      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "68344",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA nationwide, ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BRAZIL, CANADA, CHILE, CHINA, Croatia, CZECH REPUBLIC, DENMARK, EGYPT, EL SALVADOR, FINLAND, FRANCE, GERMANY, GREECE, GUATEMALA, HONG KONG, INDIA, IRAQ, ISRAEL, ITALY, JAPAN, KAZAKHSTAN, REPUBLIC OF KOREA, MALAYSIA, NETHERLANDS, NEW ZEALAND, OMAN, PERU, POLAND, ROMANIA, RUSSIAN FEDERATION, SAUDI ARABIA, SINGAPORE, SPAIN, SWEDEN, SWITZERLAND, SYRIAN ARAB REPUBLIC, TAIWAN, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, VENEZUELA, LEBANON, REPUBLIC OF MALDOVIA, COSTA RICA, GABON, INDONESIA, JAMAICA, NIGERIA, NORWAY, NEPAL, PAKISTAN, PANAMA, PHILIPPINES, SENEGAL, SOUTH AFRICA, THAILAND, TUNISIA, and CAMBODIA.",
      "recall_number": "Z-2068-2014",
      "product_description": "DATEX-OHMEDA S/5 SINGLE-WIDTH AIRWAY MODULE, E-MINIC AND ACCESSORIES.     DATEX-OHMEDA S/5TM FM WITH L-FICU04 ANDL-FICUO4A SOFTWARE AND N-FCREC MODULE **(OUS only).     E-miniC and accessories is indicated for monitoring C02 and respiration rate of all hospital patients.",
      "product_quantity": "2,948 (761 units US, 2,187 units OUS)",
      "reason_for_recall": "Failure of the CO2 detector in Single-width Airway and Extension modules, which may cause a slow continuous decrease of measured Et-/FiCO2 values. Incorrect Et CO2/Fi CO2 value may impair clinical decision making for both mechanically and spontaneously ventilated patients because of incorrect low CO2 values.",
      "recall_initiation_date": "20140611",
      "center_classification_date": "20140718",
      "termination_date": "20160329",
      "report_date": "20140730",
      "code_info": "Serial # Model #   6837005 E-MINIC-00  6838210 E-MINIC-00  6845207 E-MINIC-00  6845218 E-MINIC-00  6845228 E-MINIC-00  6845248 E-MINIC-00  6860768 E-MINIC-00  6860858 E-MINIC-00  6876048 E-MINIC-00  6876050 E-MINIC-00  6876052 E-MINIC-00  6876054 E-MINIC-00  6878048 E-MINIC-00  6888716 E-MINIC-00  6888726 E-MINIC-00  6888755 E-MINIC-00  6889935 E-MINIC-00  6889950 E-MINIC-00  6891094 E-MINIC-00  6891114 E-MINIC-00  6891115 E-MINIC-00  6891116 E-MINIC-00  6891117 E-MINIC-00  6891121 E-MINIC-00  6891130 E-MINIC-00  6891137 E-MINIC-00  6891139 E-MINIC-00  6892616 E-MINIC-00  6892626 E-MINIC-00  6827924 E-MINIC-00  6829318 E-MINIC-00  6829321 E-MINIC-00  6832785 E-MINIC-00  6832787 E-MINIC-00  6832788 E-MINIC-00  6832789 E-MINIC-00  6832791 E-MINIC-00  6832795 E-MINIC-00  6832796 E-MINIC-00  6832797 E-MINIC-00  6832802 E-MINIC-00  6832803 E-MINIC-00  6832804 E-MINIC-00  6832805 E-MINIC-00  6832813 E-MINIC-00  6832815 E-MINIC-00  6832824 E-MINIC-00  6832825 E-MINIC-00  6832831 E-MINIC-00  6834874 E-MINIC-00  6834877 E-MINIC-00  6834880 E-MINIC-00  6834883 E-MINIC-00  6834884 E-MINIC-00  6834885 E-MINIC-00  6834886 E-MINIC-00  6834887 E-MINIC-00  6834888 E-MINIC-00  6834889 E-MINIC-00  6834890 E-MINIC-00  6834891 E-MINIC-00  6834892 E-MINIC-00  6834893 E-MINIC-00  6834894 E-MINIC-00  6834895 E-MINIC-00  6834896 E-MINIC-00  6834898 E-MINIC-00  6834899 E-MINIC-00  6834902 E-MINIC-00  6834903 E-MINIC-00  6834904 E-MINIC-00  6834907 E-MINIC-00  6834909 E-MINIC-00  6834910 E-MINIC-00  6834911 E-MINIC-00  6834913 E-MINIC-00  6834914 E-MINIC-00  6834915 E-MINIC-00  6834916 E-MINIC-00  6834917 E-MINIC-00  6834918 E-MINIC-00  6834919 E-MINIC-00  6834920 E-MINIC-00  6834921 E-MINIC-00  6834922 E-MINIC-00  6836999 E-MINIC-00  6837002 E-MINIC-00  6837003 E-MINIC-00  6837010 E-MINIC-00  6837011 E-MINIC-00  6837013 E-MINIC-00  6838187 E-MINIC-00  6838196 E-MINIC-00  6838197 E-MINIC-00  6838198 E-MINIC-00  6842192 E-MINIC-00  6842196 E-MINIC-00  6842198 E-MINIC-00  6842212 E-MINIC-00  6842221 E-MINIC-00  6848886 E-MINIC-00  6848887 E-MINIC-00  6848888 E-MINIC-00  6852650 E-MINIC-00  6852655 E-MINIC-00  6855674 E-MINIC-00  6860683 E-MINIC-00  6860685 E-MINIC-00  6860698 E-MINIC-00  6860750 E-MINIC-00  6873156 E-MINIC-00  6879423 E-MINIC-00  6883991 E-MINIC-00  6883994 E-MINIC-00  6883995 E-MINIC-00  6883997 E-MINIC-00  6884000 E-MINIC-00  6884007 E-MINIC-00  6884014 E-MINIC-00  6884016 E-MINIC-00  6884022 E-MINIC-00  6884023 E-MINIC-00  6887178 E-MINIC-00  6887182 E-MINIC-00  6887183 E-MINIC-00  6887185 E-MINIC-00  6887186 E-MINIC-00  6887187 E-MINIC-00  6887189 E-MINIC-00  6887190 E-MINIC-00  6887194 E-MINIC-00  6887195 E-MINIC-00  6887199 E-MINIC-00  6887201 E-MINIC-00  6887203 E-MINIC-00  6887206 E-MINIC-00  6887209 E-MINIC-00  6887210 E-MINIC-00  6887221 E-MINIC-00  6887222 E-MINIC-00  6888655 E-MINIC-00  6888670 E-MINIC-00  6888672 E-MINIC-00  6888681 E-MINIC-00  6888683 E-MINIC-00  6888684 E-MINIC-00  6888686 E-MINIC-00  6888688 E-MINIC-00  6888699 E-MINIC-00  6888703 E-MINIC-00  6888749 E-MINIC-00  6888750 E-MINIC-00  6888754 E-MINIC-00  6888771 E-MINIC-00  6888772 E-MINIC-00  6888773 E-MINIC-00  6889916 E-MINIC-00  6889939 E-MINIC-00  6889967 E-MINIC-00  6891144 E-MINIC-00  6891146 E-MINIC-00  6891148 E-MINIC-00  6891150 E-MINIC-00  6891152 E-MINIC-00  6891162 E-MINIC-00  6891180 E-MINIC-00  6891182 E-MINIC-00  6891184 E-MINIC-00  6896279 E-MINIC-00  6896303 E-MINIC-00  6838209 E-MINIC-00  6838211 E-MINIC-00  6838212 E-MINIC-00  6892655 E-MINIC-00  6892720 E-MINIC-00  6832823 E-MINIC-00  6838243 E-MINIC-00  6838252 E-MINIC-00  6838256 E-MINIC-00  6838262 E-MINIC-00  6842175 E-MINIC-00  6842181 E-MINIC-00  6842190 E-MINIC-00  6842204 E-MINIC-00  6842206 E-MINIC-00  6842217 E-MINIC-00  6842218 E-MINIC-00  6842219 E-MINIC-00  6842224 E-MINIC-00  6842225 E-MINIC-00  6842226 E-MINIC-00  6842227 E-MINIC-00  6842228 E-MINIC-00  6842230 E-MINIC-00  6842231 E-MINIC-00  6842233 E-MINIC-00  6842234 E-MINIC-00  6842236 E-MINIC-00  6842239 E-MINIC-00  6842242 E-MINIC-00  6842244 E-MINIC-00  6848877 E-MINIC-00  6848878 E-MINIC-00  6848881 E-MINIC-00  6848883 E-MINIC-00  6848884 E-MINIC-00  6848885 E-MINIC-00  6848889 E-MINIC-00  6848895 E-MINIC-00  6848897 E-MINIC-00  6852617 E-MINIC-00  6852641 E-MINIC-00  6852656 E-MINIC-00  6852662 E-MINIC-00  6852667 E-MINIC-00  6855627 E-MINIC-00  6855630 E-MINIC-00  6855719 E-MINIC-00  6860661 