      "status": "Terminated",
      "city": "Schwabmunchen",
      "state": "N/A",
      "country": "Germany",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83358",
      "recalling_firm": "Medical Intelligence Medizintechnik Gmbh",
      "address_1": "Robert-bosch-str. 8",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide Distribution - US Nationwide",
      "recall_number": "Z-2082-2019",
      "product_description": "IGuide System, software versions: 2.2.0, 2.2.1, 2.2.2    Product Usage:    Linear Accelerator. The intended use of the device is the control of accurate patient positioning with assistance of a 30 Tracking System in a radiotherapy environment.",
      "product_quantity": "496 units",
      "reason_for_recall": "Sometimes, although the HexaPOD has reached the target position, the Constant Correlation Check (CCC) can be too sensitive and iGUIDE may incorrectly demand a verification scan.",
      "recall_initiation_date": "20190718",
      "center_classification_date": "20190725",
      "termination_date": "20230811",
      "report_date": "20190731",
      "code_info": "UDI (GTIN)  iGuide 2.2.0 = 04056719001704  iGuide 2.2.1 = 04056719001742   iGuide 2.2.2 = 04056719002039    Installed on the following units:  Master Devices with serial numbers:  VERSA HD/30002624-003/154166  VERSA HD/30000782-003/153788  VERSA HD/30000782-003/153788  VERSA HD/30000782-003/153788  ELEKTA INFINITY/19175-003/154363  VERSA HD/30001056-002/109016  VERSA HD/30001009-001/153427  ELEKTA INFINITY/11766-003/152570  VERSA HD/11766-004/153649  ELEKTA SYNERGY/11766-002/151717  VERSA HD/17114-001/153383  VERSA HD/19677-005/109018  VERSA HD/30003910-LIN001/109005  VERSA HD/30003903-LIN001/109006  VERSA HD/30003903-LIN002/109035  VERSA HD/30001230-003/153428  VERSA HD/15081-LIN003/109003  ELEKTA SYNERGY/15081-002/152131  VERSA HD/12486-003/109036  VERSA HD/15240-003/109033  ELEKTA SYNERGY/15240-001/152128  ELEKTA SYNERGY/15240-002/152419  ELEKTA SYNERGY/20967-001/153012  VERSA HD/30003249-001/154299  VERSA HD/14756-002/153003  ELEKTA SYNERGY/14756-001/151916  ELEKTA SYNERGY/12579-003/152379  ELEKTA AXESSE/18225-001/152691  VERSA HD/18225-003/153900  VERSA HD/30001349-001/154087  VERSA HD/20965-001/153008  VERSA HD/30002428-LIN001/154024  ELEKTA AXESSE/11654-003/152617  VERSA HD/11654-006/109013  VERSA HD/10394-AKHEE/153248  VERSA HD/12570-#001/154395  VERSA HD/12570-#002/154435  VERSA HD/12490-KLG02E/154043  VERSA HD/12490-KLG04/153699  VERSA HD/17152-#001/156066  ELEKTA SYNERGY/10266-IBK02E/152807  VERSA HD/10266-IBK05/153055  VERSA HD/10266-IBK01E/154680  ELEKTA SYNERGY/10266-IBK04/151431  VERSA HD/11517-C8230/154221  VERSA HD/11517-C8220/154473  ELEKTA SYNERGY/11517-C8210/151888  ELEKTA SYNERGY/11517-C8200/151483  VERSA HD/10587-001/154730  VERSA HD/16805-002/153976  