{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Carlsbad",
      "state": "CA",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88127",
      "recalling_firm": "Respironics California, LLC",
      "address_1": "2271 Cosmos Ct",
      "address_2": "N/A",
      "postal_code": "92011-1517",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: U.S (nationwide) to states of.: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SC, TN, TX, UT, VA, VT, WA, WI, WV, and WY; and O.U.S. countries of: Andorra, United Arab Emirates, Albania, Argentina, Austria, Australia, Belgium, Canada, Switzerland, Chile, China, Colombia, Czech Republic, Germany, Denmark, Spain, Finland, France, United Kingdom, French Guiana, Guadeloupe, Greece, Hong Kong, Indonesia, Ireland, India, Iceland, Italy, Jordan, Kenya, Kuwait, Lebanon, Latvia, Libya, Morocco, Martinique, Malta, Mexico, Malaysia, New Caledonia, Nigeria, Netherlands, Norway, New Zealand, Philippines, Poland, Portugal, Qatar, R¿union, Russian Federation, Saudi Arabia, Sweden, Singapore, Slovenia, Thailand, and South Africa.",
      "recall_number": "Z-2073-2021",
      "product_description": "Philips Respironics V60 Ventilator with Software Versions 3.00 and 3.10.    The Respironics V60/V60 Plus Ventilator is an assist ventilator and is intended to augment patient breathing. It is intended for spontaneously breathing individuals who require mechanical ventilation: patients with respiratory failure, chronic respiratory insufficiency, or obstructive sleep apnea in a hospital or other institutional settings under the direction of a physician.",
      "product_quantity": "Total of V60/V60 Plus = 23,115 systems",
      "reason_for_recall": "It has been identified that ventilators equipped with High Flow Therapy (Software Versions 3.00 and 3.10), that includes a designed safety mechanism that limits the amount of pressure that can be delivered to a patient, in situations where the system pressure reaches the maximum limit and sounds the \"Cannot Reach Target Flow\"  alarm , the system will reduce pressure and simultaneously decreasing the flow rate to a level below what was set by the clinicians.  If the reason for the increase in pressure is not resolved, the ventilator will continue to provide the lower flow rate, which result in patients experiencing oxygen desaturation which can be characterized as moderate or severe hypoxemia.",
      "recall_initiation_date": "20210618",
      "center_classification_date": "20210721",
      "report_date": "20210728",
      "code_info": "Model V60 Ventilator: P/N 1053613, 1053614, 1053615, 1053616, 1053617, DU1053617, U1053617, 1076716, 1137292    US System Serial Number/UDI Code:  100351272 (01)00884838020054(21)100351272  100006691   100193211   100220566 (01)00884838020054(21)100220566  100313757 (01)00884838020054(21)100313757  100313412 (01)00884838020054(21)100313412  100080135   100081569   100110187   100121703   100013472   100230650 (01)00884838020054(21)100230650  100374411 (01)00884838020054(21)100374411  100054305   100068863   100188390   100181283   100030036   100030039   100030046   100030040   100043349   100030041   100037246   201029401 (01)00884838020054(21)201029401  100278366 (01)00884838020054(21)100278366  100088340   100131320   201030489 (01)00884838020054(21)201030489  100057026   100074563   201005638   201001514   100027860   201005635   100154662   100056752   100130449   100154846   100078004   100055123   100191452   100077709   100015080   100191788   100129710   100028046   100077079   201017909   100224361 (01)00884838020054(21)100224361  100043830   100024302   100133355   100077493   100077503   100039424   100177846   201011206   100076877   100191271   100152171   100056385   100120289   100262177 (01)00884838020054(21)100262177  100120262   100047561   100156074   100104165   100115189   201026331 (01)00884838020054(21)201026331  100057549   100133002   100258925 (01)00884838020054(21)100258925  201030702 (01)00884838020054(21)201030702  100040748   100290605 (01)00884838020054(21)100290605  100044617   100147649   100038426   100049697   100318142 (01)00884838020054(21)100318142  100062636   100047178   100130954   100017106   100172675   100265434 (01)00884838020054(21)100265434  100097540   100174538   100028947   201033184 (01)00884838020054(21)201033184  100158623   100184938   100184879   100094731   100325804 (01)00884838020054(21)100325804  100117611   100048641   100267973 (01)00884838020054(21)100267973  100148068   100416709 (01)00884838020054(21)100416709  100118800   100177855   100023735   100025030   100040300   100279146 (01)00884838020054(21)100279146  100008228   100054315   100008224   100008438   100258476 (01)00884838020054(21)100258476  100008146   100006219   100063484   100074097   100349391 (01)00884838020054(21)100349391  100064083   100187263   100168073   100280881 (01)00884838020054(21)100280881  100256575 (01)00884838020054(21)100256575  100245312 (01)00884838020054(21)100245312  100038024   100088492   