  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Carlsbad",
      "address_1": "5818 El Camino Real",
      "reason_for_recall": "The firm initiated this recall of certain lots of Illico Ti Cannulated Polyaxial Screws 7.5mm x 40mm implants, due to an off-center cannulation hole at the proximal end of the bone screw. There is a potential risk that when the screw is mated to a polyaxial screwdriver, the guide wire may not pass through the screwdriver.",
      "address_2": "",
      "product_quantity": "333 screws",
      "code_info": "Part Number 73875-40.  Lots Affected:   647289,  648923,  649244,  649556,  649813,  650040,  650262,  650589.",
      "center_classification_date": "20130827",
      "distribution_pattern": "Worldwide distribution: US (nationwide) including states of:US states: AZ, CA, FL, GA, IL, MD, NJ, NY, OK,  PA, TN, TX,and WA ; and countries of: Spain, France, Japan.  International: Spain, France, Japan.",
      "state": "CA",
      "product_description": "Illico Ti Cannulated Polyaxial Screw 7.5mm X 40mm. The Illico MIS Posterior Fixation System is intended to facilitate the surgical correction of noncervical spinal deformities by providing temporary internal fixation and stabilization during bone graft healing and/or fusion mass development.",
      "report_date": "20130904",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Alphatec Spine, Inc.",
      "recall_number": "Z-2086-2013",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "product_type": "Devices",
      "event_id": "65672",
      "termination_date": "20141009",
      "more_code_info": "",
      "recall_initiation_date": "20130628",
      "postal_code": "92008-8816",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}