  "results": [
    {
      "status": "Terminated",
      "city": "Grove City",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79813",
      "recalling_firm": "Tosoh Bioscience Inc",
      "address_1": "3600 Gantz Rd",
      "address_2": "N/A",
      "postal_code": "43123-1895",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: US (Nationwide) and countries of; Canada, Mexico, Columbia, Chile, Uruguay, Nicaragua, Argentina, Peru, Chile, Honduras. Equador, Caymen Islands, Panama, and Costa Rica.",
      "recall_number": "Z-2072-2018",
      "product_description": "ST-AIA PACK Myoglobin; Part Number: 025297     Assay, Cardiac Marker",
      "product_quantity": "733",
      "reason_for_recall": "Asfotase Alfa (Strensiq) interferes with certain Tosoh assays which potentially causes falsely increased or falsely decreased immunoassay test results, depending upon the assay.",
      "recall_initiation_date": "20180305",
      "center_classification_date": "20180605",
      "termination_date": "20200831",
      "report_date": "20180613",
      "code_info": "All Lot Codes"
    }
  ]
}