E-MINIC-00  6860720 E-MINIC-00  6860799 E-MINIC-00  6860821 E-MINIC-00  6860823 E-MINIC-00  6860825 E-MINIC-00  6860827 E-MINIC-00  6867789 E-MINIC-00  6870556 E-MINIC-00  6870574 E-MINIC-00  6870594 E-MINIC-00  6873165 E-MINIC-00  6873182 E-MINIC-00  6873190 E-MINIC-00  6876007 E-MINIC-00  6878050 E-MINIC-00  6878054 E-MINIC-00  6878056 E-MINIC-00  6883990 E-MINIC-00  6883998 E-MINIC-00  6884001 E-MINIC-00  6884006 E-MINIC-00  6884008 E-MINIC-00  6884010 E-MINIC-00  6884015 E-MINIC-00  6884020 E-MINIC-00  6887188 E-MINIC-00  6887198 E-MINIC-00  6887204 E-MINIC-00  6887211 E-MINIC-00  6887215 E-MINIC-00  6887216 E-MINIC-00  6887226 E-MINIC-00  6888661 E-MINIC-00  6888711 E-MINIC-00  6888719 E-MINIC-00  6888724 E-MINIC-00  6889920 E-MINIC-00  6889936 E-MINIC-00  6889942 E-MINIC-00  6889943 E-MINIC-00  6889948 E-MINIC-00  6889954 E-MINIC-00  6889957 E-MINIC-00  6889958 E-MINIC-00  6889959 E-MINIC-00  6889963 E-MINIC-00  6889966 E-MINIC-00  6891059 E-MINIC-00  6891069 E-MINIC-00  6891070 E-MINIC-00  6891080 E-MINIC-00  6891081 E-MINIC-00  6891090 E-MINIC-00  6891131 E-MINIC-00  6891143 E-MINIC-00  6892659 E-MINIC-00  6892665 E-MINIC-00  6892676 E-MINIC-00  6892680 E-MINIC-00  6892681 E-MINIC-00  6892694 E-MINIC-00  6892700 E-MINIC-00  6892701 E-MINIC-00  6895014 E-MINIC-00  6896275 E-MINIC-00  6896372 E-MINIC-00  6898678 E-MINIC-00  6898687 E-MINIC-00  6898696 E-MINIC-00  6838247 E-MINIC-00  6838251 E-MINIC-00  6838254 E-MINIC-00  6838259 E-MINIC-00  6842167 E-MINIC-00  6842182 E-MINIC-00  6868294 E-MINIC-00  6868320 E-MINIC-00  6870587 E-MINIC-00  6873125 E-MINIC-00  6873128 E-MINIC-00  6873130 E-MINIC-00  6873131 E-MINIC-00  6873139 E-MINIC-00  6873141 E-MINIC-00  6824774 E-MINIC-00  6824782 E-MINIC-00  6824783 E-MINIC-00  6860780 E-MINIC-00  6845224 E-MINIC-00  6845235 E-MINIC-00  6824790 E-MINIC-00  6824792 E-MINIC-00  6824797 E-MINIC-00  6824799 E-MINIC-00  6824800 E-MINIC-00  6824802 E-MINIC-00  6824804 E-MINIC-00  6824811 E-MINIC-00  6824812 E-MINIC-00  6824818 E-MINIC-00  6832801 E-MINIC-00  6832817 E-MINIC-00  6834901 E-MINIC-00  6836976 E-MINIC-00  6836980 E-MINIC-00  6829311 E-MINIC-00  6848912 E-MINIC-00  6860830 E-MINIC-00  6832821 E-MINIC-00  6842180 E-MINIC-00  6860849 E-MINIC-00  6889956 E-MINIC-00  6891173 E-MINIC-00  6892607 E-MINIC-00  6832790 E-MINIC-00  6845227 E-MINIC-00  6891055 E-MINIC-00  6892614 E-MINIC-00  6822945 E-MINIC-00  6822949 E-MINIC-00  6832794 E-MINIC-00  6832812 E-MINIC-00  6836997 E-MINIC-00  6836998 E-MINIC-00  6838214 E-MINIC-00  6838224 E-MINIC-00  6838228 E-MINIC-00  6838236 E-MINIC-00  6838237 E-MINIC-00  6838245 E-MINIC-00  6842177 E-MINIC-00  6842241 E-MINIC-00  6845206 E-MINIC-00  6848890 E-MINIC-00  6848898 E-MINIC-00  6848908 E-MINIC-00  6848909 E-MINIC-00  6848913 E-MINIC-00  6848915 E-MINIC-00  6848916 E-MINIC-00  6848942 E-MINIC-00  6852653 E-MINIC-00  6852661 E-MINIC-00  6852678 E-MINIC-00  6860438 E-MINIC-00  6860667 E-MINIC-00  6860777 E-MINIC-00  6870568 E-MINIC-00  6873123 E-MINIC-00  6873154 E-MINIC-00  6888695 E-MINIC-00  6888712 E-MINIC-00  6888738 E-MINIC-00  6891128 E-MINIC-00  6891168 E-MINIC-00  6891175 E-MINIC-00  6891176 E-MINIC-00  6891178 E-MINIC-00  6891181 E-MINIC-00  6892618 E-MINIC-00  6892625 E-MINIC-00  6892646 E-MINIC-00  6892693 E-MINIC-00  6892698 E-MINIC-00  6868321 E-MINIC-00  6842220 E-MINIC-00  6848914 E-MINIC-00  6848924 E-MINIC-00  6848936 E-MINIC-00  6852611 E-MINIC-00  6852612 E-MINIC-00  6852619 E-MINIC-00  6852622 E-MINIC-00  6852664 E-MINIC-00  6852666 E-MINIC-00  6852668 E-MINIC-00  6852669 E-MINIC-00  6852674 E-MINIC-00  6852675 E-MINIC-00  6852676 E-MINIC-00  6852677 E-MINIC-00  6852679 E-MINIC-00  6852680 E-MINIC-00  6852681 E-MINIC-00  6852682 E-MINIC-00  6855629 E-MINIC-00  6855639 E-MINIC-00  6855643 E-MINIC-00  6855647 E-MINIC-00  6855649 E-MINIC-00  6855651 E-MINIC-00  6855653 E-MINIC-00  6855660 E-MINIC-00  6855661 E-MINIC-00  6855664 E-MINIC-00  6855671 E-MINIC-00  6855673 E-MINIC-00  6855678 E-MINIC-00  6855680 E-MINIC-00  6855681 E-MINIC-00  6855683 E-MINIC-00  6855685 E-MINIC-00  6855686 E-MINIC-00  6855691 E-MINIC-00  6855695 E-MINIC-00  6855699 E-MINIC-00  6855701 E-MINIC-00  6855702 E-MINIC-00  6855704 E-MINIC-00  6855705 E-MINIC-00  6855709 E-MINIC-00  6855712 E-MINIC-00  6855713 E-MINIC-00  6855715 E-MINIC-00  6855717 E-MINIC-00  6855718 E-MINIC-00  6855722 E-MINIC-00  6855725 E-MINIC-00  6883984 E-MINIC-00  6884013 E-MINIC-00  6884019 E-MINIC-00  6895016 E-MINIC-00  6896270 E-MINIC-00  6896308 E-MINIC-00  6896310 E-MINIC-00  6896312 E-MINIC-00  6896334 E-MINIC-00  6896351 E-MINIC-00  6896356 E-MINIC-00  6896357 E-MINIC-00  6898635 E-MINIC-00  6898659 E-MINIC-00  6898679 E-MINIC-00  6898693 E-MINIC-00  6898697 E-MINIC-00  6898706 E-MINIC-00  6898707 E-MINIC-00  6898708 E-MINIC-00  6898715 E-MINIC-00  6898716 E-MINIC-00  6898720 E-MINIC-00  6898724 E-MINIC-00  6898726 E-MINIC-00  6898727 E-MINIC-00  6898729 E-MINIC-00  6898732 E-MINIC-00  6832827 E-MINIC-00  6842222 E-MINIC-00  6842238 E-MINIC-00  6855656 E-MINIC-00  6855677 E-MINIC-00  6855720 E-MINIC-00  6855728 E-MINIC-00  6855729 E-MINIC-00  6855732 E-MINIC-00  6860439 E-MINIC-00  6860444 E-MINIC-00  6860450 E-MINIC-00  6860670 E-MINIC-00  6860672 E-MINIC-00  6860761 E-MINIC-00  6860772 E-MINIC-00  6860773 E-MINIC-00  6860778 E-MINIC-00  6883980 E-MINIC-00  6883986 E-MINIC-00  6883988 E-MINIC-00  6884003 E-MINIC-00  6884004 E-MINIC-00  6884005 E-MINIC-00  6887205 E-MINIC-00  6888687 E-MINIC-00  6888693 E-MINIC-00  6889965 E-MINIC-00  6891151 E-MINIC-00  6891174 E-MINIC-00  6892656 E-MINIC-00  6892672 E-MINIC-00  6892697 E-MINIC-00  6894976 E-MINIC-00  6894998 E-MINIC-00  6895005 E-MINIC-00  6895021 E-MINIC-00  6895022 E-MINIC-00  6895023 E-MINIC-00  6895025 E-MINIC-00  6896262 E-MINIC-00  6896264 E-MINIC-00  6896266 E-MINIC-00  6896268 E-MINIC-00  6896269 E-MINIC-00  6896271 E-MINIC-00  6896272 E-MINIC-00  6896273 E-MINIC-00  6896274 E-MINIC-00  6896276 E-MINIC-00  6896280 E-MINIC-00  6896282 E-MINIC-00  6896283 E-MINIC-00  6896287 E-MINIC-00  6896290 E-MINIC-00  6896291 E-MINIC-00  6896292 E-MINIC-00  6896294 E-MINIC-00  6896296 E-MINIC-00  6896297 E-MINIC-00  