  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Mahwah",
      "address_1": "325 Corporate Drive",
      "reason_for_recall": "Stryker has updated the instruction for use (IFU) (product correction)for the ABGII Modular and Rejuvenate Modular Hip Systems.  This is based on a reported rate of less than one percent for revisions potentially associated with fretting and or corrosion at or about the modular neck junction. On June 29, 2012, Stryker has decided to remove the affected product from the field since there may be an",
      "address_2": "",
      "product_quantity": "9003 units",
      "code_info": "510 K092406    all lots    (4845-4XXX and NLS/NLV)",
      "center_classification_date": "20120726",
      "distribution_pattern": "Nationwide Distribution",
      "state": "NJ",
      "product_description": "ABG II Modular Hip Stem  Stryker Howmedica Osteonics Corp. Mahwah, NJ 07430  Stryker Ireland Carrigtwohill Industrial Estate, Carrigtwohill County Cork, Ireland.    The ABG II Modular range of implants is constituted of 8 right stems and 8 left stems. The necks are made of GADS Vitallium, a proprietary Co Cr alloy that Stryker has developed. GADS stands for Gas Atomized Dispersion Strengthened. IT is a Co Cr alloy that is a corrosion resistant and demonstrates qualities of improved fatigue strength (with respect to regular Co Cr alloy). Neck Range - The ABG II modular neck range is composed of 10 reversible neck implants. There are different options in terms of length, version and neck angle. The AG II Modular necks come in two different lengths: 28 mm (short) and 36 mm (long). The necks are offered in two version angles ; 0 and 7. The 0 and 7 necks are available in 125, 130 and 135 neck angles.",
      "report_date": "20120801",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Stryker Howmedica Osteonics Corp.",
      "recall_number": "Z-2089-2012",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "62456",
      "more_code_info": "",
      "recall_initiation_date": "20120430",
      "postal_code": "07430",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Ongoing"
    }
  ]
}