VERSA HD/30003118-002/154744  VERSA HD/30003118-001/154743  VERSA HD/30003118-003/154745  VERSA HD/30003118-004/154746  ELEKTA INFINITY/14770-003/153005  iGUIDE Software/14770-001/2.2.2  ELEKTA INFINITY/12473-012/152957  ELEKTA INFINITY/12473-013/152956  ELEKTA INFINITY/15012-008/152853  ELEKTA INFINITY/15012-007/153015  ELEKTA INFINITY/13219-004/153852  ELEKTA INFINITY/12274-021/153049  ELEKTA INFINITY/14771-001/152692  ELEKTA INFINITY/14771-002/152979  VERSA HD/20491-001/154325  ELEKTA INFINITY/12678-003/154786  VERSA HD/20531-001/154705  ELEKTA INFINITY/15291-002/154761  ELEKTA INFINITY/12709-002/154790  VERSA HD/30003232-001/154528  iGUIDE Software/30003232-001/2.2.2  ELEKTA INFINITY/30004215-001/154734  VERSA HD/30002674-003/154149  ELEKTA INFINITY/13128-002/154630  ELEKTA INFINITY/20612-001/154775  VERSA HD/19096-001/154701  VERSA HD/30003818-001/154760  ELEKTA AXESSE/30001262-001/154969  ELEKTA INFINITY/12659-003/154894  ELEKTA INFINITY/21118-002/154423  ELEKTA INFINITY/13431-008/154806  ELEKTA INFINITY/17556-002/154977  VERSA HD/12634-003/154900  ELEKTA INFINITY/30000925-002/154619  ELEKTA INFINITY/30003865-001/154605  VERSA HD/19919-002/154421  ELEKTA INFINITY/17128-004/154475  VERSA HD/12668-005/154551  VERSA HD/12668-005/154551  ELEKTA INFINITY/19137-002/154699  ELEKTA SYNERGY/13123-002/154715  VERSA HD/30003134-L14501/154398  ELEKTA SYNERGY/13166-ZLI002/154005  ELEKTA SYNERGY/10562-008/151787  ELEKTA SYNERGY/10562-004/151666  VERSA HD/10562-009/152843  VERSA HD/10562-010/152912  VERSA HD/10419-007/154808  VERSA HD/30003123-001/154324  VERSA HD/10102-001/154723  VERSA HD/12500-004/154198  VERSA HD/12510-003/153876  IGUIDE¿ 2.2.2 FCO MOD KIT/12523-003/2.2.2  VERSA HD/12525-008/154077  VERSA HD/12525-009/154981  ELEKTA SYNERGY/14785-001/152376  VERSA HD/12508-003/154550  VERSA HD/11666-004/153104  VERSA HD/15250-002/153872  VERSA HD/11868-012/156052  VERSA HD/12538-003/153836  VERSA HD/18578-001/154258  VERSA HD/18578-002/154449  VERSA HD/18578-005/154807  VERSA HD/12457-003/154148  VERSA HD/12537-004/109028  VERSA HD/10278-002/156022  VERSA HD/10278-001/154947  ELEKTA SYNERGY/13185-F06501/151852  ELEKTA AXESSE/10795-K08501/152366  ELEKTA SYNERGY/11863-B03601/152069  VERSA HD/10376-002/153711  VERSA HD/10376-003/153706  ELEKTA SYNERGY/10194-001/153580  ELEKTA SYNERGY/10993-M06501/152417  ELEKTA SYNERGY/10993-M06601/152565  VERSA HD/10993-M06101/153359  VERSA HD/18404-D07701/153106  VERSA HD/11488-S07701/153279  VERSA HD/11735-D04701/154057  ELEKTA SYNERGY/18480-002/153856  VERSA HD/30001101-S17501/153249  VERSA HD/30003074-K12502/154383  ELEKTA SYNERGY/12455-A01701/152630  VERSA HD/12455-A01801/154530  ELEKTA SYNERGY/11153-A03601/151455  ELEKTA SYNERGY/11153-A03501/151429  VERSA HD/13192-E03502/154750  IGUIDE¿ 2.2.2 FCO MOD KIT/13192-E03501/2.2.2  VERSA