6896298 E-MINIC-00  6896301 E-MINIC-00  6896307 E-MINIC-00  6896309 E-MINIC-00  6896311 E-MINIC-00  6896313 E-MINIC-00  6896315 E-MINIC-00  6896317 E-MINIC-00  6896322 E-MINIC-00  6896323 E-MINIC-00  6896332 E-MINIC-00  6896333 E-MINIC-00  6896335 E-MINIC-00  6896336 E-MINIC-00  6896337 E-MINIC-00  6896338 E-MINIC-00  6896339 E-MINIC-00  6896340 E-MINIC-00  6896341 E-MINIC-00  6896342 E-MINIC-00  6896343 E-MINIC-00  6896344 E-MINIC-00  6896345 E-MINIC-00  6896346 E-MINIC-00  6896347 E-MINIC-00  6896348 E-MINIC-00  6896349 E-MINIC-00  6896350 E-MINIC-00  6896352 E-MINIC-00  6896353 E-MINIC-00  6896354 E-MINIC-00  6896355 E-MINIC-00  6896358 E-MINIC-00  6896359 E-MINIC-00  6896360 E-MINIC-00  6896361 E-MINIC-00  6896362 E-MINIC-00  6896363 E-MINIC-00  6896364 E-MINIC-00  6896365 E-MINIC-00  6896366 E-MINIC-00  6896367 E-MINIC-00  6896368 E-MINIC-00  6896369 E-MINIC-00  6896370 E-MINIC-00  6896371 E-MINIC-00  6896373 E-MINIC-00  6896374 E-MINIC-00  6896375 E-MINIC-00  6896376 E-MINIC-00  6898627 E-MINIC-00  6898628 E-MINIC-00  6898629 E-MINIC-00  6898631 E-MINIC-00  6898632 E-MINIC-00  6898633 E-MINIC-00  6898634 E-MINIC-00  6898636 E-MINIC-00  6898638 E-MINIC-00  6898639 E-MINIC-00  6898640 E-MINIC-00  6898641 E-MINIC-00  6898642 E-MINIC-00  6898643 E-MINIC-00  6898644 E-MINIC-00  6898645 E-MINIC-00  6898646 E-MINIC-00  6898649 E-MINIC-00  6898650 E-MINIC-00  6898651 E-MINIC-00  6898653 E-MINIC-00  6898654 E-MINIC-00  6898655 E-MINIC-00  6898656 E-MINIC-00  6898657 E-MINIC-00  6898658 E-MINIC-00  6898660 E-MINIC-00  6898661 E-MINIC-00  6898662 E-MINIC-00  6898663 E-MINIC-00  6898664 E-MINIC-00  6898665 E-MINIC-00  6898666 E-MINIC-00  6898667 E-MINIC-00  6898668 E-MINIC-00  6898669 E-MINIC-00  6898670 E-MINIC-00  6898671 E-MINIC-00  6898672 E-MINIC-00  6898673 E-MINIC-00  6898674 E-MINIC-00  6898675 E-MINIC-00  6898676 E-MINIC-00  6898677 E-MINIC-00  6898680 E-MINIC-00  6898681 E-MINIC-00  6898682 E-MINIC-00  6898683 E-MINIC-00  6898684 E-MINIC-00  6898685 E-MINIC-00  6898686 E-MINIC-00  6898688 E-MINIC-00  6898691 E-MINIC-00  6898699 E-MINIC-00  6898700 E-MINIC-00  6898703 E-MINIC-00  6898705 E-MINIC-00  6898710 E-MINIC-00  6898713 E-MINIC-00  6898717 E-MINIC-00  6898718 E-MINIC-00  6898719 E-MINIC-00  6898722 E-MINIC-00  6898723 E-MINIC-00  6898725 E-MINIC-00  6898728 E-MINIC-00  6898735 E-MINIC-00  6898737 E-MINIC-00  6898742 E-MINIC-00  6898746 E-MINIC-00  6898747 E-MINIC-00  6898748 E-MINIC-00  6898758 E-MINIC-00  6898759 E-MINIC-00  6898764 E-MINIC-00  6898766 E-MINIC-00  6898767 E-MINIC-00  6898773 E-MINIC-00  6898775 E-MINIC-00  6898776 E-MINIC-00  6898777 E-MINIC-00  6824807 E-MINIC-00  6824814 E-MINIC-00  6824815 E-MINIC-00  6824816 E-MINIC-00  6824821 E-MINIC-00  6827897 E-MINIC-00  6827938 E-MINIC-00  6829322 E-MINIC-00  6832806 E-MINIC-00  6832807 E-MINIC-00  6832808 E-MINIC-00  6832816 E-MINIC-00  6891063 E-MINIC-00  6891078 E-MINIC-00  6891084 E-MINIC-00  6891110 E-MINIC-00  6891113 E-MINIC-00  6891120 E-MINIC-00  6845231 E-MINIC-00  6855657 E-MINIC-00  6855662 E-MINIC-00  6855687 E-MINIC-00  6855688 E-MINIC-00  6860731 E-MINIC-00  6887197 E-MINIC-00  6888696 E-MINIC-00  6888698 E-MINIC-00  6888702 E-MINIC-00  6888713 E-MINIC-00  6888732 E-MINIC-00  6888764 E-MINIC-00  6889955 E-MINIC-00  6889971 E-MINIC-00  6891041 E-MINIC-00  6891054 E-MINIC-00  6891064 E-MINIC-00  6891067 E-MINIC-00  6891082 E-MINIC-00  6891086 E-MINIC-00  6891088 E-MINIC-00  6891091 E-MINIC-00  6891092 E-MINIC-00  6891101 E-MINIC-00  6891102 E-MINIC-00  6891103 E-MINIC-00  6891105 E-MINIC-00  6891106 E-MINIC-00  6891111 E-MINIC-00  6892639 E-MINIC-00  6892651 E-MINIC-00  6892654 E-MINIC-00  6892678 E-MINIC-00  6892682 E-MINIC-00  6892721 E-MINIC-00  6892722 E-MINIC-00  6896261 E-MINIC-00  6852613 E-MINIC-00  6852621 E-MINIC-00  6870549 E-MINIC-00  6870564 E-MINIC-00  6892663 E-MINIC-00  6891153 E-MINIC-00  6891155 E-MINIC-00  6891156 E-MINIC-00  6832822 E-MINIC-00  6838235 E-MINIC-00  6838238 E-MINIC-00  6838240 E-MINIC-00  6892689 E-MINIC-00  6892699 E-MINIC-00  6822971 E-MINIC-00  6842243 E-MINIC-00  6848891 E-MINIC-00  6873143 E-MINIC-00  6883992 E-MINIC-00  6884011 E-MINIC-00  6884017 E-MINIC-00  6827922 E-MINIC-00  6827931 E-MINIC-00  6827932 E-MINIC-00  6827937 E-MINIC-00  6829313 E-MINIC-00  6848905 E-MINIC-00  6891166 E-MINIC-00  6852660 E-MINIC-00  6887213 E-MINIC-00  6888704 E-MINIC-00  6888737 E-MINIC-00  6888742 E-MINIC-00  6888743 E-MINIC-00  6832811 E-MINIC-00  6834923 E-MINIC-00  6834924 E-MINIC-00  6834927 E-MINIC-00  6834932 E-MINIC-00  6836970 E-MINIC-00  6836972 E-MINIC-00  6836978 E-MINIC-00  6837014 E-MINIC-00  6838190 E-MINIC-00  6896278 E-MINIC-00  6870593 E-MINIC-00  6873121 E-MINIC-00  6860684 E-MINIC-00  6860847 E-MINIC-00  6888678 E-MINIC-00  6891119 E-MINIC-00  6891154 E-MINIC-00  6892621 E-MINIC-00  6892627 E-MINIC-00  6892633 E-MINIC-00  6892692 E-MINIC-00  6894986 E-MINIC-00  6896305 E-MINIC-00  6834933 E-MINIC-00  6836975 E-MINIC-00  6836977 E-MINIC-00  6883967 E-MINIC-00  6887217 E-MINIC-00  6887219 E-MINIC-00  6888757 E-MINIC-00  6888765 E-MINIC-00  6888770 E-MINIC-00  6889915 E-MINIC-00  6889917 E-MINIC-00  6889919 E-MINIC-00  6889928 E-MINIC-00  6889932 E-MINIC-00  6891122 E-MINIC-00  6891123 E-MINIC-00  6891124 E-MINIC-00  6891125 E-MINIC-00  6891126 E-MINIC-00  6891132 E-MINIC-00  6891134 E-MINIC-00  6892719 E-MINIC-00  6845246 E-MINIC-00  6824787 E-MINIC-00  6832793 E-MINIC-00  6832798 E-MINIC-00  6832799 E-MINIC-00  6838258 E-MINIC-00  6838261 E-MINIC-00  6870581 E-MINIC-00  6870566 E-MINIC-00  6870572 E-MINIC-00  6870573 E-MINIC-00  6824788 E-MINIC-00  6824793 E-MINIC-00  6824813 E-MINIC-00  6827892 E-MINIC-00  6827904 E-MINIC-00  6827907 E-MINIC-00  6827912 E-MINIC-00  6827913 E-MINIC-00  6827914 E-MINIC-00  6827927 E-MINIC-00  6870582 E-MINIC-00  6873158 E-MINIC-00  6845232 E-MINIC-00  6827916 E-MINIC-00  6827935 E-MINIC-00  6836982 E-MINIC-00  6836995 E-MINIC-00  6837001 E-MINIC-00  6838186 E-MINIC-00  6838189 E-MINIC-00  6838193 E-MINIC-00  6838201 E-MINIC-00  6838204 E-MINIC-00  6838244 E-MINIC-00  6838246 E-MINIC-00  6855716 E-MINIC-00  6855721 E-MINIC-00  6860759 E-MINIC-00  6860791 