  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Brea",
      "address_1": "250 S Kraemer Blvd",
      "reason_for_recall": "Beckman Coulter has identified the following drugs as potential sources of interference: Metamizole (Dipyrone), Sulfasalazine, and Sulfapyridine and may result in false low results.",
      "address_2": "",
      "product_quantity": "50,645 units total (43,816 units in US)",
      "code_info": "all lots",
      "center_classification_date": "20170512",
      "distribution_pattern": "Worldwide Distribution -- Australia, Bangladesh, Belgium, Brunei, Bulgaria, Canada, China, Colombia, Czech Republic, Egypt, Eritrea, Finland, France, Germany, Ghana, Greece, Guatemala, Hong Kong, India, Ireland, Italy, Japan, Jordan, Kuwait, Lebanon, Lithuania, Macau, Malaysia, Mayotte, Mexico, Netherlands, New Zealand, Oman, Pakistan, Panama, Paraguay, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland,  Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, United States including Puerto Rico, and Vietnam.",
      "state": "CA",
      "product_description": "Salicylate Reagent (SALY 2 x 45), Catalog No. 378194.    Intended for quantitative determination of salicylate concentration in human serum or plasma.",
      "report_date": "20170524",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Beckman Coulter Inc.",
      "recall_number": "Z-2084-2017",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "76962",
      "termination_date": "20180531",
      "more_code_info": "",
      "recall_initiation_date": "20170403",
      "postal_code": "92821-6232",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}