  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-05-06",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Concord",
      "address_1": "4040 Nelson Ave",
      "reason_for_recall": "Potential safety issue when using the IMMEDIATE RESUMPTION feature after a Control Console interlock during the delivery of flattening-filter-free (Multiple-X) beams or segments. In a very rare scenario, the RT Therapist will send again the previously completed beam or segment for delivery to the control console, thus repeating the already delivered Monitor Units of the same beam or segment.",
      "address_2": "",
      "product_quantity": "18",
      "code_info": "ARTISTE, Part #: 08139789,  ONCOR Avant Garde, Part #: 05863472,  ONCOR Impression Plus, Part #: 05857912,  ONCOR Expression, Part #: 07360717",
      "center_classification_date": "20120724",
      "distribution_pattern": "Worldwide Distribution, including the state of Florida (USA) and the countries of Germany, India, Poland, Thailand, and the Republic of Korea.",
      "state": "CA",
      "product_description": "ARTISTE or ONCOR Digital Linear Accelerators with syngo RT Therapist version 4.2/MOSAIQ OIS Systems and Multiple-X, Manufactured by SIEMENS AG.  The ARTISTE and ONCOR family of Linear Accelerators are used to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer.",
      "report_date": "20120801",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Siemens Medical Solutions USA,  Inc",
      "recall_number": "Z-2073-2012",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "62519",
      "termination_date": "20121018",
      "more_code_info": "",
      "recall_initiation_date": "20120625",
      "postal_code": "94520-1200",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}