E-MINIC-00  6883975 E-MINIC-00  6883976 E-MINIC-00  6883977 E-MINIC-00  6883978 E-MINIC-00  6883979 E-MINIC-00  6883983 E-MINIC-00  6888697 E-MINIC-00  6888700 E-MINIC-00  6888707 E-MINIC-00  6892643 E-MINIC-00  6892660 E-MINIC-00  6892675 E-MINIC-00  6892683 E-MINIC-00  6892695 E-MINIC-00  6892702 E-MINIC-00  6892703 E-MINIC-00  6894971 E-MINIC-00  6894972 E-MINIC-00  6894983 E-MINIC-00  6896327 E-MINIC-00  6898702 E-MINIC-00  6898714 E-MINIC-00  6898721 E-MINIC-00  6898730 E-MINIC-00  6898731 E-MINIC-00  6860718 E-MINIC-00  6873178 E-MINIC-00  6873185 E-MINIC-00  6873193 E-MINIC-00  6876009 E-MINIC-00  6876010 E-MINIC-00  6876011 E-MINIC-00  6876015 E-MINIC-00  6876017 E-MINIC-00  6876018 E-MINIC-00  6876026 E-MINIC-00  6876037 E-MINIC-00  6876041 E-MINIC-00  6876042 E-MINIC-00  6876051 E-MINIC-00  6878075 E-MINIC-00  6878076 E-MINIC-00  6878077 E-MINIC-00  6878078 E-MINIC-00  6878079 E-MINIC-00  6878082 E-MINIC-00  6878085 E-MINIC-00  6878089 E-MINIC-00  6879405 E-MINIC-00  6879406 E-MINIC-00  6879408 E-MINIC-00  6879410 E-MINIC-00  6879411 E-MINIC-00  6879412 E-MINIC-00  6879413 E-MINIC-00  6879414 E-MINIC-00  6879416 E-MINIC-00  6879419 E-MINIC-00  6879426 E-MINIC-00  6879432 E-MINIC-00  6879433 E-MINIC-00  6879440 E-MINIC-00  6879442 E-MINIC-00  6879443 E-MINIC-00  6879444 E-MINIC-00  6879446 E-MINIC-00  6879447 E-MINIC-00  6879448 E-MINIC-00  6879449 E-MINIC-00  6879451 E-MINIC-00  6879452 E-MINIC-00  6879453 E-MINIC-00  6879454 E-MINIC-00  6883965 E-MINIC-00  6887179 E-MINIC-00  6887212 E-MINIC-00  6860806 E-MINIC-00  6860832 E-MINIC-00  6860833 E-MINIC-00  6867788 E-MINIC-00  6867790 E-MINIC-00  6822940 E-MINIC-00  6822942 E-MINIC-00  6822946 E-MINIC-00  6822948 E-MINIC-00  6827890 E-MINIC-00  6827921 E-MINIC-00  6827928 E-MINIC-00  6827929 E-MINIC-00  6827936 E-MINIC-00  6829315 E-MINIC-00  6829316 E-MINIC-00  6834925 E-MINIC-00  6834926 E-MINIC-00  6834929 E-MINIC-00  6834930 E-MINIC-00  6834931 E-MINIC-00  6836969 E-MINIC-00  6836971 E-MINIC-00  6836973 E-MINIC-00  6836974 E-MINIC-00  6836979 E-MINIC-00  6836981 E-MINIC-00  6836983 E-MINIC-00  6836984 E-MINIC-00  6836985 E-MINIC-00  6836986 E-MINIC-00  6836987 E-MINIC-00  6836988 E-MINIC-00  6836989 E-MINIC-00  6836990 E-MINIC-00  6836991 E-MINIC-00  6836992 E-MINIC-00  6836993 E-MINIC-00  6837000 E-MINIC-00  6837007 E-MINIC-00  6837009 E-MINIC-00  6837012 E-MINIC-00  6837016 E-MINIC-00  6837018 E-MINIC-00  6838185 E-MINIC-00  6838188 E-MINIC-00  6838199 E-MINIC-00  6838200 E-MINIC-00  6838202 E-MINIC-00  6838203 E-MINIC-00  6838205 E-MINIC-00  6838217 E-MINIC-00  6838219 E-MINIC-00  6838222 E-MINIC-00  6838263 E-MINIC-00  6842191 E-MINIC-00  6842193 E-MINIC-00  6842195 E-MINIC-00  6842203 E-MINIC-00  6842205 E-MINIC-00  6842207 E-MINIC-00  6842215 E-MINIC-00  6845201 E-MINIC-00  6845205 E-MINIC-00  6845208 E-MINIC-00  6845226 E-MINIC-00  6845237 E-MINIC-00  6845242 E-MINIC-00  6848907 E-MINIC-00  6848934 E-MINIC-00  6848940 E-MINIC-00  6848949 E-MINIC-00  6852616 E-MINIC-00  6852618 E-MINIC-00  6852624 E-MINIC-00  6852633 E-MINIC-00  6852665 E-MINIC-00  6852670 E-MINIC-00  6852689 E-MINIC-00  6852691 E-MINIC-00  6852698 E-MINIC-00  6852701 E-MINIC-00  6855626 E-MINIC-00  6855631 E-MINIC-00  6855668 E-MINIC-00  6855692 E-MINIC-00  6855700 E-MINIC-00  6860790 E-MINIC-00  6860795 E-MINIC-00  6860796 E-MINIC-00  6860805 E-MINIC-00  6860828 E-MINIC-00  6860837 E-MINIC-00  6867794 E-MINIC-00  6867827 E-MINIC-00  6867832 E-MINIC-00  6868301 E-MINIC-00  6868305 E-MINIC-00  6868306 E-MINIC-00  6868307 E-MINIC-00  6868310 E-MINIC-00  6868311 E-MINIC-00  6868313 E-MINIC-00  6868314 E-MINIC-00  6868315 E-MINIC-00  6868322 E-MINIC-00  6868324 E-MINIC-00  6868327 E-MINIC-00  6868328 E-MINIC-00  6868330 E-MINIC-00  6868331 E-MINIC-00  6868332 E-MINIC-00  6868333 E-MINIC-00  6868337 E-MINIC-00  6868340 E-MINIC-00  6868343 E-MINIC-00  6870550 E-MINIC-00  6870552 E-MINIC-00  6870553 E-MINIC-00  6870555 E-MINIC-00  6870557 E-MINIC-00  6870558 E-MINIC-00  6870562 E-MINIC-00  6870565 E-MINIC-00  6870576 E-MINIC-00  6870583 E-MINIC-00  6870584 E-MINIC-00  6870585 E-MINIC-00  6870586 E-MINIC-00  6870588 E-MINIC-00  6870591 E-MINIC-00  6870595 E-MINIC-00  6873122 E-MINIC-00  6873126 E-MINIC-00  6873127 E-MINIC-00  6873129 E-MINIC-00  6873134 E-MINIC-00  6873135 E-MINIC-00  6873136 E-MINIC-00  6873145 E-MINIC-00  6873148 E-MINIC-00  6873161 E-MINIC-00  6876008 E-MINIC-00  6876012 E-MINIC-00  6878066 E-MINIC-00  6878083 E-MINIC-00  6887224 E-MINIC-00  6888694 E-MINIC-00  6888709 E-MINIC-00  6888715 E-MINIC-00  6888723 E-MINIC-00  6888740 E-MINIC-00  6889940 E-MINIC-00  6891142 E-MINIC-00  6891149 E-MINIC-00  6891185 E-MINIC-00  6891187 E-MINIC-00  6892608 E-MINIC-00  6892620 E-MINIC-00  6892684 E-MINIC-00  6892696 E-MINIC-00  6894975 E-MINIC-00  6894978 E-MINIC-00  6894980 E-MINIC-00  6896318 E-MINIC-00  6896320 E-MINIC-00  6898701 E-MINIC-00  6898711 E-MINIC-00  6892609 E-MINIC-00  6852648 E-MINIC-00  6832786 E-MINIC-00  6827889 E-MINIC-00  6837015 E-MINIC-00  6848899 E-MINIC-00  6848901 E-MINIC-00  6848902 E-MINIC-00  6848911 E-MINIC-00  6848925 E-MINIC-00  6848930 E-MINIC-00  6855645 E-MINIC-00  6860714 E-MINIC-00  6860723 E-MINIC-00  6860728 E-MINIC-00  6883982 E-MINIC-00  6888729 E-MINIC-00  6888748 E-MINIC-00  6891159 E-MINIC-00  6891161 E-MINIC-00  6892604 E-MINIC-00  6892705 E-MINIC-00  6892708 E-MINIC-00  6892709 E-MINIC-00  6894973 E-MINIC-00  6837006 E-MINIC-00  6838191 E-MINIC-00  6838194 E-MINIC-00  6838216 E-MINIC-00  6838223 E-MINIC-00  6838227 E-MINIC-00  6838232 E-MINIC-00  6838242 E-MINIC-00  6818561 E-MINIC-00  6824817 E-MINIC-00  6827893 E-MINIC-00  6827920 E-MINIC-00  6827926 E-MINIC-00  6827930 E-MINIC-00  6829310 E-MINIC-00  6829312 E-MINIC-00  6829314 E-MINIC-00  6829320 E-MINIC-00  6832809 E-MINIC-00  6832810 E-MINIC-00  6832819 E-MINIC-00  6834928 E-MINIC-00  6838213 E-MINIC-00  6838215 E-MINIC-00  6838218 E-MINIC-00  6838226 E-MINIC-00  6838233 E-MINIC-00  6838234 E-MINIC-00  6838241 E-MINIC-00  6838249 E-MINIC-00  6842172 E-MINIC-00  6845210 E-MINIC-00  6845222 E-MINIC-00  6845233 E-MINIC-00  6848896 E-MINIC-00  6848917 E-MINIC-00  6848921 E-MINIC-00  