      "status": "Ongoing",
      "city": "Carlsbad",
      "state": "CA",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88127",
      "recalling_firm": "Respironics California, LLC",
      "address_1": "2271 Cosmos Ct",
      "address_2": "N/A",
      "postal_code": "92011-1517",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: U.S (nationwide) to states of.: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SC, TN, TX, UT, VA, VT, WA, WI, WV, and WY; and O.U.S. countries of: Andorra, United Arab Emirates, Albania, Argentina, Austria, Australia, Belgium, Canada, Switzerland, Chile, China, Colombia, Czech Republic, Germany, Denmark, Spain, Finland, France, United Kingdom, French Guiana, Guadeloupe, Greece, Hong Kong, Indonesia, Ireland, India, Iceland, Italy, Jordan, Kenya, Kuwait, Lebanon, Latvia, Libya, Morocco, Martinique, Malta, Mexico, Malaysia, New Caledonia, Nigeria, Netherlands, Norway, New Zealand, Philippines, Poland, Portugal, Qatar, R¿union, Russian Federation, Saudi Arabia, Sweden, Singapore, Slovenia, Thailand, and South Africa.",
      "recall_number": "Z-2074-2021",
      "product_description": "Philips Respironics V60 Plus Ventilator with Software Versions 3.00 and 3.10.    The Respironics V60/V60 Plus Ventilator is an assist ventilator and is intended to augment patient breathing. It is intended for spontaneously breathing individuals who require mechanical ventilation: patients with respiratory failure, chronic respiratory insufficiency, or obstructive sleep apnea in a hospital or other institutional settings under the direction of a physician.",
      "product_quantity": "Total of V60/V60 Plus = 23,115 systems",
      "reason_for_recall": "It has been identified that ventilators equipped with High Flow Therapy (Software Versions 3.00 and 3.10), that includes a designed safety mechanism that limits the amount of pressure that can be delivered to a patient, in situations where the system pressure reaches the maximum limit and sounds the \"Cannot Reach Target Flow\"  alarm , the system will reduce pressure and simultaneously decreasing the flow rate to a level below what was set by the clinicians.  If the reason for the increase in pressure is not resolved, the ventilator will continue to provide the lower flow rate, which result in patients experiencing oxygen desaturation which can be characterized as moderate or severe hypoxemia.",
      "recall_initiation_date": "20210618",
      "center_classification_date": "20210721",
      "report_date": "20210728",
      "code_info": "Model V60 Plus Ventilator: P/N 1138747 and 1137276;    US System Serial Number/UDI Code:  100363973/ ;  100366525/ ;  100366542/ ;  100371876/ (01)00884838090996(21)100371876;  100373102/ (01)00884838090996(21)100373102;  100374459/ (01)00884838090996(21)100374459;  100376350/ (01)00884838090996(21)100376350;  100377311/ (01)00884838090996(21)100377311;  100377324/ (01)00884838090996(21)100377324;  100380441/ (01)00884838090996(21)100380441;  100382407/ (01)00884838090996(21)100382407;  100382452/ (01)00884838090996(21)100382452;  100382461/ (01)00884838090996(21)100382461;  100384413/ (01)00884838090996(21)100384413;  100384748/ (01)00884838090996(21)100384748;  100385425/ (01)00884838090996(21)100385425;  100385428/ (01)00884838090996(21)100385428;  100385429/ (01)00884838090996(21)100385429;  100385447/ (01)00884838090996(21)100385447;  100385529/ (01)00884838090996(21)100385529;  100386081/ (01)00884838090996(21)100386081;  100386137/ (01)00884838090996(21)100386137;  100386179/ (01)00884838090996(21)100386179;  100389986/ (01)00884838090996(21)100389986;  100390076/ (01)00884838090996(21)100390076;  100392652/ (01)00884838090996(21)100392652;  100392801/ (01)00884838090996(21)100392801;  100392809/ (01)00884838090996(21)100392809;  100392899/ (01)00884838090996(21)100392899;  100393482/ (01)00884838090996(21)100393482;  100393544/ (01)00884838090996(21)100393544;  100393555/ (01)00884838090996(21)100393555;  100393571/ (01)00884838090996(21)100393571;  100393741/ (01)00884838090996(21)100393741;  100393749/ (01)00884838090996(21)100393749;  100395503/ (01)00884838090996(21)100395503;  100395807/ (01)00884838090996(21)100395807;  100396158/ (01)00884838090996(21)100396158;  100396540/ (01)00884838090996(21)100396540;  100396598/ (01)00884838090996(21)100396598;  100396747/ (01)00884838090996(21)100396747;  100398796/ (01)00884838090996(21)100398796;  100398852/ (01)00884838090996(21)100398852;  100403796/ (01)00884838090996(21)100403796;  100403798/ (01)00884838090996(21)100403798;  100403813/ (01)00884838090996(21)100403813;  100403828/ (01)00884838090996(21)100403828;  100403830/ (01)00884838090996(21)100403830;  100403832/ (01)00884838090996(21)100403832;  100403842/ (01)00884838090996(21)100403842;  100404111/ (01)00884838090996(21)100404111;  100404113/ (01)00884838090996(21)100404113;  100404114/ (01)00884838090996(21)100404114;  100404115/ (01)00884838090996(21)100404115;  100404117/ (01)00884838090996(21)100404117;  100404118/ (01)00884838090996(21)100404118;  100404124/ (01)00884838090996(21)100404124;  100404131/ (01)00884838090996(21)100404131;  100404192/ (01)00884838090996(21)100404192;  100405715/ (01)00884838090996(21)100405715;  100405718/ (01)00884838090996(21)100405718;  100405741/ (01)00884838090996(21)100405741;  100405742/ (01)00884838090996(21)100405742;  100406077/ (01)00884838090996(21)100406077;  100406208/ (01)00884838090996(21)100406208;  100408124/ (01)00884838090996(21)100408124;  100408126/ (01)00884838090996(21)100408126;  100408129/ (01)00884838090996(21)100408129;  100408133/ (01)00884838090996(21)100408133;  100408137/ (01)00884838090996(21)100408137;  100408144/ (01)00884838090996(21)100408144;  100408149/ (01)00884838090996(21)100408149;  100408419/ (01)00884838090996(21)100408419;  100408882/ (01)00884838090996(21)100408882;  100408886/ (01)00884838090996(21)100408886;  100408899/ (01)00884838090996(21)100408899;  100408901/ (01)00884838090996(21)100408901;  100409389/ (01)00884838090996(21)100409389;  100409394/ (01)00884838090996(21)100409394;  100409537/ (01)00884838090996(21)100409537;  100409560/ (01)00884838090996(21)100409560;  100409712/ (01)00884838090996(21)100409712;  100409734/ (01)00884838090996(21)100409734;  100409932/ (01)00884838090996(21)100409932;  100409937/ (01)00884838090996(21)100409937;  100410001/ (01)00884838090996(21)100410001;  100410024/ (01)00884838090996(21)100410024;  100412283/ (01)00884838090996(21)100412283;  100412293/ (01)00884838090996(21)100412293;  100412450/ (01)00884838090996(21)100412450;  100412596/ (01)00884838090996(21)100412596;  100412940/ (01)00884838090996(21)100412940;  100412943/ (01)00884838090996(21)100412943;  100412976/ (01)00884838090996(21)100412976;  100412980/ (01)00884838090996(21)100412980;  100412981/ (01)00884838090996(21)100412981;  100413021/ (01)00884838090996(21)100413021;  100413027/ (01)00884838090996(21)100413027;  100413028/ (01)00884838090996(21)100413028;  100413030/ (01)00884838090996(21)100413030;  100413033/ (01)00884838090996(21)100413033;  100413524/ (01)00884838090996(21)100413524;  100413571/ (01)00884838090996(21)100413571;  100413663/ (01)00884838090996(21)100413663;  100413670/ (01)00884838090996(21)100413670;  100413672/ (01)00884838090996(21)100413672;  100413837/ (01)00884838090996(21)100413837;  100413905/ (01)00884838090996(21)100413905;  100413990/ (01)00884838090996(21)100413990;  100414030/ (01)00884838090996(21)100414030;  100414034/ (01)00884838090996(21)100414034;  100414256/ (01)00884838090996(21)100414256;  100414346/ (01)00884838090996(21)100414346;  100414349/ (01)00884838090996(21)100414349;  100414484/ (01)00884838090996(21)100414484;  100414490/ (01)00884838090996(21)100414490;  100414496/ (01)00884838090996(21)100414496;  100414498/ (01)00884838090996(21)100414498;  100414503/ (01)00884838090996(21)100414503;  100414510/ (01)00884838090996(21)100414510;  100414527/ (01)00884838090996(21)100414527;  100414530/ (01)00884838090996(21)100414530;  100415081/ (01)00884838090996(21)100415081;  100415124/ (01)00884838090996(21)100415124;  100416665/ (01)00884838090996(21)100416665;  100416680/ (01)00884838090996(21)100416680;  100416717/ (01)00884838090996(21)100416717;  100417376/ (01)00884838090996(21)100417376;  100417558/ (01)00884838090996(21)100417558;  100417661/ (01)00884838090996(21)100417661;  100417668/ (01)00884838090996(21)100417668;  100417684/ (01)00884838090996(21)100417684;  100417892/ (01)00884838090996(21)100417892;  100418072/ (01)00884838090996(21)100418072;  100418075/ (01)00884838090996(21)100418075;  100418093/ (01)00884838090996(21)100418093;  100418094/ (01)00884838090996(21)100418094;  100418494/ (01)00884838090996(21)100418494;  100418618/ (01)00884838090996(21)100418618;  100419145/ (01)00884838090996(21)100419145;  100419163/ (01)00884838090996(21)100419163;  100419962/ (01)00884838090996(21)100419962;  100419968/ (01)00884838090996(21)100419968;  100419976/ (01)00884838090996(21)100419976;  100420066/ (01)00884838090996(21)100420066;  100420219/ (01)00884838090996(21)100420219;  100420305/ (01)00884838090996(21)100420305;  100420315/ (01)00884838090996(21)100420315;  100420989/ (01)00884838090996(21)100420989;  100420990/ (01)00884838090996(21)100420990;  100421014/ (01)00884838090996(21)100421014;  100421320/ (01)00884838090996(21)100421320;  100421554/ (01)00884838090996(21)100421554;  100421861/ (01)00884838090996(21)100421861;  100422301/ (01)00884838090996(21)100422301;  100422313/ (01)00884838090996(21)100422313;  100423017/ (01)00884838090996(21)100423017;  100423029/ (01)00884838090996(21)100423029;  100423344/ ;  100423346/ (01)00884838090996(21)100423346;  100423348/ ;  100423352/ (01)00884838090996(21)100423352;  100423551/ (01)00884838090996(21)100423551;  100423558/ (01)00884838090996(21)100423558;  100423682/ (01)00884838090996(21)100423682;  100423683/ ;  100423685/ (01)00884838090996(21)100423685;  100423814/ (01)00884838090996(21)100423814;  100423846/ (01)00884838090996(21)100423846;  100423851/ (01)00884838090996(21)100423851;  100423877/ (01)00884838090996(21)100423877;  100423906/ (01)00884838090996(21)100423906;  100423924/ (01)00884838090996(21)100423924;  100423930/ (01)00884838090996(21)100423930;  100423934/ (01)00884838090996(21)100423934;  100424157/ (01)00884838090996(21)100424157;  100424778/ (01)00884838090996(21)100424778;  100425150/ (01)00884838090996(21)100425150;  100425183/ (01)00884838090996(21)100425183;  100425226/ (01)00884838090996(21)100425226;  100425229/ ;  100425230/ (01)00884838090996(21)100425230;  100425235/ (01)00884838090996(21)100425235;  100425696/ ;  100425742/ (01)00884838090996(21)100425742;  100427796/ (01)00884838090996(21)100427796;  100427967/ ;  100428999/ (01)00884838090996(21)100428999;  100429553/ (01)00884838090996(21)100429553;  100429568/ (01)00884838090996(21)100429568;  100431103/ ;  100431133/ (01)00884838090996(21)100431133;  100431231/ (01)00884838090996(21)100431231;  100431316/ (01)00884838090996(21)100431316;  100431449/ ;  100431724/ (01)00884838090996(21)100431724;  100431755/ ;  100431756/ ;  100432709/ ;  100432710/ ;  100433807/ ;  100433814/ ;  100433815/ (01)00884838090996(21)100433815;  100433817/ (01)00884838090996(21)100433817;  100433827/ (01)00884838090996(21)100433827;  100433837/ (01)00884838090996(21)100433837;  100433901/ ;  100433903/ ;  100433909/ (01)00884838090996(21)100433909;  100433910/ ;  100433911/ (01)00884838090996(21)100433911;  100434067/ (01)00884838090996(21)100434067;  100434069/ ;  100434077/ ;  100434078/ (01)00884838090996(21)100434078;  100434180/ (01)00884838090996(21)100434180;  100434185/ ;  100434197/ (01)00884838090996(21)100434197;  100435673/ (01)00884838089280(21)100435673;  100436278/ (01)00884838090996(21)100436278;  100436322/ (01)00884838090996(21)100436322;  100436598/ (01)00884838090996(21)100436598;  100436641/ (01)00884838090996(21)100436641;  100436649/ (01)00884838090996(21)100436649;  100436654/ (01)00884838090996(21)100436654;  100436661/ ;  100436680/ (01)00884838090996(21)100436680;  100436685/ (01)00884838090996(21)100436685;  100436691/ (01)00884838090996(21)100436691;  100436696/ (01)00884838090996(21)100436696;  100436700/ (01)00884838090996(21)100436700;  100436799/ (01)00884838090996(21)100436799;  100436809/ (01)00884838090996(21)100436809;  100437253/ (01)00884838090996(21)100437253;  