HD/18474-002/154774  VERSA HD/18474-003/154982  VERSA HD/10719-T01701/153765  ELEKTA SYNERGY/10719-T01017/151846  ELEKTA SYNERGY/11653-W02501/151009  ELEKTA SYNERGY/11653-W02601/152584  VERSA HD/30004402-LIN001/154815  ELEKTA SYNERGY/11095-003/151576  VERSA HD/30003361-LIN001/154369  VERSA HD/10125-V-SELT/154167  VERSA HD/10125-V-SIGN/154226  iGUIDE Software/ACCOUNT-LOCATION/2.2.0  ELEKTA SYNERGY/12456-004/152779  VERSA HD/30003351-001/154306  VERSA HD/18650-001/154593  VERSA HD/30003382-001/154257  VERSA HD/30003225-001/154285  iGUIDE Software/ACCOUNT-LOCATION/2.2.0  ELEKTA AXESSE/18649-001/152519  VERSA HD/30003199-001/154225  VERSA HD/30003223-001/154320  VERSA HD/30002974-001/153455  VERSA HD/30001991-001/154412  VERSA HD/18280-002/154792  VERSA HD/30004177-001/154615  VERSA HD/30003093-001/154400  VERSA HD/30003093-002/154401  ELEKTA INFINITY/30003093-004/154403  VERSA HD/30003030-001/154233  VERSA HD/30003838-001/154541  VERSA HD/16686-002/154214  VERSA HD/16809-002/154839  ELEKTA SYNERGY/30001993-001/153533  VERSA HD/18630-002/154875  VERSA HD/30001597-001/154462  VERSA HD/30004346-001/154768  VERSA HD/10930-001/154142  ELEKTA SYNERGY/13206-002/152460  VERSA HD/30004237-LIN001/154714  VERSA HD/17373-LIN002/154284  iGUIDE Software/ACCOUNT-LOCATION/2.2.0  VERSA HD/18607-005/154990  VERSA HD/11755-005/154500  VERSA HD/11755-004/154279  VERSA HD/30003951-001/154598  ELEKTA SYNERGY/10924-002/151550  HexaPOD evo Module/11319-004/118046  VERSA HD/11320-004/156026  VERSA HD/11005-002/154236  VERSA HD/12576-003/154548  iGUIDE Software/12576-003/2.2.2  VERSA HD/10024-003/154984  VERSA HD/17416-002/154901  VERSA HD/12469-005/154174  VERSA HD/18753-001/154666  VERSA HD/17253-002/154824  VERSA HD/30000525-001/109023  VERSA HD/19242-003/154845  VERSA HD/30001314-001/154653  VERSA HD/30001635-002/154013  VERSA HD/17301-002/154629  VERSA HD/18743-001/153978  ELEKTA SYNERGY/10626-002/152895  VERSA HD/18784-002/156065  VERSA HD/18784-001/154930  ELEKTA SYNERGY/30000468-LIN001/154241  ELEKTA SYNERGY/30000468-LIN001/154241  ELEKTA INFINITY/15903-ONC001/153869  VERSA HD/16572-ONC001/153689  ELEKTA SYNERGY/16572-ONC002/153710  iGUIDE Software/30000332-ONC001/2.2.2  ELEKTA SYNERGY/16553-ONC001/152505  VERSA HD/30000330-ONC001/153905  VERSA HD/11612-LIN001/154254  ELEKTA INFINITY/15719-LIN001/154572  ELEKTA SYNERGY/16208-ONC001/153201  ELEKTA SYNERGY/13300-ONC001/151832  VERSA HD/13300-LIN002/154832  VERSA HD/12999-LIN001/154840  ELEKTA INFINITY/15940-LIN001/154159  VERSA HD/30000240-LIN001/154827  ELEKTA INFINITY/19930-LIN001/154209  ELEKTA INFINITY/16148-ONC001/154134  ELEKTA SYNERGY/30000390-LIN001/154256  ELEKTA SYNERGY/13324-ONC001/152762  VERSA HD/12600-LIN001/154436  VERSA HD/16194-LIN001/154849  Versa HD/16310-LIN002/109024  ELEKTA SYNERGY/16310-LIN001/154441  ELEKTA