6848938 E-MINIC-00  6848946 E-MINIC-00  6848950 E-MINIC-00  6852602 E-MINIC-00  6852607 E-MINIC-00  6852620 E-MINIC-00  6852626 E-MINIC-00  6852627 E-MINIC-00  6852629 E-MINIC-00  6852635 E-MINIC-00  6852642 E-MINIC-00  6852644 E-MINIC-00  6852646 E-MINIC-00  6852647 E-MINIC-00  6852652 E-MINIC-00  6852654 E-MINIC-00  6852658 E-MINIC-00  6852663 E-MINIC-00  6852686 E-MINIC-00  6855625 E-MINIC-00  6855635 E-MINIC-00  6855675 E-MINIC-00  6855694 E-MINIC-00  6855706 E-MINIC-00  6855708 E-MINIC-00  6855710 E-MINIC-00  6855711 E-MINIC-00  6855714 E-MINIC-00  6855731 E-MINIC-00  6860430 E-MINIC-00  6860431 E-MINIC-00  6860433 E-MINIC-00  6860434 E-MINIC-00  6860445 E-MINIC-00  6860447 E-MINIC-00  6860448 E-MINIC-00  6860449 E-MINIC-00  6860452 E-MINIC-00  6860662 E-MINIC-00  6860664 E-MINIC-00  6860669 E-MINIC-00  6860689 E-MINIC-00  6860719 E-MINIC-00  6860729 E-MINIC-00  6860740 E-MINIC-00  6860752 E-MINIC-00  6860753 E-MINIC-00  6860755 E-MINIC-00  6860798 E-MINIC-00  6867831 E-MINIC-00  6868308 E-MINIC-00  6868316 E-MINIC-00  6868317 E-MINIC-00  6868318 E-MINIC-00  6868329 E-MINIC-00  6873176 E-MINIC-00  6873192 E-MINIC-00  6876032 E-MINIC-00  6876043 E-MINIC-00  6876044 E-MINIC-00  6884002 E-MINIC-00  6888665 E-MINIC-00  6827919 E-MINIC-00  6848920 E-MINIC-00  6855726 E-MINIC-00  6860826 E-MINIC-00  6860851 E-MINIC-00  6860852 E-MINIC-00  6867791 E-MINIC-00  6867792 E-MINIC-00  6867796 E-MINIC-00  6867798 E-MINIC-00  6867809 E-MINIC-00  6867816 E-MINIC-00  6867818 E-MINIC-00  6867822 E-MINIC-00  6867828 E-MINIC-00  6868304 E-MINIC-00  6870551 E-MINIC-00  6888736 E-MINIC-00  6888766 E-MINIC-00  6889937 E-MINIC-00  6889961 E-MINIC-00  6891138 E-MINIC-00  6891147 E-MINIC-00  6891157 E-MINIC-00  6891167 E-MINIC-00  6891169 E-MINIC-00  6891177 E-MINIC-00  6891179 E-MINIC-00  6892613 E-MINIC-00  6892631 E-MINIC-00  6892657 E-MINIC-00  6892662 E-MINIC-00  6892664 E-MINIC-00  6892707 E-MINIC-00  6892712 E-MINIC-00  6892713 E-MINIC-00  6892716 E-MINIC-00  6896277 E-MINIC-00  6896306 E-MINIC-00  6896314 E-MINIC-00  6896326 E-MINIC-00  6896331 E-MINIC-00  6896288 E-MINIC-00  6896293 E-MINIC-00  6852672 E-MINIC-00  6891186 E-MINIC-00  6860781 E-MINIC-00  6860786 E-MINIC-00  6887184 E-MINIC-00  6892690 E-MINIC-00  6818574 E-MINIC-00  6822938 E-MINIC-00  6822939 E-MINIC-00  6822943 E-MINIC-00  6822944 E-MINIC-00  6822947 E-MINIC-00  6822951 E-MINIC-00  6822952 E-MINIC-00  6822953 E-MINIC-00  6822954 E-MINIC-00  6822955 E-MINIC-00  6822956 E-MINIC-00  6822957 E-MINIC-00  6822958 E-MINIC-00  6822959 E-MINIC-00  6822960 E-MINIC-00  6822962 E-MINIC-00  6822964 E-MINIC-00  6822965 E-MINIC-00  6822967 E-MINIC-00  6824777 E-MINIC-00  6824779 E-MINIC-00  6824820 E-MINIC-00  6832783 E-MINIC-00  6832784 E-MINIC-00  6832792 E-MINIC-00  6832800 E-MINIC-00  6832818 E-MINIC-00  6832828 E-MINIC-00  6832830 E-MINIC-00  6834897 E-MINIC-00  6834900 E-MINIC-00  6836994 E-MINIC-00  6837004 E-MINIC-00  6837008 E-MINIC-00  6837017 E-MINIC-00  6838192 E-MINIC-00  6838195 E-MINIC-00  6838206 E-MINIC-00  6838207 E-MINIC-00  6838208 E-MINIC-00  6838239 E-MINIC-00  6838253 E-MINIC-00  6838257 E-MINIC-00  6842188 E-MINIC-00  6842197 E-MINIC-00  6842202 E-MINIC-00  6842208 E-MINIC-00  6842209 E-MINIC-00  6842216 E-MINIC-00  6842223 E-MINIC-00  6842229 E-MINIC-00  6842232 E-MINIC-00  6842235 E-MINIC-00  6842237 E-MINIC-00  6842240 E-MINIC-00  6845203 E-MINIC-00  6845204 E-MINIC-00  6845212 E-MINIC-00  6845214 E-MINIC-00  6845215 E-MINIC-00  6845216 E-MINIC-00  6845217 E-MINIC-00  6845219 E-MINIC-00  6845220 E-MINIC-00  6845221 E-MINIC-00  6845223 E-MINIC-00  6845225 E-MINIC-00  6845229 E-MINIC-00  6845230 E-MINIC-00  6845234 E-MINIC-00  6845236 E-MINIC-00  6845238 E-MINIC-00  6845239 E-MINIC-00  6845240 E-MINIC-00  6845244 E-MINIC-00  6845249 E-MINIC-00  6845250 E-MINIC-00  6848900 E-MINIC-00  6848922 E-MINIC-00  6848926 E-MINIC-00  6848927 E-MINIC-00  6848928 E-MINIC-00  6848932 E-MINIC-00  6848933 E-MINIC-00  6848937 E-MINIC-00  6848945 E-MINIC-00  6852609 E-MINIC-00  6852614 E-MINIC-00  6852615 E-MINIC-00  6852623 E-MINIC-00  6852625 E-MINIC-00  6852630 E-MINIC-00  6852631 E-MINIC-00  6852634 E-MINIC-00  6852636 E-MINIC-00  6852637 E-MINIC-00  6852639 E-MINIC-00  6852640 E-MINIC-00  6852657 E-MINIC-00  6852673 E-MINIC-00  6852685 E-MINIC-00  6852690 E-MINIC-00  6852692 E-MINIC-00  6852693 E-MINIC-00  6852694 E-MINIC-00  6852695 E-MINIC-00  6852696 E-MINIC-00  6852697 E-MINIC-00  6852700 E-MINIC-00  6855624 E-MINIC-00  6855628 E-MINIC-00  6855632 E-MINIC-00  6855633 E-MINIC-00  6855634 E-MINIC-00  6855636 E-MINIC-00  6855637 E-MINIC-00  6855655 E-MINIC-00  6855658 E-MINIC-00  6855659 E-MINIC-00  6855663 E-MINIC-00  6855666 E-MINIC-00  6855682 E-MINIC-00  6855697 E-MINIC-00  6860660 E-MINIC-00  6860680 E-MINIC-00  6860682 E-MINIC-00  6860686 E-MINIC-00  6860690 E-MINIC-00  6860713 E-MINIC-00  6860748 E-MINIC-00  6860756 E-MINIC-00  6860758 E-MINIC-00  6860787 E-MINIC-00  6860803 E-MINIC-00  6860810 E-MINIC-00  6860811 E-MINIC-00  6860812 E-MINIC-00  6860817 E-MINIC-00  6860822 E-MINIC-00  6860834 E-MINIC-00  6860838 E-MINIC-00  6860843 E-MINIC-00  6860850 E-MINIC-00  6860853 E-MINIC-00  6860857 E-MINIC-00  6867795 E-MINIC-00  6867799 E-MINIC-00  6867800 E-MINIC-00  6867801 E-MINIC-00  6867802 E-MINIC-00  6867819 E-MINIC-00  6867821 E-MINIC-00  6867824 E-MINIC-00  6867830 E-MINIC-00  6868303 E-MINIC-00  6868335 E-MINIC-00  6868336 E-MINIC-00  6868338 E-MINIC-00  6868339 E-MINIC-00  6868341 E-MINIC-00  6870554 E-MINIC-00  6870563 E-MINIC-00  6870569 E-MINIC-00  6870570 E-MINIC-00  6870575 E-MINIC-00  6870589 E-MINIC-00  6870590 E-MINIC-00  6870592 E-MINIC-00  6873117 E-MINIC-00  6873138 E-MINIC-00  6873140 E-MINIC-00  6873147 E-MINIC-00  6873155 E-MINIC-00  6873187 E-MINIC-00  6876016 E-MINIC-00  6888666 E-MINIC-00  6888689 E-MINIC-00  6888714 E-MINIC-00  6888717 E-MINIC-00  6888718 E-MINIC-00  6888727 E-MINIC-00  6888730 E-MINIC-00  6888753 E-MINIC-00  6888760 E-MINIC-00  6888762 E-MINIC-00  6888769 E-MINIC-00  6889947 E-MINIC-00  6889949 E-MINIC-00  6889953 E-MINIC-00  6889964 E-MINIC-00  6889969 E-MINIC-00  6889972 E-MINIC-00  6889973 E-MINIC-00  6891039 E-MINIC-00  