  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Plainsboro",
      "address_1": "311 Enterprise Dr",
      "reason_for_recall": "The affected K-wires do not meet specification requirements. Specifically, a) K-wires were made of 316LVM stainless steel instead of 304 stainless steel. b) K-wires have passivated surface finish instead of electro-polish surface finish. c) K-wires have incorrect length and diameter.",
      "address_2": "",
      "product_quantity": "182 units",
      "code_info": "Lot 237830TE",
      "center_classification_date": "20160628",
      "distribution_pattern": "Nationwide Distribution to OK, CA, TX, WA, GA, NY, MS, NY, PA",
      "state": "NJ",
      "product_description": "K-wire, Smooth Double-pointed 0.8 x 90mm, a component of the METASURG Capture Screw System   The K-wire is used in conjunction with the 2.0 mm cannulated screws for the METASURG Capture Screw System. The K-wire is used by the physician to insert into the bone and to act as a guide for drilling, countersinking and installing screws",
      "report_date": "20160706",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Integra LifeSciences Corp.",
      "recall_number": "Z-2084-2016",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "74224",
      "termination_date": "20161116",
      "more_code_info": "",
      "recall_initiation_date": "20160520",
      "postal_code": "08536-3344",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}