100017237   100039542   100245557 (01)00884838020054(21)100245557  100275362 (01)00884838020054(21)100275362  100053364   100190401   100237792 (01)00884838020054(21)100237792  100059977   100360168 (01)00884838020054(21)100360168  100268194 (01)00884838020054(21)100268194  100147646   100063411   201024993 (01)00884838020054(21)201024993  100112558   100113474   100266415 (01)00884838020054(21)100266415  100031093   100037777   100051583   100105808   100411005 (01)00884838020054(21)100411005  100148200   100376334 (01)00884838020054(21)100376334  100057733   100091066   100122067   100073597   100057124   100146619   100302964 (01)00884838020054(21)100302964  100091075   100120976   100147661   100032534   MX0000332 (01)00884838020054(21)MX0000332  201028104 (01)00884838020054(21)201028104  100130950   100360674 (01)00884838020054(21)100360674  100094723   201032131 (01)00884838020054(21)201032131  100347825 (01)00884838020054(21)100347825  100043722   100114570   100266417 (01)00884838020054(21)100266417  100043812   100312855 (01)00884838020054(21)100312855  100066482   100066913   100057973   100240908 (01)00884838020054(21)100240908  100360608 (01)00884838020054(21)100360608  100132982   100172584   100105301   MX0000811 (01)00884838020054(21)MX0000811  100049922   201014454   100045057   100119310   100279355 (01)00884838020054(21)100279355  100256638 (01)00884838020054(21)100256638  100379488 (01)00884838020054(21)100379488  100230368 (01)00884838020054(21)100230368  100012543   100058126   100095885   100072279   100074017   100028855   100272829 (01)00884838020054(21)100272829  100027744   100015201   100058120   100134489   100065687   100192989   100266252 (01)00884838020054(21)100266252  201030363 (01)00884838020054(21)201030363  100080380   100060798   100010101   100073350   100050880   201019491   100034592   100059844   100292155 (01)00884838020054(21)100292155  100065690   100240306 (01)00884838020054(21)100240306  100092442   100084643   100063471   100091560   100184278   201008243   100051577   100292954 (01)00884838020054(21)100292954  100095009   100092540   100170409   100251069 (01)00884838020054(21)100251069  100184876   201022432 (01)00884838020054(21)201022432  100249314 (01)00884838020054(21)100249314  100188585   100141803   100028067   100265833 (01)00884838020054(21)100265833  100112787   100238524 (01)00884838020054(21)100238524  100023629   100191117   100023474   100060967   100066569   100148673   100092302   100042111   201029330 (01)00884838020054(21)201029330  100094822   100150330   100084010   100113272   100378839 (01)00884838020054(21)100378839  100104483   100043449   100082989   100132958   100090020   100031686   100029521   100084770   100042021   100080984   100091092   100096788   100175175   100235999 (01)00884838020054(21)100235999  201029470 (01)00884838020054(21)201029470  100015985   100162248   100235912 (01)00884838020054(21)100235912  100168717   100284772 (01)00884838020054(21)100284772  100082777   100071612   100385452 (01)00884838020054(21)100385452  100029517   100319179 (01)00884838020054(21)100319179  100035364   100141536   100033471   100344538 (01)00884838020054(21)100344538  201000384   100330907 (01)00884838020054(21)100330907  100209178 (01)00884838020054(21)100209178  100229296 (01)00884838020054(21)100229296  201019552   100185513   100186030   100047820   100009102   100015661   100251121 (01)00884838020054(21)100251121  100283321 (01)00884838020054(21)100283321  100005114   100066315   100292966 (01)00884838020054(21)100292966  100146517   100267865 (01)00884838020054(21)100267865  201026954 (01)00884838020054(21)201026954  100043440   100037836   100060802   100038505   100078562   100037657   100031685   100147672   201035397 (01)00884838020054(21)201035397  100258290 (01)00884838020054(21)100258290  100108292   100118814   100101557   100057971   100302866 (01)00884838020054(21)100302866  100041202   100058679   100057990   201030455 (01)00884838020054(21)201030455  100054197   100029524   100012045   100073413   100063498   100081182   100081568   201019506   100072280   100065053   100303070 (01)00884838020054(21)100303070  100239144 (01)00884838020054(21)100239144  100193608   100031689   100143204   100034694   100179762   100273519 (01)00884838020054(21)100273519  100077060   100048293   100071616   201033026 (01)00884838020054(21)201033026  100066726   100059847   100162238   100062768   100088753   100101568   100056764   100033470   100097539   100035601   100030461   100105810   100174144   100063885   100090005   100103817   100151611   201017722   100120829   100308332 (01)00884838020054(21)100308332  100258815 (01)00884838020054(21)100258815  100028566   100005396   100080859   100105828   100061374   100227575 (01)00884838020054(21)100227575  100366009 (01)00884838020054(21)100366009  100356232 (01)00884838020054(21)100356232  100357485 (01)00884838020054(21)100357485  100044776   100043442   100029669   100351203 (01)00884838020054(21)100351203  