6891040 E-MINIC-00  6891042 E-MINIC-00  6891046 E-MINIC-00  6891050 E-MINIC-00  6891053 E-MINIC-00  6891058 E-MINIC-00  6891060 E-MINIC-00  6891061 E-MINIC-00  6891062 E-MINIC-00  6891065 E-MINIC-00  6891066 E-MINIC-00  6891068 E-MINIC-00  6891071 E-MINIC-00  6891072 E-MINIC-00  6891073 E-MINIC-00  6891074 E-MINIC-00  6891075 E-MINIC-00  6891076 E-MINIC-00  6891085 E-MINIC-00  6891104 E-MINIC-00  6891127 E-MINIC-00  6891133 E-MINIC-00  6891135 E-MINIC-00  6891140 E-MINIC-00  6891141 E-MINIC-00  6891164 E-MINIC-00  6892603 E-MINIC-00  6892605 E-MINIC-00  6892606 E-MINIC-00  6892610 E-MINIC-00  6892611 E-MINIC-00  6892612 E-MINIC-00  6892615 E-MINIC-00  6892617 E-MINIC-00  6892619 E-MINIC-00  6892622 E-MINIC-00  6892623 E-MINIC-00  6892624 E-MINIC-00  6892628 E-MINIC-00  6892629 E-MINIC-00  6892630 E-MINIC-00  6892634 E-MINIC-00  6892638 E-MINIC-00  6892641 E-MINIC-00  6892645 E-MINIC-00  6892648 E-MINIC-00  6892652 E-MINIC-00  6892658 E-MINIC-00  6892668 E-MINIC-00  6892687 E-MINIC-00  6892704 E-MINIC-00  6892710 E-MINIC-00  6892715 E-MINIC-00  6894974 E-MINIC-00  6894979 E-MINIC-00  6895029 E-MINIC-00  6896285 E-MINIC-00  6896300 E-MINIC-00  6896304 E-MINIC-00  6898692 E-MINIC-00  6898752 E-MINIC-00  6898760 E-MINIC-00  6898765 E-MINIC-00  6822973 E-MINIC-00  6822974 E-MINIC-00  6824780 E-MINIC-00  6824784 E-MINIC-00  6827908 E-MINIC-00  6827909 E-MINIC-00  6827925 E-MINIC-00  6827933 E-MINIC-00  6827934 E-MINIC-00  6829317 E-MINIC-00  6889941 E-MINIC-00  6891057 E-MINIC-00  6898745 E-MINIC-00  6822968 E-MINIC-00  6822969 E-MINIC-00  6824773 E-MINIC-00  6860692 E-MINIC-00  6860746 E-MINIC-00  6860807 E-MINIC-00  6860808 E-MINIC-00  6860835 E-MINIC-00  6860855 E-MINIC-00  6860856 E-MINIC-00  6867797 E-MINIC-00  6868323 E-MINIC-00  6883981 E-MINIC-00  6883985 E-MINIC-00  6883987 E-MINIC-00  6883989 E-MINIC-00  6883999 E-MINIC-00  6884021 E-MINIC-00  6888677 E-MINIC-00  6888685 E-MINIC-00  6888705 E-MINIC-00  6888710 E-MINIC-00  6889921 E-MINIC-00  6889929 E-MINIC-00  6895024 E-MINIC-00  6896263 E-MINIC-00  6898698 E-MINIC-00  6848923 E-MINIC-00  6873120 E-MINIC-00  6873124 E-MINIC-00  6873160 E-MINIC-00  6873163 E-MINIC-00  6889945 E-MINIC-00  6889946 E-MINIC-00  6889962 E-MINIC-00  6891077 E-MINIC-00  6891079 E-MINIC-00  6891097 E-MINIC-00  6891112 E-MINIC-00  6891163 E-MINIC-00  6891170 E-MINIC-00  6891172 E-MINIC-00  6891183 E-MINIC-00  6892632 E-MINIC-00  6892635 E-MINIC-00  6892671 E-MINIC-00  6892691 E-MINIC-00  6892711 E-MINIC-00  6892714 E-MINIC-00  6892717 E-MINIC-00  6892718 E-MINIC-00  6894984 E-MINIC-00  6896267 E-MINIC-00  6896324 E-MINIC-00  6898709 E-MINIC-00  6836996 E-MINIC-00  6888758 E-MINIC-00  6848919 E-MINIC-00  6898694 E-MINIC-00  6898755 E-MINIC-00  6898756 E-MINIC-00  6829319 E-MINIC-00  6832814 E-MINIC-00  6832820 E-MINIC-00  6832829 E-MINIC-00  6832832 E-MINIC-00  6842189 E-MINIC-00  6852606 E-MINIC-00  6852684 E-MINIC-00  6855638 E-MINIC-00  6855640 E-MINIC-00  6855641 E-MINIC-00  6855642 E-MINIC-00  6855644 E-MINIC-00  6855646 E-MINIC-00  6855648 E-MINIC-00  6855652 E-MINIC-00  6855654 E-MINIC-00  6855727 E-MINIC-00  6855730 E-MINIC-00  6860741 E-MINIC-00  6860789 E-MINIC-00  6860792 E-MINIC-00  6860797 E-MINIC-00  6860802 E-MINIC-00  6860809 E-MINIC-00  6860824 E-MINIC-00  6860842 E-MINIC-00  6860846 E-MINIC-00  6860848 E-MINIC-00  6860854 E-MINIC-00  6867803 E-MINIC-00  6867804 E-MINIC-00  6867805 E-MINIC-00  6867806 E-MINIC-00  6867808 E-MINIC-00  6867810 E-MINIC-00  6867812 E-MINIC-00  6867813 E-MINIC-00  6867815 E-MINIC-00  6867817 E-MINIC-00  6867820 E-MINIC-00  6867837 E-MINIC-00  6883993 E-MINIC-00  6883996 E-MINIC-00  6887191 E-MINIC-00  6887193 E-MINIC-00  6887200 E-MINIC-00  6",
      "more_code_info": "887202 E-MINIC-00  6887218 E-MINIC-00  6888656 E-MINIC-00  6888682 E-MINIC-00  6888691 E-MINIC-00  6896257 E-MINIC-00  6896258 E-MINIC-00  6884009 E-MINIC-00  6884012 E-MINIC-00  6884018 E-MINIC-00  6894985 E-MINIC-00  6822941 E-MINIC-00-EN  6822950 E-MINIC-00-EN  6822961 E-MINIC-00-EN  6822963 E-MINIC-00-EN  6822966 E-MINIC-00-EN  6822970 E-MINIC-00-EN  6822972 E-MINIC-00-EN  6824775 E-MINIC-00-EN  6824776 E-MINIC-00-EN  6824778 E-MINIC-00-EN  6824781 E-MINIC-00-EN  6824785 E-MINIC-00-EN  6824786 E-MINIC-00-EN  6824789 E-MINIC-00-EN  6824791 E-MINIC-00-EN  6824794 E-MINIC-00-EN  6824795 E-MINIC-00-EN  6824796 E-MINIC-00-EN  6824798 E-MINIC-00-EN  6824801 E-MINIC-00-EN  6824803 E-MINIC-00-EN  6824805 E-MINIC-00-EN  6824806 E-MINIC-00-EN  6824809 E-MINIC-00-EN  6824810 E-MINIC-00-EN  6824819 E-MINIC-00-EN  6824822 E-MINIC-00-EN  6827894 E-MINIC-00-EN  6827895 E-MINIC-00-EN  6827898 E-MINIC-00-EN  6827899 E-MINIC-00-EN  6827900 E-MINIC-00-EN  6827901 E-MINIC-00-EN  6827902 E-MINIC-00-EN  6827903 E-MINIC-00-EN  6827905 E-MINIC-00-EN  6827906 E-MINIC-00-EN  6827910 E-MINIC-00-EN  6827911 E-MINIC-00-EN  6827915 E-MINIC-00-EN  6827917 E-MINIC-00-EN  6827923 E-MINIC-00-EN  6834875 E-MINIC-00-EN  6834876 E-MINIC-00-EN  6834878 E-MINIC-00-EN  6834879 E-MINIC-00-EN  6834881 E-MINIC-00-EN  6834882 E-MINIC-00-EN  6834905 E-MINIC-00-EN  6834906 E-MINIC-00-EN  6834908 E-MINIC-00-EN  6834912 E-MINIC-00-EN  6838220 E-MINIC-00-EN  6838221 E-MINIC-00-EN  6838225 E-MINIC-00-EN  6838229 E-MINIC-00-EN  6838230 E-MINIC-00-EN  6838248 E-MINIC-00-EN  6838250 E-MINIC-00-EN  6838260 E-MINIC-00-EN  6838264 E-MINIC-00-EN  6842166 E-MINIC-00-EN  6842168 E-MINIC-00-EN  6842169 E-MINIC-00-EN  6842170 E-MINIC-00-EN  6842171 E-MINIC-00-EN  6842173 E-MINIC-00-EN  6842178 E-MINIC-00-EN  6842179 E-MINIC-00-EN  6842184 E-MINIC-00-EN  6842185 E-MINIC-00-EN  6842186 E-MINIC-00-EN  6842187 E-MINIC-00-EN  6842194 E-MINIC-00-EN  6842199 E-MINIC-00-EN  6842200 E-MINIC-00-EN  6842201 E-MINIC-00-EN  6842210 E-MINIC-00-EN  6842211 E-MINIC-00-EN  6842213 E-MINIC-00-EN  6842214 E-MINIC-00-EN  6845202 E-MINIC-00-EN  6845209 E-MINIC-00-EN  6845211 E-MINIC-00-EN  6845213 E-MINIC-00-EN  6845241 E-MINIC-00-EN  6845243 E-MINIC-00-EN  6845245 E-MINIC-00-EN  6845247 E-MINIC-00-EN  6848872 E-MINIC-00-EN  6848873 E-MINIC-00-EN  6848874 E-MINIC-00-EN  6848875 E-MINIC-00-EN  6848876 E-MINIC-00-EN  6848879 E-MINIC-00-EN  6848882 