INFINITY/16222-ONC001/153677  ELEKTA SYNERGY/30000494-LIN001/154220  VERSA HD/15753-LIN001/154694  VERSA HD/16375-LIN001/154172  VERSA HD/12974-LIN002/154534  ELEKTA SYNERGY/15807-LIN001/154770  ELEKTA INFINITY/30000387-ONC001/153646  ELEKTA INFINITY/30000387-ONC001/153646  ELEKTA SYNERGY/16145-ONC001/152351  ELEKTA INFINITY/30000349-ONC001/153236  ELEKTA AXESSE/30000380-ONC001/153204  ELEKTA AXESSE/30000380-ONC001/153204  ELEKTA INFINITY/19933-LIN001/154208  ELEKTA SYNERGY/16386-LIN001/154782  VERSA HD/12266-LIN001/154481  ELEKTA AXESSE/30000602-ONC001/153102  ELEKTA SYNERGY/15933-LIN001/154326  ELEKTA SYNERGY/15933-LIN001/154326  VERSA HD/16225-ONC001/153998  IGUIDE¿ 2.2.2 FCO MOD KIT/10425-LIN001/2.2.2  VERSA HD/10425-LIN001/154628  ELEKTA INFINITY/15954-LIN001/154766  ELEKTA INFINITY/30000477-LIN001/153982  VERSA HD/30003036-LIN001/153776  ELEKTA SYNERGY/15614-ONC001/152260  VERSA HD/10375-ONC002/154679  VERSA HD/16578-LIN001/154812  VERSA HD/15861-LIN001/154158  ELEKTA SYNERGY/30002370-LIN001/154196  ELEKTA SYNERGY/11390-ONC001/154157  ELEKTA SYNERGY/17389-ONC001/152624  iGUIDE Software/17389-ONC001/2.2.2  VERSA HD/30000917-LIN001/154648  ELEKTA INFINITY/16262-LIN001/154213  VERSA HD/16267-LIN001/154200  ELEKTA SYNERGY/12749-ONC001/151596  VERSA HD/30001318-LIN001/154532  ELEKTA SYNERGY/15760-ONC001/154160  ELEKTA INFINITY/15916-LIN003/154516  Versa HD/30000498-LIN001/109039  VERSA HD/15850-LIN001/154197  VERSA HD/16403-LIN001/109032  ELEKTA SYNERGY/16366-LIN001/154538  ELEKTA SYNERGY/16730-LIN001/154223  VERSA HD/16730-LIN002/154222  VERSA HD/11686-007/154291  VERSA HD/11745-LIN001/154588  VERSA HD/30004417-LIN001/154846  VERSA HD/19028-LIN001/154919  VERSA HD/15471-LIN001/154492  VERSA HD/11998-LIN001/154430  ELEKTA INFINITY/30003850-LIN001/154560  VERSA HD/17033-LIN008/154751  ELEKTA INFINITY/17033-006/153294  VERSA HD/40021-LIN002/156025  VERSA HD/30004435-LIN001/154939  VERSA HD/17260-LIN002/154624  ELEKTA INFINITY/30002792-001/153748  ELEKTA INFINITY/30003292-001/154467  VERSA HD/12532-005/154239  VERSA HD/11461-L6new/154731  VERSA HD/11461-L8new/154511  VERSA HD/11771-AntHL2/154150  VERSA HD/11014-UMCU7N/154795  ELEKTA AXESSE/11014-UMCU15/152754  ELEKTA SYNERGY/13058-002/152161  ELEKTA SYNERGY/13058-003/152455  VERSA HD/13058-004/109012  VERSA HD/30000993-LIN001/109004  VERSA HD/30000993-LIN002/109017  VERSA HD/14926-004/154039  VERSA HD/20991-002/153163  VERSA HD/20991-001/153130  VERSA HD/14808-LIN001/109015  ELEKTA SYNERGY/13351-001/151966  VERSA HD/11412-004/154836  iGUIDE Software/12929-OPO801/2.2.2  VERSA HD/12929-LIN001/154890  VERSA HD/30003526-501/154686  VERSA HD/30003526-601/154716  iGUIDE Software/17176-KIE501/2.2.2  VERSA HD/17176-LIN001/154888  VERSA HD/12619-WAR801/154778  VERSA HD/12619-WAR901/154911  VERSA HD/30003322-WIM501/154251  VERSA