100356963 (01)00884838020054(21)100356963  100236545 (01)00884838020054(21)100236545  100246393 (01)00884838020054(21)100246393  201025663 (01)00884838020054(21)201025663  100099969   MX0000233 (01)00884838020054(21)MX0000233  MX0000368 (01)00884838020054(21)MX0000368  100079012   100278291 (01)00884838020054(21)100278291  100079000   100099766   100038336   100048253   100035738   100023434   100041906   100415118 (01)00884838020054(21)100415118  100064441   100312749 (01)00884838020054(21)100312749  100120430   100148191   100365899 (01)00884838020054(21)100365899  100242919 (01)00884838020054(21)100242919  100090018   100150604   100278299 (01)00884838020054(21)100278299  100045096   100236459 (01)00884838020054(21)100236459  201002834   100301490 (01)00884838020054(21)100301490  100187690   100150052   100041751   100009973   100344568 (01)00884838020054(21)100344568  100051744   100049919   100175176   100121323   100012861   100161316   100092045   100057962   100334852 (01)00884838020054(21)100334852  100058800   100126907   100246068 (01)00884838020054(21)100246068  100150665   100271325 (01)00884838020054(21)100271325  100093992   100130367   100288718 (01)00884838020054(21)100288718  100230446 (01)00884838020054(21)100230446  100021591   201028355 (01)00884838020054(21)201028355  100125239   100040868   100065834   100135922   100253808 (01)00884838020054(21)100253808  100087926   100051222   100150324   100074107   100042475   100121152   100286599 (01)00884838020054(21)100286599  100067204   100072285   100043350   100171351   100095332   100114921   100057861   100185673   100266685 (01)00884838020054(21)100266685  100293443 (01)00884838020054(21)100293443  MX0000238 (01)00884838020054(21)MX0000238  100331096 (01)00884838020054(21)100331096  100331180 (01)00884838020054(21)100331180  100240670 (01)00884838020054(21)100240670  100293958 (01)00884838020054(21)100293958  100088708   100275526 (01)00884838020054(21)100275526  100143160   100275236 (01)00884838020054(21)100275236  100194038 (01)00884838020054(21)100194038  100151427   100168250   100168622   100135783   100195058 (01)00884838020054(21)100195058  100188258   100064686   100059973   100005495   100043439   100295535 (01)00884838020054(21)100295535  100086410   100085637   100006536   100116545   100136237   100321031 (01)00884838020054(21)100321031  100179120   100164145   100121422   100146430   100062493   100299044 (01)00884838020054(21)100299044  100194457   100048295   100082338   100038502   100280731 (01)00884838020054(21)100280731  100101547   100029856   100072198   100062868   100331195 (01)00884838020054(21)100331195  100047308   100187253   100379299 (01)00884838020054(21)100379299  100037769   100062986   201009743   100047183   100261542 (01)00884838020054(21)100261542  100068667   100047288   100088497   100041912   100229287 (01)00884838020054(21)100229287  100134980   100134790   100172336   100057010   100186433   100097542   100064438   100021504   100045032   100317647 (01)00884838020054(21)100317647  100034160   100021836   100159749   100047287   100249025 (01)00884838020054(21)100249025  100187691   100267319 (01)00884838020054(21)100267319  100031559   100153032   100313539 (01)00884838020054(21)100313539  100006213   100171610   201022885 (01)00884838020054(21)201022885  100040173   201025712 (01)00884838020054(21)201025712  100109734   100232319 (01)00884838020054(21)100232319  100121691   100147641   100238964 (01)00884838020054(21)100238964  100170127   100099354   100255853 (01)00884838020054(21)100255853  100223794 (01)00884838020054(21)100223794  100063393   100089004   100226064 (01)00884838020054(21)100226064  100082340   100010317   100088401   100155080   201014575   201024359 (01)00884838020054(21)201024359  100149437   100409568 (01)00884838020054(21)100409568  201024303 (01)00884838020054(21)201024303  100086137   100122252   100223642 (01)00884838020054(21)100223642  100037640   201024282 (01)00884838020054(21)201024282  100086194   100028055   100073101   201015094   100118785   100061151   100025790   100069330   100334142 (01)00884838020054(21)100334142  100157691   100185654   100105448   100310876 (01)00884838020054(21)100310876  100301495 (01)00884838020054(21)100301495  100276891 (01)00884838020054(21)100276891  100309098 (01)00884838020054(21)100309098  201014424   100104852   100092307   100181279   100220406 (01)00884838020054(21)100220406  100168705   100006528   100182319   201025673 (01)00884838020054(21)201025673  100054314   100096083   100057539   100238137 (01)00884838020054(21)100238137  100192675   100120651   100172589   100180562   100023417   100023232   100299584 (01)00884838020054(21)100299584  100030827   100261747 (01)00884838020054(21)100261747  100272460 (01)00884838020054(21)100272460  201017686   100063013   100218795 (01)00884838020054(21)100218795  100012694   100047381   100047382   100062883   100088746   100324163 (01)00884838020054(21)100324163  201027220 (01)00884838020054(21)201027220  100034697   