100437269/ (01)00884838090996(21)100437269;  100447800/ (01)00884838090996(21)100447800;  100447802/ (01)00884838090996(21)100447802;  100447803/ (01)00884838090996(21)100447803;  100447807/ (01)00884838090996(21)100447807;  100447817/ (01)00884838090996(21)100447817;  100448385/ (01)00884838090996(21)100448385;  100448386/ (01)00884838090996(21)100448386;  100448501/ (01)00884838090996(21)100448501;  100448509/ (01)00884838090996(21)100448509;  100448516/ (01)00884838090996(21)100448516;  100448519/ (01)00884838090996(21)100448519;  100448563/ (01)00884838090996(21)100448563;  100448570/ (01)00884838090996(21)100448570;  100448572/ (01)00884838090996(21)100448572;  100448587/ (01)00884838090996(21)100448587;  100448588/ (01)00884838090996(21)100448588;  100448729/ (01)00884838090996(21)100448729;  100448730/ (01)00884838090996(21)100448730;  100448733/ (01)00884838090996(21)100448733;  100455183/ (01)00884838090996(21)100455183;  100458043/ (01)00884838090996(21)100458043;  100458826/ (01)00884838090996(21)100458826;  MX0000539/ (01)00884838090996(21)MX0000539;  MX0000558/ (01)00884838090996(21)MX0000558;  MX0001763/ (01)00884838090996(21)MX0001763;  MX0002491/ (01)00884838090996(21)MX0002491;  MX0002704/ (01)00884838090996(21)MX0002704;  MX0002716/ (01)00884838090996(21)MX0002716;  MX0002723/ (01)00884838090996(21)MX0002723;  MX0002728/ (01)00884838090996(21)MX0002728;  MX0002731/ (01)00884838090996(21)MX0002731;  MX0002737/ (01)00884838090996(21)MX0002737;  MX0002751/ (01)00884838090996(21)MX0002751;  MX0002761/ (01)00884838090996(21)MX0002761;  MX0002764/ (01)00884838090996(21)MX0002764;  MX0002768/ (01)00884838090996(21)MX0002768;  MX0002819/ (01)00884838090996(21)MX0002819;  MX0003003/ (01)00884838090996(21)MX0003003;  MX0003012/ (01)00884838090996(21)MX0003012;  MX0003460/ (01)00884838090996(21)MX0003460;  MX0003474/ (01)00884838090996(21)MX0003474;  MX0003496/ (01)00884838090996(21)MX0003496;  MX0003503/ (01)00884838090996(21)MX0003503;  MX0003505/ (01)00884838090996(21)MX0003505;  MX0003546/ (01)00884838090996(21)MX0003546;  MX0003555/ (01)00884838090996(21)MX0003555;  MX0004467/ (01)00884838090996(21)MX0004467;  MX0004469/ (01)00884838090996(21)MX0004469;  MX0004470/ (01)00884838090996(21)MX0004470;  MX0007172/ (01)00884838090996(21)MX0007172;  MX0007173/ (01)00884838090996(21)MX0007173;  MX0007179/ (01)00884838090996(21)MX0007179;  MX0007198/ ;  MX0007230/ (01)00884838090996(21)MX0007230;  MX0007242/ (01)00884838090996(21)MX0007242;  MX0007243/ (01)00884838090996(21)MX0007243;  MX0007253/ (01)00884838090996(21)MX0007253;  MX0007538/ (01)00884838090996(21)MX0007538;  MX0007541/ (01)00884838090996(21)MX0007541;  MX0007542/ (01)00884838090996(21)MX0007542;  MX0007577/ (01)00884838090996(21)MX0007577;  MX0007585/ ;  MX0007587/ (01)00884838090996(21)MX0007587;  MX0007592/ (01)00884838090996(21)MX0007592;  MX0007606/ (01)00884838090996(21)MX0007606;  MX0007837/ (01)00884838090996(21)MX0007837;  MX0007860/ (01)00884838090996(21)MX0007860;  MX0008995/ (01)00884838090996(21)MX0008995;  MX0009104/ (01)00884838090996(21)MX0009104;  