E-MINIC-00-EN  6848893 E-MINIC-00-EN  6848894 E-MINIC-00-EN  6848903 E-MINIC-00-EN  6848904 E-MINIC-00-EN  6848906 E-MINIC-00-EN  6848910 E-MINIC-00-EN  6848918 E-MINIC-00-EN  6848929 E-MINIC-00-EN  6848935 E-MINIC-00-EN  6848939 E-MINIC-00-EN  6848941 E-MINIC-00-EN  6848943 E-MINIC-00-EN  6848944 E-MINIC-00-EN  6848947 E-MINIC-00-EN  6848948 E-MINIC-00-EN  6848951 E-MINIC-00-EN  6852603 E-MINIC-00-EN  6852604 E-MINIC-00-EN  6852605 E-MINIC-00-EN  6852608 E-MINIC-00-EN  6852610 E-MINIC-00-EN  6852628 E-MINIC-00-EN  6852632 E-MINIC-00-EN  6852638 E-MINIC-00-EN  6852643 E-MINIC-00-EN  6852645 E-MINIC-00-EN  6852649 E-MINIC-00-EN  6852659 E-MINIC-00-EN  6852671 E-MINIC-00-EN  6852683 E-MINIC-00-EN  6852687 E-MINIC-00-EN  6852688 E-MINIC-00-EN  6852699 E-MINIC-00-EN  6855650 E-MINIC-00-EN  6855665 E-MINIC-00-EN  6855667 E-MINIC-00-EN  6855669 E-MINIC-00-EN  6855670 E-MINIC-00-EN  6855672 E-MINIC-00-EN  6855676 E-MINIC-00-EN  6855679 E-MINIC-00-EN  6855684 E-MINIC-00-EN  6855689 E-MINIC-00-EN  6855690 E-MINIC-00-EN  6855693 E-MINIC-00-EN  6855696 E-MINIC-00-EN  6855698 E-MINIC-00-EN  6855703 E-MINIC-00-EN  6855707 E-MINIC-00-EN  6855723 E-MINIC-00-EN  6855724 E-MINIC-00-EN  6855733 E-MINIC-00-EN  6860429 E-MINIC-00-EN  6860432 E-MINIC-00-EN  6860435 E-MINIC-00-EN  6860436 E-MINIC-00-EN  6860437 E-MINIC-00-EN  6860440 E-MINIC-00-EN  6860441 E-MINIC-00-EN  6860442 E-MINIC-00-EN  6860443 E-MINIC-00-EN  6860446 E-MINIC-00-EN  6860451 E-MINIC-00-EN  6860453 E-MINIC-00-EN  6860663 E-MINIC-00-EN  6860666 E-MINIC-00-EN  6860668 E-MINIC-00-EN  6860671 E-MINIC-00-EN  6860673 E-MINIC-00-EN  6860674 E-MINIC-00-EN  6860675 E-MINIC-00-EN  6860676 E-MINIC-00-EN  6860677 E-MINIC-00-EN  6860678 E-MINIC-00-EN  6860679 E-MINIC-00-EN  6860681 E-MINIC-00-EN  6860687 E-MINIC-00-EN  6860688 E-MINIC-00-EN  6860691 E-MINIC-00-EN  6860693 E-MINIC-00-EN  6860694 E-MINIC-00-EN  6860695 E-MINIC-00-EN  6860696 E-MINIC-00-EN  6860697 E-MINIC-00-EN  6860699 E-MINIC-00-EN  6860701 E-MINIC-00-EN  6860702 E-MINIC-00-EN  6860703 E-MINIC-00-EN  6860704 E-MINIC-00-EN  6860705 E-MINIC-00-EN  6860706 E-MINIC-00-EN  6860707 E-MINIC-00-EN  6860708 E-MINIC-00-EN  6860709 E-MINIC-00-EN  6860710 E-MINIC-00-EN  6860711 E-MINIC-00-EN  6860712 E-MINIC-00-EN  6860715 E-MINIC-00-EN  6860716 E-MINIC-00-EN  6860717 E-MINIC-00-EN  6860721 E-MINIC-00-EN  6860722 E-MINIC-00-EN  6860724 E-MINIC-00-EN  6860725 E-MINIC-00-EN  6860726 E-MINIC-00-EN  6860727 E-MINIC-00-EN  6860730 E-MINIC-00-EN  6860732 E-MINIC-00-EN  6860733 E-MINIC-00-EN  6860734 E-MINIC-00-EN  6860735 E-MINIC-00-EN  6860736 E-MINIC-00-EN  6860737 E-MINIC-00-EN  6860738 E-MINIC-00-EN  6860739 E-MINIC-00-EN  6860742 E-MINIC-00-EN  6860743 E-MINIC-00-EN  6860744 E-MINIC-00-EN  6860745 E-MINIC-00-EN  6860747 E-MINIC-00-EN  6860749 E-MINIC-00-EN  6860751 E-MINIC-00-EN  6860754 E-MINIC-00-EN  6860757 E-MINIC-00-EN  6860760 E-MINIC-00-EN  6860762 E-MINIC-00-EN  6860763 E-MINIC-00-EN  6860764 E-MINIC-00-EN  6860765 E-MINIC-00-EN  6860766 E-MINIC-00-EN  6860767 E-MINIC-00-EN  6860769 E-MINIC-00-EN  6860770 E-MINIC-00-EN  6860771 E-MINIC-00-EN  6860774 E-MINIC-00-EN  6860775 E-MINIC-00-EN  6860776 E-MINIC-00-EN  6860779 E-MINIC-00-EN  6860782 E-MINIC-00-EN  6860783 E-MINIC-00-EN  6860784 E-MINIC-00-EN  6860785 E-MINIC-00-EN  6860788 E-MINIC-00-EN  6860793 E-MINIC-00-EN  6860794 E-MINIC-00-EN  6860800 E-MINIC-00-EN  6860801 E-MINIC-00-EN  6860804 E-MINIC-00-EN  6860813 E-MINIC-00-EN  6860814 E-MINIC-00-EN  6860815 E-MINIC-00-EN  6860816 E-MINIC-00-EN  6860818 E-MINIC-00-EN  6860819 E-MINIC-00-EN  6860820 E-MINIC-00-EN  6860829 E-MINIC-00-EN  6860831 E-MINIC-00-EN  6860836 E-MINIC-00-EN  6860839 E-MINIC-00-EN  6860840 E-MINIC-00-EN  6860841 E-MINIC-00-EN  6860844 E-MINIC-00-EN  6860845 E-MINIC-00-EN  6860859 E-MINIC-00-EN  6867793 E-MINIC-00-EN  6867807 E-MINIC-00-EN  6867811 E-MINIC-00-EN  6867814 E-MINIC-00-EN  6867823 E-MINIC-00-EN  6867825 E-MINIC-00-EN  6867826 E-MINIC-00-EN  6867829 E-MINIC-00-EN  6867833 E-MINIC-00-EN  6867834 E-MINIC-00-EN  6867835 E-MINIC-00-EN  6867836 E-MINIC-00-EN  6868295 E-MINIC-00-EN  6868296 E-MINIC-00-EN  6868297 E-MINIC-00-EN  6868298 E-MINIC-00-EN  6868299 E-MINIC-00-EN  6868300 E-MINIC-00-EN  6868302 E-MINIC-00-EN  6868309 E-MINIC-00-EN  6868312 E-MINIC-00-EN  6868319 E-MINIC-00-EN  6868325 E-MINIC-00-EN  6868326 E-MINIC-00-EN  6868334 E-MINIC-00-EN  6868342 E-MINIC-00-EN  6870559 E-MINIC-00-EN  6870560 E-MINIC-00-EN  6870561 E-MINIC-00-EN  6870567 E-MINIC-00-EN  6870571 E-MINIC-00-EN  6870577 E-MINIC-00-EN  6870578 E-MINIC-00-EN  6870579 E-MINIC-00-EN  6870580 E-MINIC-00-EN  6870596 E-MINIC-00-EN  6870597 E-MINIC-00-EN  6870598 E-MINIC-00-EN  6873118 E-MINIC-00-EN  6873119 E-MINIC-00-EN  6873132 E-MINIC-00-EN  6873133 E-MINIC-00-EN  6873137 E-MINIC-00-EN  6873142 E-MINIC-00-EN  6873144 E-MINIC-00-EN  6873146 E-MINIC-00-EN  6873149 E-MINIC-00-EN  6873150 E-MINIC-00-EN  6873151 E-MINIC-00-EN  6873152 E-MINIC-00-EN  6873153 E-MINIC-00-EN  6873157 E-MINIC-00-EN  6873159 E-MINIC-00-EN  6873162 E-MINIC-00-EN  6873164 E-MINIC-00-EN  6873166 E-MINIC-00-EN  6873167 E-MINIC-00-EN  6873168 E-MINIC-00-EN  6873169 E-MINIC-00-EN  6873170 E-MINIC-00-EN  6873171 E-MINIC-00-EN  6873172 E-MINIC-00-EN  6873173 E-MINIC-00-EN  6873174 E-MINIC-00-EN  6873175 E-MINIC-00-EN  6873177 E-MINIC-00-EN  6873179 E-MINIC-00-EN  6873180 E-MINIC-00-EN  6873181 E-MINIC-00-EN  6873183 E-MINIC-00-EN  6873184 E-MINIC-00-EN  6873186 E-MINIC-00-EN  6873188 E-MINIC-00-EN  6873189 E-MINIC-00-EN  6873191 E-MINIC-00-EN  6873194 E-MINIC-00-EN  6873195 E-MINIC-00-EN  6873196 E-MINIC-00-EN  6876005 E-MINIC-00-EN  6876006 E-MINIC-00-EN  6876013 E-MINIC-00-EN  6876014 E-MINIC-00-EN  6876019 E-MINIC-00-EN  6876020 E-MINIC-00-EN  6876021 E-MINIC-00-EN  6876022 E-MINIC-00-EN  6876023 E-MINIC-00-EN  6876024 E-MINIC-00-EN  6876025 E-MINIC-00-EN  6876027 E-MINIC-00-EN  6876028 E-MINIC-00-EN  6876029 E-MINIC-00-EN  6876030 E-MINIC-00-EN  6876031 E-MINIC-00-EN  6876033 E-MINIC-00-EN  6876034 E-MINIC-00-EN  6876035 E-MINIC-00-EN  6876036 E-MINIC-00-EN  6876038 E-MINIC-00-EN  6876039 E-MINIC-00-EN  6876040 E-MINIC-00-EN  6876045 E-MINIC-00-EN  