100034995   100086615   100049797   100224968 (01)00884838020054(21)100224968  100080993   201008497   100044788   100297480 (01)00884838020054(21)100297480  100070679   201000381   100014079   201027348 (01)00884838020054(21)201027348  100118787   100152929   100135503   100158627   100086650   100033326   100027739   100066854   100156075   100321931 (01)00884838020054(21)100321931  100377146 (01)00884838020054(21)100377146  100230719   100119582   201031729 (01)00884838020054(21)201031729  100011302   100141943   100006603   100023238   100330574 (01)00884838020054(21)100330574  100047307   100155060   100034979   100073835   100341094 (01)00884838020054(21)100341094  100157491   100016355   100065697   100262209 (01)00884838020054(21)100262209  100028482   100117761   100006941   100026966   100066178   100026828   100028476   100027422   100024475   100085371   100113868   201034495 (01)00884838020054(21)201034495  100409926 (01)00884838020054(21)100409926  MX0001199 (01)00884838020054(21)MX0001199  100113860   100101195   100313756 (01)00884838020054(21)100313756  100295529 (01)00884838020054(21)100295529  100093789   100246120 (01)00884838020054(21)100246120  100066581   100297483 (01)00884838020054(21)100297483  100066100   100135284   100121708   100086648   100073509   100113054   100032482   100275980 (01)00884838020054(21)100275980  100164291   100064784   100048647   100227589 (01)00884838020054(21)100227589  100027749   100114949   100087819   100334369 (01)00884838020054(21)100334369  100382512 (01)00884838020054(21)100382512  100320594 (01)00884838020054(21)100320594  100313743 (01)00884838020054(21)100313743  100120822   100378826 (01)00884838020054(21)100378826  100059956   100141152   100240716 (01)00884838020054(21)100240716  100246461 (01)00884838020054(21)100246461  100152770   100248825 (01)00884838020054(21)100248825  100131669   100235921 (01)00884838020054(21)100235921  100088608   100068592   100004546   100075675   100048498   MX0001029 (01)00884838020054(21)MX0001029  100029840   100094726   100073833   100054303   100063004   100236027 (01)00884838020054(21)100236027  100069013   100050757   100043431   100042570   100062357   100218811 (01)00884838020054(21)100218811  100265685 (01)00884838020054(21)100265685  100083739   100150664   100340614 (01)00884838020054(21)100340614  100004997   100112866   100004542   100047719   100044803   100390066 (01)00884838020054(21)100390066  201010452   100106893   100302368 (01)00884838020054(21)100302368  100032471   100359425 (01)00884838020054(21)100359425  100241825 (01)00884838020054(21)100241825  100048256   100095120   100283162 (01)00884838020054(21)100283162  100013791   100094890   100092434   100195050 (01)00884838020054(21)100195050  100302583 (01)00884838020054(21)100302583  100318179 (01)00884838020054(21)100318179  100080061   100172522   201002590   100044618   100062882   100049352   100232302 (01)00884838020054(21)100232302  201021064   100084013   100059724   100028470   100154215   100270684 (01)00884838020054(21)100270684  100176309   100131473   100304029 (01)00884838020054(21)100304029  100024299   100246582 (01)00884838020054(21)100246582  100066578   100065703   201016744   100143128   201025704 (01)00884838020054(21)201025704  201025598 (01)00884838020054(21)201025598  201010639   100351240 (01)00884838020054(21)100351240  100102513   100235909 (01)00884838020054(21)100235909  100014282   100041757   100033455   100075690   100246581 (01)00884838020054(21)100246581  100023721   100042671   100149602   100009517   100130526   100074092   100142048   100004031   100302580 (01)00884838020054(21)100302580  100064333   100041755   100042481   100007972   100178181   100148066   100045099   100058794   100255972 (01)00884838020054(21)100255972  201002796   100283355 (01)00884838020054(21)100283355  201000837   100008138   100165440   100121320   100112871   100094735   201002835   100091090   100082096   100333811 (01)00884838020054(21)100333811  100119209   100055120   100185262   100299287 (01)00884838020054(21)100299287  100356940 (01)00884838020054(21)100356940  100040181   201024768 (01)00884838020054(21)201024768  100143208   100118802   100009939   100050812   100298084 (01)00884838020054(21)100298084  100191144   100331194 (01)00884838020054(21)100331194  100349414 (01)00884838020054(21)100349414  100303494 (01)00884838020054(21)100303494  100265837 (01)00884838020054(21)100265837  100095092   100005599   100109421   100024832   100028344   100051618   100041075   100072195   100194448   100094936   100148677   100337476 (01)00884838020054(21)100337476  100150328   100071648   100266407 (01)00884838020054(21)100266407  100275754 (01)00884838020054(21)100275754  100148006   100158343   100016751   100258818 (01)00884838020054(21)100258818  100231061 (01)00884838020054(21)100231061  100186424   201001645   MX0001033 (01)00884838020054(21)MX0001033  100133093   100011328   100113271   100379344 (01)00884838020054(21)100379344  100093878   100016528   