  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Williston",
      "state": "VT",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83310",
      "recalling_firm": "LABORIE MEDICAL TECHNOLOGIES, CORP",
      "address_1": "400 Avenue D Ste 10",
      "address_2": "N/A",
      "postal_code": "05495-7828",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US distribution in AZ, and PA.",
      "recall_number": "Z-2076-2019",
      "product_description": "U3-3 Solar Blue, Laptop integrated power supply used with the Solar Blue Urodynamics for assessing lower urinary tract dysfunctions",
      "product_quantity": "2 US",
      "reason_for_recall": "Power supply unit (PS-SBL) not correctly labeled, and insulation material used at the low voltage DC end does not meet all applicable flammability rating requirements. may lead to a potential fire hazard",
      "recall_initiation_date": "20190717",
      "center_classification_date": "20190725",
      "termination_date": "20230710",
      "report_date": "20190731",
      "code_info": "Serial Numbers: PE18-7SBL1452, PE18-7SBL 1414"
    }
  ]
}