38 mm and 42 mm. They come in multiple neck angles and versions in each length. They are 127, 130 and 132. The versions are 0, 8, 16 degrees. In total there are 16 unique neck implants in the Rejuvenate Modular System.",
      "report_date": "20120801",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Stryker Howmedica Osteonics Corp.",
      "recall_number": "Z-2090-2012",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "62456",
      "more_code_info": "",
      "recall_initiation_date": "20120430",
      "postal_code": "07430",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Ongoing"
    }
  ]
}