HD/30003322-WIM601/154252  ELEKTA INFINITY/30003322-LIN001/154889  VERSA HD/12439-003/154763  ELEKTA INFINITY/18974-501/154837  VERSA HD/13041-LIN001/154921  VERSA HD/18959-LIN001/154975  VERSA HD/12475-LJU08/153679  IGUIDE¿ 2.2.2 FCO MOD KIT/30003117-001/2.2.2  VERSA HD/30003117-001/154438  VERSA HD/30003263-001/154935  ELEKTA INFINITY/12558-002/153631  IGUIDE¿ 2.2.2 FCO MOD KIT/30003724-001/2.2.2  VERSA HD/18521-002/154732  VERSA HD/18524-001/154681  VERSA HD/18527-001/154373  VERSA HD/30004046-001/154596  ELEKTA INFINITY/30001015-002/154468  VERSA HD/10708-001/154687  VERSA HD/18509-001/154513  VERSA HD/11261-003/154913  VERSA HD/10860-004/154124  VERSA HD/11071-005/154544  VERSA HD/13373-002/156004  VERSA HD/18534-001/154777  VERSA HD/30002863-001/154333  VERSA HD/10142-001/154706  VERSA HD/15020-04/154219  VERSA HD/15020-03/154529  iGUIDE Software/15020-04/2.2.2  VERSA HD/10157-013/154867  VERSA HD/10157-014/154862  VERSA HD/19658-RIV01/152951  VERSA HD/30004475-LIN001/154971  VERSA HD/10283-LIN002/154929  VERSA HD/16913-001/154739  VERSA HD/11487-LIN004/154542  ELEKTA INFINITY/30001447-LIN002/154780  VERSA HD/16909-LIN003/154406  VERSA HD/10498-LIN003/154678  VERSA HD/10620-LIN001/154811  VERSA HD/10620-LIN002/154794  iGUIDE Software/30003464-001/2.2.2.0  VERSA HD/30003464-001/154456  VERSA HD/30004284-001/154738  VERSA HD/30002988-001/154112  VERSA HD/30002988-001/154112  VERSA HD/30003000-001/154682  VERSA HD/30004023-001/154576  VERSA HD/10472-010/154354  IGUIDE¿ 2.2.2 FCO MOD KIT/13271-014/2.2.2  VERSA HD/13271-015/154689  ELEKTA SYNERGY/11407-011/152890  VERSA HD/10886-021/156063  ELEKTA SYNERGY/10886-012/151827  ELEKTA SYNERGY/11630-012/153212  ELEKTA SYNERGY/11489-016/151556  VERSA HD/30002724-01/154207  VERSA HD/30002724-03/154494  VERSA HD/30002724-02/154647  VERSA HD/11637-012/154347  VERSA HD/11637-013/154702  VERSA HD/30002401-001/154063  ELEKTA SYNERGY/10622-001/152822  VERSA HD/17023-001/154886  VERSA HD/12873-003/154755  VERSA HD/30000021-002/154843  VERSA HD/14117-001/153376  VERSA HD/14117-003/154559  VERSA HD/14117-004/154709  VERSA HD/11746-012/154426  ELEKTA INFINITY/13991-001/153072  ELEKTA INFINITY/11090-006/154244  VERSA HD/30000095-002/154697  VERSA HD/10434-003/153752  VERSA HD/13570-002/153847  VERSA HD/13570-003/154193  VERSA HD/11876-007/154501  VERSA HD/11876-008/154887  iGUIDE Workstation/14147-RO00/CLHD7T2  ELEKTA INFINITY/13851-001/154604  ELEKTA INFINITY/13851-002/154204  ELEKTA INFINITY/13852-001/154422  ELEKTA INFINITY/30000354-001/153342  ELEKTA INFINITY/30000354-002/153536  ELEKTA INFINITY/13571-001/154916  ELEKTA INFINITY/13523-001/154206  VERSA HD/10451-004/154105  VERSA HD/11923-003/154413  VERSA HD/30003099-001/154372  VERSA HD/30004313-001/154879  VERSA HD/30004307-001/154892  ELEKTA INFINITY/30000702-001/153164  ELEKTA