201031330 (01)00884838020054(21)201031330  100131170   100051574   201001686   100303800 (01)00884838020054(21)100303800  100149365   201034533 (01)00884838020054(21)201034533  100059952   100289913 (01)00884838020054(21)100289913  100261402 (01)00884838020054(21)100261402  100011271   100321034 (01)00884838020054(21)100321034  100093097   201031096 (01)00884838020054(21)201031096  100330563 (01)00884838020054(21)100330563  100321411 (01)00884838020054(21)100321411  100104724   201019556   100161306   100313201 (01)00884838020054(21)100313201  100319184 (01)00884838020054(21)100319184  201014577   100263575 (01)00884838020054(21)100263575  100174647   100091076   100045118   100040321   100074501   100104746   100029684   100099348   100162182   100019586   100072853   100049289   100104176   100041382   100219461 (01)00884838020054(21)100219461  100030612   100194307 (01)00884838020054(21)100194307  100141908   100146616   100114574   100103903   100098328   201015055   100008064   100080055   100184935   100141151   100321030 (01)00884838020054(21)100321030  100151261   100094893   201002554   100376321 (01)00884838020054(21)100376321  100129531   100100458   100021832   100271329 (01)00884838020054(21)100271329  100032480   100313205 (01)00884838020054(21)100313205  100317516 (01)00884838020054(21)100317516  100016741   201014801   100308413 (01)00884838020054(21)100308413  100084648   100072762   100047833   100093241   100037840   100030457   100079232   201026061 (01)00884838020054(21)201026061  100042652   100160324   100038601   100028868   201013195   100160571   201022517 (01)00884838020054(21)201022517  100017108   100141905   100267095 (01)00884838020054(21)100267095  100042239   100094899   100041918   100065811   100301611 (01)00884838020054(21)100301611  100059963   100165031   100038517   201019540   100313200 (01)00884838020054(21)100313200  201009840   201019632   100069024   100104379   100109740   100101669   100095636   100065162   100246095 (01)00884838020054(21)100246095  100226211 (01)00884838020054(21)100226211  100069460   100101552   100102516   100035611   100093715   100096086   100035435   100066486   100182917   100084209   100093899   100287018 (01)00884838020054(21)100287018  100080132   100172596   100356227 (01)00884838020054(21)100356227  100109731   201021047   100263576 (01)00884838020054(21)100263576  100088719   100158858   100048646   100087729   100009516   100075678   100120692   201018025   100168616   100009560   MX0001282 (01)00884838020054(21)MX0001282  100184947   100073400   100026608   100131419   100283908 (01)00884838020054(21)100283908  100086414   100150050   100120436   100193230   100023426   100231683 (01)00884838020054(21)100231683  100006702   100263123 (01)00884838020054(21)100263123  100162173   100012868   100147666   100366526 (01)00884838020054(21)100366526  100032466   100152913   100405716 (01)00884838020054(21)100405716  100085041   100133402   100012013   100097820   100048191   100141127   100146937   100325532 (01)00884838020054(21)100325532  100122265   100109638   100266148 (01)00884838020054(21)100266148  100303075 (01)00884838020054(21)100303075  100275316 (01)00884838020054(21)100275316  100113480   100094902   100149266   201015963   100039431   100038926   100188504   100039407   100188604   100034823   100098502   100072440   100272251 (01)00884838020054(21)100272251  100254975 (01)00884838020054(21)100254975  100295496 (01)00884838020054(21)100295496  100255292 (01)00884838020054(21)100255292  201026729 (01)00884838020054(21)201026729  100358924 (01)00884838020054(21)100358924  201000812   100090444   100364111 (01)00884838020054(21)100364111  100188448   100071618   100295209 (01)00884838020054(21)100295209  201029414 (01)00884838020054(21)201029414  100179763   100059719   100130366   100063470   201026989 (01)00884838020054(21)201026989  100031668   100141800   100237788 (01)00884838020054(21)100237788  100085816   100082652   201017931   100112564   201017003   100064349   100177838   100126519   100060789   100341090 (01)00884838020054(21)100341090  201001581   100182554   100043824   100238926 (01)00884838020054(21)100238926  100030481   100126512   201035555 (01)00884838020054(21)201035555  100025991   100258929 (01)00884838020054(21)100258929  100072203   100042033   100063898   100081325   100021509   100193625   100308031 (01)00884838020054(21)100308031  100092035   100026943   201025122 (01)00884838020054(21)201025122  100065168   100271321 (01)00884838020054(21)100271321  100337334 (01)00884838020054(21)100337334  100265704 (01)00884838020054(21)100265704  100112656   100295495 (01)00884838020054(21)100295495  100240321 (01)00884838020054(21)100240321  100038802   100067313   100179135   100102211   100102210   100086960   100066312   100303488 (01)00884838020054(21)100303488  100037046   100219469 (01)00884838020054(21)100219469  100007639   100160180   100185636   100236227 (01)00884838020054(21)100236227  100255157 (01)00884838020054(21)100255157  100057739   