  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-05-06",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Rochester",
      "address_1": "1777 E Henrietta Rd",
      "reason_for_recall": "Getinge Disinfection AB is voluntarily correcting Washer Disinfectors, Getinge Electrically Heated 9100-Series manufactured between 2009-05-04 through 09-26-2011. This action is based on a report from the ground fault breaker manufacturer of a production error.",
      "address_2": "",
      "product_quantity": "8 washers",
      "code_info": "Serial numbers: W50024258; W50024259; W50026573; W50025790; W50017381; W50020296; W50028138; W50020952",
      "center_classification_date": "20120726",
      "distribution_pattern": "Worldwide Distribution - USA including Maryland, Indiana, New Mexico, Massachuesetts, Texas, Kansas and the country of Canada",
      "state": "NY",
      "product_description": "Getinge brand --- large capacity, floor loading jet spray, washers for mechanical washing, intermediate level, thermal disinfection and drying of moisture and temperature stable Hospital case carts, sterilization containers, material handling carts, stands and utensils. --- Model numbers Getinge 9120, 9122, 9125, and 9128",
      "report_date": "20120801",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Getinge USA Inc",
      "recall_number": "Z-2079-2012",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "61974",
      "termination_date": "20160518",
      "more_code_info": "",
      "recall_initiation_date": "20120525",
      "postal_code": "14623-3133",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}