MX0009145/ (01)00884838090996(21)MX0009145;  MX0009758/ (01)00884838090996(21)MX0009758;  MX0010728/ (01)00884838090996(21)MX0010728;  MX0010733/ (01)00884838090996(21)MX0010733;  MX0010735/ (01)00884838090996(21)MX0010735;  MX0011083/ (01)00884838090996(21)MX0011083;  MX0013445/ (01)00884838090996(21)MX0013445;  MX0013515/ (01)00884838090996(21)MX0013515;  MX0015091/ ;  MX0015092/ (01)00884838090996(21)MX0015092;  MX0015099/ (01)00884838090996(21)MX0015099;  MX0015110/ (01)00884838090996(21)MX0015110;  MX0015123/ (01)00884838090996(21)MX0015123;  MX0015124/ (01)00884838090996(21)MX0015124;  MX0015130/ ;  MX0015144/ (01)00884838090996(21)MX0015144;  MX0015146/ (01)00884838090996(21)MX0015146;  MX0015158/ (01)00884838090996(21)MX0015158;  MX0015168/ (01)00884838090996(21)MX0015168;  MX0015172/ (01)00884838090996(21)MX0015172;  MX0015469/ (01)00884838090996(21)MX0015469;  100436648/ (01)00884838090996(21)100436648;  100436324/ (01)00884838090996(21)100436324;        OUS(International) System Serial Number/UDI Code:  100280989/ ;  100282854/ ;  100289689/ ;  100289695/ ;  100289696/ ;  100289709/ ;  100290610/ ;  100290613/ ;  100290614/ ;  100290616/ ;  100290648/ ;  100290654/ ;  100290664/ ;  100291847/ ;  100292342/ ;  100296023/ ;  100297858/ ;  100297860/ ;  100298767/ ;  100298928/ ;  100302008/ ;  100302015/ ;  100302023/ ;  100302027/ ;  100302260/ ;  100304762/ ;  100304779/ ;  100305136/ ;  100305439/ ;  100305742/ ;  100305743/ ;  100305745/ ;  100305746/ ;  100305747/ ;  100305831/ ;  100306474/ ;  100309481/ ;  100309494/ ;  100309574/ ;  100309640/ ;  100309643/ ;  100309645/ ;  100309735/ ;  100311163/ (01)00884838089280(21)100311163;  100311175/ ;  100314682/ ;  100314684/ ;  100314877/ ;  100315149/ ;  100315152/ ;  100315157/ ;  100315614/ ;  100315897/ ;  100318607/ ;  100318610/ ;  100318612/ ;  100318614/ ;  100319291/ ;  100319302/ ;  100319775/ ;  100319777/ ;  100320373/ ;  100320645/ ;  100321730/ ;  100321910/ ;  100321913/ ;  100322636/ ;  100322640/ ;  100322876/ ;  100322898/ ;  100322903/ ;  100325105/ ;  100325106/ ;  100325107/ ;  100325111/ ;  100325113/ ;  100325116/ ;  100327540/ ;  100327544/ ;  100335583/ ;  100336136/ ;  100336757/ ;  100336758/ ;  100336902/ ;  100337081/ ;  100337087/ ;  100337093/ ;  100337263/ ;  100338949/ ;  100338950/ ;  100338953/ ;  100339023/ ;  100339027/ ;  100339232/ (01)00884838089280(21)100339232;  100339234/ (01)00884838089280(21)100339234;  100339236/ ;  100339246/ ;  100339315/ ;  100339317/ ;  100339324/ ;  100339497/ (01)00884838089280(21)100339497;  100340912/ ;  100341120/ ;  100341166/ ;  100341191/ ;  100341366/ ;  100341397/ ;  100341866/ ;  100341882/ ;  100341890/ ;  100341908/ ;  100341909/ ;  100341910/ ;  100341999/ ;  100342525/ ;  100342534/ ;  100342536/ ;  100342539/ ;  100342545/ ;  100342547/ ;  100342548/ ;  100342559/ ;  100342561/ ;  100342565/ ;  100342566/ ;  100342571/ ;  100342579/ ;  100342890/ ;  100342896/ ;  100342897/ ;  100342935/ ;  100342942/ ;  100342943/ ;  100342947/ ;  100342950/ ;  100342968/ ;  100342977/ ;  100342978/ ;  100342979/ ;  