100066441   100081565   100074008   100098850   100061143   100369254 (01)00884838020054(21)100369254  100245567 (01)00884838020054(21)100245567  100230208 (01)00884838020054(21)100230208  100230214 (01)00884838020054(21)100230214  100021949   100236039 (01)00884838020054(21)100236039  100040164   100101197   100330231 (01)00884838020054(21)100330231  100193008   100109283   100060791   100133416   100095557   100041758   100050633   201003311   100113086   100182908   100094536   100100485   100312925 (01)00884838020054(21)100312925  100195186 (01)00884838020054(21)100195186  100380041 (01)00884838020054(21)100380041  100224094 (01)00884838020054(21)100224094  100055003   100042116   100224977 (01)00884838020054(21)100224977  100246093 (01)00884838020054(21)100246093  100028851   100040320   100014423   100048204   201029993 (01)00884838020054(21)201029993  100064423   100177871   100056772   100118886   100042650   100054201   100060709   100120425   201027365 (01)00884838020054(21)201027365  100028944   100273497 (01)00884838020054(21)100273497  100028950   100034411   100082333   100135924   100035427   100157487   100007041   100091067   100031678   100109218   100057998   100037768   100222615 (01)00884838020054(21)100222615  100087268   100029519   100029678   100102514   100029509   201015058   100029690   100227577 (01)00884838020054(21)100227577  100299596 (01)00884838020054(21)100299596  100229562 (01)00884838020054(21)100229562  100140437   100230444 (01)00884838020054(21)100230444  100048636   100066307   100057871   100186214   100010165   100172524   100109736   100112659   201008826   100010382   100250719 (01)00884838020054(21)100250719  100181999   100182620   100331541 (01)00884838020054(21)100331541  100043346   100154405   100017113   100245839 (01)00884838020054(21)100245839  100004085   100035421   100240305 (01)00884838020054(21)100240305  100267137 (01)00884838020054(21)100267137  100034384   100308633 (01)00884838020054(21)100308633  100093248   100191900   100010517   100082082   100081555   100034700   100093794   100035829   100088750   100081556   201027069 (01)00884838020054(21)201027069  100082085   100044785   100044624   100321103 (01)00884838020054(21)100321103  100081191   100310588 (01)00884838020054(21)100310588  100034684   100105945   100066842   100158398   100066228   201035577 (01)00884838020054(21)201035577  100240130 (01)00884838020054(21)100240130  100105302   100182589   100158467   201003072   100091038   100290548 (01)00884838020054(21)100290548  201022554 (01)00884838020054(21)201022554  100157685   100157507   100149587   100323688 (01)00884838020054(21)100323688  100188860   100042468   100057963   100052079   100265436 (01)00884838020054(21)100265436  100246127 (01)00884838020054(21)100246127  100072843   100026620   100073353   100168982   100135274   100039408   201016621   100148657   100259394 (01)00884838020054(21)100259394  100109268   100223781 (01)00884838020054(21)100223781  100372400 (01)00884838020054(21)100372400  100032536   100068393   201003183   100295215 (01)00884838020054(21)100295215  100082095   100249308 (01)00884838020054(21)100249308  100072281   100049349   100090296   100005141   100069116   100049908   100019431   100230486 (01)00884838020054(21)100230486  100124740   100281722 (01)00884838020054(21)100281722  100051223   100184907   100052280   100087275   100113092   100068382   100114937   100093098   100142052   100131425   100075573   100098762   100042110   100061150   100288829 (01)00884838020054(21)100288829  100119698   100121832   100130800   100130268   201014657   100117224   100117200   100175403   100062635   100023728   201011043   100043833   100117740   100134874   100076883   100190707   100068572   201009846   100094051   100060696   100027878   100170121   100170109   100159165   100069117   100045031   100073348   100102759   100327223 (01)00884838020054(21)100327223  201032016 (01)00884838020054(21)201032016  100194710 (01)00884838020054(21)100194710  100302369 (01)00884838020054(21)100302369  100030460   100095334   100141507   100034039   100063602   100008649   100275527 (01)00884838020054(21)100275527  100057523   100066318   100037651   100085374   100096873   100085363   100037832   100097609   100172587   100037833   201009784   100149311   201006557   100067202   100034162   100047722   100085375   100379369 (01)00884838020054(21)100379369  100085367   100295633 (01)00884838020054(21)100295633  100299585 (01)00884838020054(21)100299585  100272103 (01)00884838020054(21)100272103  100250575 (01)00884838020054(21)100250575  100062642   100300900 (01)00884838020054(21)100300900  100299448 (01)00884838020054(21)100299448  100289849 (01)00884838020054(21)100289849  100031667   100066905   100187244   100030834   201038152 (01)00884838020054(21)201038152  100263673 (01)00884838020054(21)100263673  100299300 (01)00884838020054(21)100299300  100006365   100037868   100131483   100073175   201001308   100048648   100047181   100322246 (01)00884838020054(21)100322246  100280732 (01)00884838020054(21)100280732  100062995   100331190 (01)00884838020054(21)100331190  100141745   201029415 (01)00884838020054(21)201029415  100306793 (01)00884838020054(21)100306793  100157681   100063415   100253619 (01)00884838020054(21)100253619  201022503 (01)00884838020054(21)201022503  100184877   201026966 (01)00884838020054(21)201026966  100392865 (01)00884838020054(21)100392865  100010514   100114580   100263197 (01)00884838020054(21)100263197  100316710 (01)00884838020054(21)100316710  100161352   100240668 (01)00884838020054(21)100240668  100057022   100149316   100140073   100288399 (01)00884838020054(21)100288399  100066474   100024472   100029500   100065698   100025788   100251067 (01)00884838020054(21)100251067  100043732   201032024 (01)00884838020054(21)201032024  201032088 (01)00884838020054(21)201032088  100101277   100283104 (01)00884838020054(21)100283104  100194732 (01)00884838020054(21)100194732  100081558   100047388   100050003   100016012   100118898   100020219   100141923   100048278   100062890   100121707   100064991   100110114   100065228   100041749   100042392   100371791 (01)00884838020054(21)100371791  100112577   201034617 (01)00884838020054(21)201034617  201034622 (01)00884838020054(21)201034622  201019550   100148007   100065700   100031000   100240662 (01)00884838020054(21)100240662  MX0001821 (01)00884838020054(21)MX0001821  100303069 (01)00884838020054(21)100303069  100103239   100113098   201029515 (01)00884838020054(21)201029515  100093442   100068595   201030025 (01)00884838020054(21)201030025  100092059   100273693 (01)00884838020054(21)100273693  100158649   100193806   100427792   100121330   100006895   100266679 (01)00884838020054(21)100266679  100107656   100368889 (01)00884838020054(21)100368889    OUS (International) System Serial Number/UDI Code:  100410528 (01)00884838089297(21)100410528  100342563   100261002 (01)00884838009851(21)100261002  100391108 (01)00884838089297(21)100391108  100345521   100027294   201033669 (01)00884838009851(21)201033669  100078694   100029332   100378456 (01)00884838009851(21)100378456  100143739   100144969   100243519 (01)00884838009851(21)100243519  100341392 (01)00884838009851(21)100341392  201029945 (01)00884838009851(21)201029945  100221160 (01)00884838009851(21)100221160  100368048 (01)00884838009851(21)100368048  100071881   100132030   100087603   100345113   100343567   100026763   100343580   100125173   100257206 (01)00884838009851(21)100257206  100020049   100250344 (01)00884838025776(21)100250344  100142326   201029762 (01)00884838009844(21)201029762  100170107   100021689   100060108   100169767   100030132   100248621 (01)00884838009851(21)100248621  100026671   100230020 (01)00884838009844(21)100230020  100094352   100406417 (01)00884838089297(21)100406417  100027974   100103663   201004367   100256852 (01)00884838009851(21)100256852  100158086   100267691 (01)00884838009851(21)100267691  100132108   100004667   100007329   100036438   100352131 (01)00884838009851(21)100352131  100175815   100139917   100189486   100071130   100172257   201013817   201002709   100078696   100003246   100086773   100118670   100088051   100189489   201002273   100247417 (01)00884838009851(21)100247417  100346921 (01)00884838009851(21)100346921  100176338   100278791 (01)00884838009851(21)100278791  100285001 (01)00884838009851(21)100285001  100190085   100006111   100230029 (01)00884838009851(21)100230029  100244201 (01)00884838009844(21)100244201  100179385   100239534 (01)00884838009851(21)100239534  201029776 (01)00884838009851(21)201029776  100110677   100243224 (01)00884838009851(21)100243224  100348087 (01)00884838009851(21)100348087  100144874   100179382   100229633 (01)00884838009844(21)100229633  100378459 (01)00884838089297(21)100378459  100410514   100315894   100126161   100133620   100094393   100144981   100230026 (01)00884838009851(21)100230026  100233964 (01)00884838009851(21)100233964  100015334   201022999 (01)00884838009851(21)201022999  100244590 (01)00884838009851(21)100244590  100028670   201029115 (01)00884838009851(21)201029115  100006114   100118695   100176847   100110846   100371670 (01)00884838009851(21)100371670  100189477   100189480   100078367   100169650   100172258   100381925 (01)00884838009851(21)100381925  201010821   100166085   100043299   100045681   201029918 (01)00884838009851(21)201029918  100234781 (01)00884838009851(21)100234781  100219997 (01)00884838009851(21)100219997  100082433   100104065   100024697   100011191   100259879 (01)00884838009851(21)100259879  201033640 (01)00884838009851(21)201033640  100237375 (01)00884838009851(21)100237375  100247101 (01)00884838009851(21)100247101  100289690   100189819   100254748 (01)00884838009851(21)100254748  100116378   201033644 (01)00884838009851(21)201033644  100237374 (01)00884838009844(21)100237374  201002710   100104055",
