  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Feldkirchen",
      "state": "",
      "country": "Germany",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65773",
      "recalling_firm": "Brainlab AG",
      "address_1": "Kapellenstrasse 12",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Distribution only to AR., and the countries of Australia, Germany and Japan.",
      "recall_number": "Z-2083-2013",
      "product_description": "BRAINLAB; FRAMELESS SRS QA TARGET POINTER    Robotics is a device used to compensate rotational patient misalignment (roll and pitch) in a linear accelerator environment for stereotactic radiosurgery or radiotherapy procedures. The Frameless Radiosurgery Components are a device used for fixation, localization and repositioning of the patient's: head and neck; and head, neck, and shoulders, in a linear accelerator environment for stereotactic radiosurgery/radiotherapy procedures.",
      "product_quantity": "6 units",
      "reason_for_recall": "The Frameless SRS QA Target Pointer - Pointer Cap with the engraved cross hairs may become loose even when only a moderate force is applied to the Pointer Cap.",
      "recall_initiation_date": "20130716",
      "center_classification_date": "20130826",
      "termination_date": "20170313",
      "report_date": "20130904",
      "code_info": "Part No. 43410",
      "more_code_info": ""
    }
  ]
}