  "results": [
    {
      "status": "Terminated",
      "city": "Snoqualmie",
      "state": "WA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "68611",
      "recalling_firm": "Spacelabs Healthcare Inc",
      "address_1": "35301 SE Center St",
      "address_2": "N/A",
      "postal_code": "98065-9216",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US in the state of Georgia and in the countries of :  Malaysia, Paraguay, and France.",
      "recall_number": "Z-2077-2014",
      "product_description": "Spacelabs Healthcare Ultraview SL Multigas Module, Model 92518 (Software V1.00.00-09).  Provides a means of monitoring a variety of gas concentrations and alert clinical personnel when the concentration of anesthetic agent(s), oxygen, carbon dioxide or nitrous oxide moves outside of user-defined limits.",
      "product_quantity": "8 units total (3 units in US)",
      "reason_for_recall": "With this software version, the Minimum Alveolar Concentration (MAC) value displayed is extremely high (20.1) and physiologically invalid.",
      "recall_initiation_date": "20140618",
      "center_classification_date": "20140717",
      "termination_date": "20141208",
      "report_date": "20140723",
      "code_info": "Serial Numbers:  2518-000183         2518-000184         2518-000185         2518-000186         2518-000206         2518-000207         2518-000208         2518-000209"
    }
  ]
}