100342983/ ;  100343100/ ;  100343103/ ;  100343104/ ;  100343105/ ;  100343112/ ;  100343113/ ;  100343124/ ;  100343128/ ;  100343131/ ;  100343138/ ;  100343142/ ;  100343153/ ;  100343160/ ;  100343569/ ;  100343622/ ;  100343628/ (01)00884838089280(21)100343628;  100343681/ ;  100344206/ ;  100344207/ ;  100344210/ ;  100344653/ ;  100344655/ ;  100344663/ ;  100344665/ ;  100344667/ ;  100344681/ ;  100344685/ ;  100344687/ ;  100344697/ ;  100344859/ ;  100345114/ ;  100345116/ ;  100345127/ ;  100345130/ ;  100345132/ ;  100345141/ ;  100345326/ ;  100345518/ ;  100345564/ ;  100345588/ ;  100345617/ ;  100346214/ ;  100346215/ ;  100346216/ ;  100346219/ ;  100346242/ ;  100346246/ ;  100346253/ ;  100346256/ ;  100346288/ ;  100346290/ ;  100346292/ ;  100346293/ ;  100346299/ ;  100346322/ ;  100346329/ ;  100346849/ ;  100346854/ ;  100346856/ ;  100346858/ ;  100346878/ ;  100346912/ ;  100348113/ ;  100348130/ ;  100348138/ ;  100348149/ ;  100348173/ ;  100348175/ ;  100348442/ ;  100348447/ ;  100350134/ ;  100350669/ ;  100350677/ ;  100350712/ ;  100350715/ ;  100350903/ ;  100351553/ ;  100351557/ ;  100351559/ ;  100351560/ ;  100351561/ ;  100351569/ ;  100351574/ ;  100351581/ ;  100351587/ ;  100351708/ ;  100351709/ ;  100351716/ ;  100351723/ ;  100352069/ ;  100352090/ ;  100352099/ ;  100352100/ ;  100352106/ ;  100352117/ ;  100352136/ ;  100352745/ ;  100352752/ ;  100352781/ ;  100352801/ ;  100352810/ ;  100352815/ ;  100353244/ ;  100353254/ ;  100353271/ ;  100353280/ ;  100353288/ (01)00884838089280(21)100353288;  100353589/ ;  100354439/ ;  100354471/ ;  100354475/ ;  100354705/ (01)00884838089280(21)100354705;  100355164/ (01)00884838089280(21)100355164;  100355673/ ;  100355856/ ;  100355896/ ;  100357138/ ;  100357735/ ;  100357777/ ;  100358173/ ;  100358181/ ;  100358197/ ;  100358349/ ;  100359518/ ;  100359519/ ;  100359609/ ;  100359617/ (01)00884838089280(21)100359617;  100359645/ ;  100359647/ ;  100359870/ ;  100359873/ ;  100360147/ ;  100360213/ ;  100360371/ ;  100361205/ ;  100361424/ ;  100361429/ ;  100361504/ ;  100361578/ ;  100361585/ ;  100361766/ ;  100361768/ ;  100361963/ ;  100361973/ ;  100362515/ ;  100363001/ ;  100363012/ ;  100363014/ ;  100363021/ ;  100363026/ ;  100363042/ ;  100363719/ ;  100363771/ ;  100363781/ ;  100364568/ ;  100364600/ ;  100365160/ ;  100365172/ ;  100367184/ ;  100367279/ ;  100368019/ ;  100368075/ ;  100368622/ ;  100369687/ ;  100369690/ ;  100369698/ ;  100369719/ ;  100369908/ ;  100370031/ ;  100371204/ ;  100375137/ ;  100375327/ ;  100375328/ ;  100375337/ ;  100375412/ ;  100375444/ (01)00884838089280(21)100375444;  100375448/ ;  100375456/ ;  100375460/ ;  100375478/ ;  100375479/ ;  100376361/ (01)00884838089280(21)100376361;  100376366/ (01)00884838089280(21)100376366;  100378646/ (01)00884838089280(21)100378646;  100378659/ ;  100380816/ (01)00884838089280(21)100380816;  100381005/ (01)00884838089280(21)100381005;  100381265/ ;  100381339/ (01)00884838089280(21)100381339;  100381857/ (01)00884838089280(21)100381857;  100383292/ (01)00884838089280(21)100383292;  