  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "West Chester",
      "address_1": "1302 Wrights Ln E",
      "reason_for_recall": "The Guide Blocks for the 2 Column Plate 6 Hole Head (Volar Distal Radius Plate System) were found to be labeled incorrectly. Part no. 03.111.600 (Right) was found in the package labeled  part no. 03.111.601 (Left).  The laser etching denoting the orientation on the part is correct.",
      "address_2": "",
      "product_quantity": "180",
      "code_info": "part no. 03.111.600, lot no. 10-5147, and part no. 03.111.601, lot no. 10-5147",
      "center_classification_date": "20140721",
      "distribution_pattern": "Distribution US Nationwide and Bermuda.",
      "state": "PA",
      "product_description": "Synthes Guide Blocks for the 2 Column Plate 6 Hole Head (Volar Distal Radius Plate System).  Orthopedic surgical instrument.",
      "report_date": "20140730",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Synthes, Inc.",
      "recall_number": "Z-2087-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "68442",
      "termination_date": "20160818",
      "more_code_info": "",
      "recall_initiation_date": "20140529",
      "postal_code": "19380-3417",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}