INFINITY/30000149-001/152801  ELEKTA INFINITY/30000150-001/152970  ELEKTA SYNERGY/13177-001/151969  ELEKTA SYNERGY/14568-001/151782  VERSA HD/14261-002/154733  IGUIDE¿ 2.2.2 FCO MOD KIT/98300-001/2.2.2  VERSA HD/14386-001/154280  VERSA HD/11635-002/154388  VERSA HD/11707-007/153991  VERSA HD/30004075-001/154613  VERSA HD/10817-010/153986  VERSA HD/10559-002/153340  VERSA HD/19465-001/154111  ELEKTA SYNERGY/12776-001/151722  VERSA HD/10064-010/153997  ELEKTA INFINITY/15053-002/152792  VERSA HD/30004347-002/153841  VERSA HD/30004347-002/153841  VERSA HD/10255-009/153843  ELEKTA INFINITY/30004150-001/154640  ELEKTA INFINITY/30004150-002/154708  ELEKTA INFINITY/10877-008/152732  ELEKTA INFINITY/10877-009/152833  ELEKTA SYNERGY/12041-002/151714  ELEKTA INFINITY/17096-001/154117  iGUIDE Workstation/30001324-001/JMRX3J2  VERSA HD/13120-008/153187  ELEKTA INFINITY/14131-001/154069  VERSA HD/14402-002/154719  VERSA HD/13728-003/154070  ELEKTA SYNERGY/10632-006/152468  VERSA HD/10632-004/153890  VERSA HD/10632-007/154710  ELEKTA INFINITY/13272-002/154182  VERSA HD/13267-002/153466  ELEKTA INFINITY/12131-002/152798  ELEKTA SYNERGY/12131-001/151410  VERSA HD/14175-004/154323  VERSA HD/14175-004/154323  ELEKTA AXESSE/11798-008/151952  ELEKTA SYNERGY/10303-005/153552  VERSA HD/30000068-001/154852  ELEKTA INFINITY/13777-002/153826  ELEKTA INFINITY/13797-001/152371  ELEKTA INFINITY/12015-003/154132  ELEKTA INFINITY/13725-001/154370  ELEKTA SYNERGY/30000194-002/151774  Versa HD/10228-004/154793  ELEKTA INFINITY/13866-001/152158  VERSA HD/13866-003/154217  ELEKTA INFINITY/15087-001/152392  ELEKTA SYNERGY/15087-003/153867  ELEKTA INFINITY/14707-001/154051  VERSA HD/10150-006/154711  VERSA HD/30000680-001/153295  ELEKTA AXESSE/11944-003/151793  VERSA HD/11944-004/153903  VERSA HD/30003708-001/154601  VERSA HD/12115-004/154190  VERSA HD/11193-003/154418  VERSA HD/10691-002/153858  VERSA HD/10342-005/154185  ELEKTA INFINITY/30003193-002/154986  VERSA HD/30002646-001/154085  VERSA HD/13546-001/154339  VERSA HD/13546-002/154340  ELEKTA INFINITY/15473-001/152200  ELEKTA SYNERGY/13314-002/152116  ELEKTA SYNERGY/11441-020/151412  VERSA HD/13359-003/154420  VERSA HD/13359-004/154419  ELEKTA SYNERGY/10356-008/152307  VERSA HD/10356-011/153897  ELEKTA SYNERGY/13212-002/151882  VERSA HD/14593-002/154243  VERSA HD/11017-008/154518  VERSA HD/11017-009/154908  VERSA HD/11638-007/154431  VERSA HD/30000079-001/154178  ELEKTA SYNERGY/13572-003/151885  ELEKTA AXESSE/10193-008/151812  VERSA HD/11076-003/153832  VERSA HD/12099-003/154046  VERSA HD/12099-004/153625  ELEKTA SYNERGY/10714-011/152522  VERSA HD/10492-LIN003/154349  ELEKTA SYNERGY/10492-LIN002/154356  VERSA HD/10492-LIN003/154349  ELEKTA INFINITY/16922-LIN002/154762  ELEKTA INFINITY/19112-LIN002/154265  VERSA HD/19112-LIN005/154925"
    }
  ]
}