  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Completed",
      "city": "Tempe",
      "state": "AZ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82513",
      "recalling_firm": "Bard Peripheral Vascular Inc",
      "address_1": "1625 W 3rd St Bldg 1",
      "address_2": "N/A",
      "postal_code": "85281-2438",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "US Nationwide Distribution in the states of FL, CO, WI, OH, IL, KS, AZ, WV, CT, PA, AL, MI and VA",
      "recall_number": "Z-2080-2019",
      "product_description": "POWERPORT CLEARVUE isp, 6F CHRONOFLEX Catheter, Intermediate NAUTILUS DELTA  Kit;   POWERPORT CLEARVUE isp, 8F CHRONOFLEX Catheter, Intermediate NAUTILUS DELTA Kit;   POWERPORT CLEARVUE Slim, 6F CHRONOFLEX Catheter and Suture Plugs, Intermediate NAUTILUS DELTA Kit;   POWERPORT CLEARVUE Slim, 8F CHRONOFLEX Catheter and Suture Plugs, Intermediate NAUTILUS DELTA Kit;   NAUTILUS DELTA Tip Confirmation System (Includes: Netbook with Pre-loaded  Software, Patient Module, and ECG Cable)    Product Usage:  Is an implantable access device designed to provide repeated access to the vascular system.",
      "product_quantity": "95 units",
      "reason_for_recall": "Certain NAUTILUS DELTA Port Kit product code / lot number combinations may be at risk of experiencing a lack of or a loss of the intravascular ECG signal during port placement procedures when used in combination with the NAUTILUS DELTA\" TCS.",
      "recall_initiation_date": "20180625",
      "center_classification_date": "20190725",
      "report_date": "20190731",
      "code_info": "REBR1147, REBR1148, REBR1149, REBT0471, RECQ2162 and REBR1150"
    }
  ]
}