      "more_code_info": "100278758 (01)00884838009851(21)100278758  100145123   100232758 (01)00884838009851(21)100232758  100055938   100395579 (01)00884838089297(21)100395579  201010183   100237630 (01)00884838009844(21)100237630  201001547   100124977   100124980   201007601   100315184 (01)00884838009851(21)100315184  201010864   100176022   100046415   100092492   100260037 (01)00884838009851(21)100260037  100186765   100043298   100152333   100163590   100039840   100278756 (01)00884838009851(21)100278756  100110675   100233967 (01)00884838009851(21)100233967  100259870 (01)00884838009851(21)100259870  100274760 (01)00884838009851(21)100274760  100351158   100081755   100279007 (01)00884838009851(21)100279007  100045726   100135544   100257671 (01)00884838009851(21)100257671  100219976 (01)00884838009851(21)100219976  100097395   201031754 (01)00884838009851(21)201031754  100170696   100070576   100144980   100029455   100174449   100029075   100139896   100136805   100110672   100243918 (01)00884838009851(21)100243918  100226891 (01)00884838009851(21)100226891  100388081 (01)00884838009851(21)100388081  100219290 (01)00884838009851(21)100219290  100015329   100026097   100027970   100239303 (01)00884838009851(21)100239303  100138231   100406482 (01)00884838009851(21)100406482  100097980   100292345 (01)00884838009851(21)100292345  100415676   201029884 (01)00884838009851(21)201029884  100277916 (01)00884838009851(21)100277916  100092187   100348161 (01)00884838009844(21)100348161  100029460   100055947   100312444 (01)00884838009851(21)100312444  100286657 (01)00884838009851(21)100286657  100026680   201032697 (01)00884838009851(21)201032697  201032759 (01)00884838009851(21)201032759  100152551   100282843 (01)00884838009851(21)100282843  100369435 (01)00884838009851(21)100369435  100341862 (01)00884838089297(21)100341862  100089812   100190080   100007194   100026666   100343575   100007411   100290662 (01)00884838009851(21)100290662  100019394   100343690   100406654 (01)00884838089297(21)100406654  100378670 (01)00884838089297(21)100378670  100259999 (01)00884838009851(21)100259999  100358186   100345848 (01)00884838089297(21)100345848  100375915 (01)00884838089297(21)100375915  100003612   100343639   100345599   100345547 (01)00884838089297(21)100345547  100042753   100343610   100343599   100037544   100314705   100345527   100354459   100375828   100343611   100404856 (01)00884838089297(21)100404856  100375832   100003339   100251169 (01)00884838009844(21)100251169  100247260 (01)00884838009851(21)100247260  100011042   100011044   100266744 (01)00884838009844(21)100266744  100003160   100271022 (01)00884838009851(21)100271022  100029459   100003243   100456025 (01)00884838009851(21)100456025  100064711   100055828   100013968   100026094   100137503   100247108 (01)00884838009851(21)100247108  100252969 (01)00884838009851(21)100252969  100189485   201002380   100065597   100037555   100055935   100231067 (01)00884838009851(21)100231067  100149705   100098083   100159650   100274445 (01)00884838009851(21)100274445  100004673   100045725   100160838   100065594   100102051   100350166 (01)00884838009851(21)100350166  100190008   100027301   100114025   100164088   100039849   100346973 (01)00884838009851(21)100346973  100219281 (01)00884838009851(21)100219281  100057375   100237625 (01)00884838009844(21)100237625  100110942   100100996   100368016 (01)00884838009851(21)100368016  100095586   100080824   100319297 (01)00884838025776(21)100319297  201027752 (01)00884838025776(21)201027752  100333778 (01)00884838025776(21)100333778  100125719   100120746   100289711   100081706   100226369 (01)00884838009851(21)100226369  100005815   100315391 (01)00884838009844(21)100315391  100252774 (01)00884838009844(21)100252774  201032861 (01)00884838009851(21)201032861  100176321   100357773 (01)00884838009851(21)100357773  100132408   100092646   100174964   100005813   201029728 (01)00884838009844(21)201029728  100026665   100063151   100274692 (01)00884838009851(21)100274692  100169364   100282846 (01)00884838009851(21)100282846  100239390 (01)00884838009851(21)100239390  100286304 (01)00884838009851(21)100286304  100231171 (01)00884838009844(21)100231171  100021402   100278998 (01)00884838009851(21)100278998  100279463 (01)00884838009851(21)100279463  100019389   100013971   100237192 (01)00884838009851(21)100237192  100149718   100060588   100020745   100361234 (01)00884838009851(21)100361234  100237302 (01)00884838009851(21)100237302  100174967   100391367 (01)00884838009844(21)100391367  100252765 (01)00884838009851(21)100252765  201029778 (01)00884838009851(21)201029778  100082434   201010835   100064708   100367183 (01)00884838009851(21)100367183  100114463   201001571   100029469   100030113   100053793   100069521   100014773   100084821   100228994 (01)00884838009851(21)100228994  100163725   100277890 (01)00884838009851(21)100277890  201015439   201029822 (01)00884838009851(21)201029822  201025393 (01)00884838009851(21)201025393  201025923 (01)00884838009851(21)201025923  100071127   100415765 (01)00884838009851(21)100415765"
    }
  ]
}