      "status": "Terminated",
      "city": "Neptune",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "68573",
      "recalling_firm": "Excelsior Medical Corp",
      "address_1": "1933 Heck Ave",
      "address_2": "N/A",
      "postal_code": "07753",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and Canada.",
      "recall_number": "Z-2070-2014",
      "product_description": "Excelsior Medical Sterile Triple Lead Tubing Set and Sterile Heavy Duty Triple Lead Tubing Set.  To be used with the Pharm-Assist Dispensing Pump.   Used for drug reconstitution and/or fluid transfer",
      "product_quantity": "4250 tubing sets (PA03 - 78 cases; BX03 - 92 cases (each case contains 25 tubing sets))",
      "reason_for_recall": "There is a potential for a leak.   A molding defect in the connector that joins the three leads to the main flow line may allow fluid to escape or air to enter the fluid path during transfer of fluids from one container to another.",
      "recall_initiation_date": "20140605",
      "center_classification_date": "20140716",
      "termination_date": "20150819",
      "report_date": "20140723",
      "code_info": "Product Code PA03 - Lot Nos.:  03030813, 03050813 and 03091213    Product Code BX03, Lot Nos.: 64031513, 64073113, 64081313, 64090513"
    }
  ]
}