  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Yorba Linda",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "71552",
      "recalling_firm": "Nobel Biocare Usa Llc",
      "address_1": "22715/22725 Savi Ranch Pkwy",
      "address_2": "N/A",
      "postal_code": "92887",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-US (nationwide) and the countries of Australia, Austria, Belgium, Canada, Finland, France, Germany, Greece, Hong Kong, Hungary, Italy, Luxembourg, Netherlands, New Zealand, Norway, Portugal, South Africa, Spain, Switzerland, and United Kingdom.",
      "recall_number": "Z-2074-2015",
      "product_description": "Abutment Retrieval Instrument Zirconia CC RP/WP, Article No. 37882 and Abutment Retrieval Kit, Article No. 37508 (Article No. 37882 was assembled in this kit).  Used to remove a zirconia abutment if the abutment is stuck in the implant connection due to a tight seat.  Endosseous dental implant instrument.",
      "product_quantity": "149 units total",
      "reason_for_recall": "One dimension of the affected instrument is incorrect. Therefore this instrument cannot be used to remove the metal adapter as intended.",
      "recall_initiation_date": "20150622",
      "center_classification_date": "20150715",
      "termination_date": "20150918",
      "report_date": "20150722",
      "code_info": "Article No. 37882, Batch No. 96149;   Article No. 37508, Batch No. 814929, 815096, 815140, and 855021.",
      "more_code_info": ""
    }
  ]
}