100384081/ (01)00884838089280(21)100384081;  100384771/ (01)00884838089280(21)100384771;  100384773/ (01)00884838089280(21)100384773;  100384806/ (01)00884838089280(21)100384806;  100385304/ (01)00884838089280(21)100385304;  100385342/ (01)00884838089280(21)100385342;  100385359/ (01)00884838089280(21)100385359;  100387277/ (01)00884838089280(21)100387277;  100387296/ (01)00884838089280(21)100387296;  100387308/ (01)00884838089280(21)100387308;  100387830/ (01)00884838089280(21)100387830;  100387874/ (01)00884838089280(21)100387874;  100387885/ (01)00884838089280(21)100387885;  100389427/ (01)00884838089280(21)100389427;  100390083/ (01)00884838089280(21)100390083;  100390089/ (01)00884838089280(21)100390089;  100391984/ (01)00884838089280(21)100391984;  100392618/ (01)00884838089280(21)100392618;  100392632/ (01)00884838089280(21)100392632;  100395357/ (01)00884838089280(21)100395357;  100399025/ (01)00884838089280(21)100399025;  100399325/ ;  100399414/ ;  100400107/ (01)00884838089280(21)100400107;  100400420/ (01)00884838089280(21)100400420;  100400924/ ;  100400929/ ;  100401134/ (01)00884838089280(21)100401134;  100401137/ ;  100402491/ ;  100402502/ (01)00884838089280(21)100402502;  100402955/ ;  100403074/ (01)00884838089280(21)100403074;  100404141/ (01)00884838089280(21)100404141;  100411608/ ;  100411615/ (01)00884838089280(21)100411615;  100426463/ (01)00884838089280(21)100426463;  100426923/ (01)00884838089280(21)100426923;  100434598/ ;  100438752/ ;  100440475/ (01)00884838089280(21)100440475;  100440835/ ;  100446264/ (01)00884838089280(21)100446264;  100456002/ (01)00884838089280(21)100456002;  201035973/ ;  201036161/ ;  201036162/ ;  201036163/ ;  201036171/ ;  201036175/ ;  201036177/ ;  201036179/ ;  201036180/ ;  201036256/ ;  201036259/ ;  201036272/ ;  201036274/ ;  201038686/ ;  201040131/ ;  201040182/ ;  201040188/ ;  201040197/ ;  201040198/ ;  201040199/ ;  201040210/ ;  201040337/ ;  201040368/ ;  201040464/ ;  201040465/ ;  201040480/ ;  201040556/ ;  201040583/ ;  201040589/ ;  201040594/ ;  201040604/ ;  201040624/ ;  201040628/ ;  201040647/ ;  HU2000081/ (01)00884838089280(21)HU2000081;  HU2000084/ (01)00884838089280(21)HU2000084;  HU2000086/ (01)00884838089280(21)HU2000086;  HU2000091/ (01)00884838089280(21)HU2000091;  HU2000092/ (01)00884838089280(21)HU2000092;  HU2000106/ (01)00884838089280(21)HU2000106;  HU2000108/ (01)00884838089280(21)HU2000108;  HU2000127/ (01)00884838089280(21)HU2000127;  HU2000131/ (01)00884838089280(21)HU2000131;  HU2000136/ (01)00884838089280(21)HU2000136;  HU2000155/ (01)00884838089280(21)HU2000155;  HU2000172/ (01)00884838089280(21)HU2000172;  HU2000201/ (01)00884838089280(21)HU2000201;  HU2000217/ ;  HU2000279/ (01)00884838089280(21)HU2000279;  HU2000309/ ;  HU2000324/ (01)00884838089280(21)HU2000324;  HU2000393/ (01)00884838089280(21)HU2000393;  HU2000427/ (01)00884838089280(21)HU2000427;  HU2000449/ (01)00884838089280(21)HU2000449;  HU2000459/ ;  HU2000527/ ;  HU2000574/ ;  HU2000602/ ;  HU2000667/ ;  HU2000706/ ;  HU2000865/ ;",
      "more_code_info": ""
    }
  ]
}