6876046 E-MINIC-00-EN  6876047 E-MINIC-00-EN  6876049 E-MINIC-00-EN  6876053 E-MINIC-00-EN  6878047 E-MINIC-00-EN  6878049 E-MINIC-00-EN  6878051 E-MINIC-00-EN  6878052 E-MINIC-00-EN  6878053 E-MINIC-00-EN  6878055 E-MINIC-00-EN  6878057 E-MINIC-00-EN  6878058 E-MINIC-00-EN  6878059 E-MINIC-00-EN  6878060 E-MINIC-00-EN  6878061 E-MINIC-00-EN  6878062 E-MINIC-00-EN  6878063 E-MINIC-00-EN  6878064 E-MINIC-00-EN  6878065 E-MINIC-00-EN  6878067 E-MINIC-00-EN  6878068 E-MINIC-00-EN  6878069 E-MINIC-00-EN  6878070 E-MINIC-00-EN  6878071 E-MINIC-00-EN  6878072 E-MINIC-00-EN  6878073 E-MINIC-00-EN  6878074 E-MINIC-00-EN  6878080 E-MINIC-00-EN  6878081 E-MINIC-00-EN  6878084 E-MINIC-00-EN  6878086 E-MINIC-00-EN  6878087 E-MINIC-00-EN  6878088 E-MINIC-00-EN  6878090 E-MINIC-00-EN  6878091 E-MINIC-00-EN  6878092 E-MINIC-00-EN  6878093 E-MINIC-00-EN  6878094 E-MINIC-00-EN  6878095 E-MINIC-00-EN  6878096 E-MINIC-00-EN  6879407 E-MINIC-00-EN  6879409 E-MINIC-00-EN  6879415 E-MINIC-00-EN  6879417 E-MINIC-00-EN  6879418 E-MINIC-00-EN  6879420 E-MINIC-00-EN  6879421 E-MINIC-00-EN  6879422 E-MINIC-00-EN  6879424 E-MINIC-00-EN  6879425 E-MINIC-00-EN  6879427 E-MINIC-00-EN  6879428 E-MINIC-00-EN  6879429 E-MINIC-00-EN  6879430 E-MINIC-00-EN  6879431 E-MINIC-00-EN  6879434 E-MINIC-00-EN  6879435 E-MINIC-00-EN  6879436 E-MINIC-00-EN  6879437 E-MINIC-00-EN  6879438 E-MINIC-00-EN  6879439 E-MINIC-00-EN  6879441 E-MINIC-00-EN  6879445 E-MINIC-00-EN  6879450 E-MINIC-00-EN  6883966 E-MINIC-00-EN  6883968 E-MINIC-00-EN  6883969 E-MINIC-00-EN  6883970 E-MINIC-00-EN  6883971 E-MINIC-00-EN  6883973 E-MINIC-00-EN  6883974 E-MINIC-00-EN  6884024 E-MINIC-00-EN  6887180 E-MINIC-00-EN  6887181 E-MINIC-00-EN  6887196 E-MINIC-00-EN  6887207 E-MINIC-00-EN  6887208 E-MINIC-00-EN  6887214 E-MINIC-00-EN  6887220 E-MINIC-00-EN  6887225 E-MINIC-00-EN  6887227 E-MINIC-00-EN  6888653 E-MINIC-00-EN  6888654 E-MINIC-00-EN  6888657 E-MINIC-00-EN  6888658 E-MINIC-00-EN  6888659 E-MINIC-00-EN  6888660 E-MINIC-00-EN  6888662 E-MINIC-00-EN  6888663 E-MINIC-00-EN  6888664 E-MINIC-00-EN  6888667 E-MINIC-00-EN  6888668 E-MINIC-00-EN  6888669 E-MINIC-00-EN  6888671 E-MINIC-00-EN  6888673 E-MINIC-00-EN  6888674 E-MINIC-00-EN  6888675 E-MINIC-00-EN  6888676 E-MINIC-00-EN  6888679 E-MINIC-00-EN  6888680 E-MINIC-00-EN  6888690 E-MINIC-00-EN  6888701 E-MINIC-00-EN  6888706 E-MINIC-00-EN  6888708 E-MINIC-00-EN  6888720 E-MINIC-00-EN  6888722 E-MINIC-00-EN  6888728 E-MINIC-00-EN  6888731 E-MINIC-00-EN  6888733 E-MINIC-00-EN  6888734 E-MINIC-00-EN  6888735 E-MINIC-00-EN  6888739 E-MINIC-00-EN  6888741 E-MINIC-00-EN  6888744 E-MINIC-00-EN  6888745 E-MINIC-00-EN  6888746 E-MINIC-00-EN  6888747 E-MINIC-00-EN  6888751 E-MINIC-00-EN  6888752 E-MINIC-00-EN  6888756 E-MINIC-00-EN  6888759 E-MINIC-00-EN  6888761 E-MINIC-00-EN  6888763 E-MINIC-00-EN  6888768 E-MINIC-00-EN  6889914 E-MINIC-00-EN  6889922 E-MINIC-00-EN  6889923 E-MINIC-00-EN  6889924 E-MINIC-00-EN  6889925 E-MINIC-00-EN  6889930 E-MINIC-00-EN  6889931 E-MINIC-00-EN  6889938 E-MINIC-00-EN  6889944 E-MINIC-00-EN  6889951 E-MINIC-00-EN  6889960 E-MINIC-00-EN  6889968 E-MINIC-00-EN  6889970 E-MINIC-00-EN  6891038 E-MINIC-00-EN  6891043 E-MINIC-00-EN  6891044 E-MINIC-00-EN  6891047 E-MINIC-00-EN  6891048 E-MINIC-00-EN  6891049 E-MINIC-00-EN  6891051 E-MINIC-00-EN  6891052 E-MINIC-00-EN  6891056 E-MINIC-00-EN  6891087 E-MINIC-00-EN  6891089 E-MINIC-00-EN  6891098 E-MINIC-00-EN  6891107 E-MINIC-00-EN  6891109 E-MINIC-00-EN  6891118 E-MINIC-00-EN  6891129 E-MINIC-00-EN  6891136 E-MINIC-00-EN  6891145 E-MINIC-00-EN  6891158 E-MINIC-00-EN  6891160 E-MINIC-00-EN  6891165 E-MINIC-00-EN  6891171 E-MINIC-00-EN  6892636 E-MINIC-00-EN  6892637 E-MINIC-00-EN  6892640 E-MINIC-00-EN  6892642 E-MINIC-00-EN  6892644 E-MINIC-00-EN  6892647 E-MINIC-00-EN  6892650 E-MINIC-00-EN  6892653 E-MINIC-00-EN  6892661 E-MINIC-00-EN  6892666 E-MINIC-00-EN  6892667 E-MINIC-00-EN  6892669 E-MINIC-00-EN  6892670 E-MINIC-00-EN  6892673 E-MINIC-00-EN  6892674 E-MINIC-00-EN  6892677 E-MINIC-00-EN  6892685 E-MINIC-00-EN  6892686 E-MINIC-00-EN  6892688 E-MINIC-00-EN  6892706 E-MINIC-00-EN  6894977 E-MINIC-00-EN  6894981 E-MINIC-00-EN  6894982 E-MINIC-00-EN  6894987 E-MINIC-00-EN  6894988 E-MINIC-00-EN  6894989 E-MINIC-00-EN  6894990 E-MINIC-00-EN  6894992 E-MINIC-00-EN  6894993 E-MINIC-00-EN  6894994 E-MINIC-00-EN  6894995 E-MINIC-00-EN  6894996 E-MINIC-00-EN  6894997 E-MINIC-00-EN  6894999 E-MINIC-00-EN  6895000 E-MINIC-00-EN  6895002 E-MINIC-00-EN  6895003 E-MINIC-00-EN  6895004 E-MINIC-00-EN  6895006 E-MINIC-00-EN  6895007 E-MINIC-00-EN  6895008 E-MINIC-00-EN  6895009 E-MINIC-00-EN  6895010 E-MINIC-00-EN  6895011 E-MINIC-00-EN  6895012 E-MINIC-00-EN  6895013 E-MINIC-00-EN  6895015 E-MINIC-00-EN  6895017 E-MINIC-00-EN  6895018 E-MINIC-00-EN  6895019 E-MINIC-00-EN  6895020 E-MINIC-00-EN  6895026 E-MINIC-00-EN  6895027 E-MINIC-00-EN  6895028 E-MINIC-00-EN  6895030 E-MINIC-00-EN  6896259 E-MINIC-00-EN  6896260 E-MINIC-00-EN  6896265 E-MINIC-00-EN  6896284 E-MINIC-00-EN  6896286 E-MINIC-00-EN  6896289 E-MINIC-00-EN  6896295 E-MINIC-00-EN  6896299 E-MINIC-00-EN  6896302 E-MINIC-00-EN  6896319 E-MINIC-00-EN  6896321 E-MINIC-00-EN  6896325 E-MINIC-00-EN  6896328 E-MINIC-00-EN  6896329 E-MINIC-00-EN  6896330 E-MINIC-00-EN  6898736 E-MINIC-00-EN  6898740 E-MINIC-00-EN  6898743 E-MINIC-00-EN  6898744 E-MINIC-00-EN  6898749 E-MINIC-00-EN  6898750 E-MINIC-00-EN  6898751 E-MINIC-00-EN  6898762 E-MINIC-00-EN  6898763 E-MINIC-00-EN  6898768 E-MINIC-00-EN  6898770 E-MINIC-00-EN  6898771 E-MINIC-00-EN  6898774 E-MINIC-00-EN  6838231 E-MINIC-00-EN  6883972 E-MINIC-00-EN  6827918 E-MINIC-00-EN  6860700 E-MINIC-00-EN  6888721 E-MINIC-00-EN  6891083 E-MINIC-00-EN  6891093 E-MINIC-00-EN  6891096 E-MINIC-00-EN  6891099 E-MINIC-00-EN  6891100 E-MINIC-00-EN  6848880 E-MINIC-00-EN  6848892 E-MINIC-00-EN  6848931 E-MINIC-00-EN  6887192 E-MINIC-00-EN  6888725 E-MINIC-00-EN  6888767 E-MINIC-00-EN  6889918 E-MINIC-00-EN  6889926 E-MINIC-00-EN  6889927 E-MINIC-00-EN  6889933 E-MINIC-00-EN  6889934 E-MINIC-00-EN  6838255 E-MINIC-00-EN  6842165 E-MINIC-00-EN  6842174 E-MINIC-00-EN  6842176 E-MINIC-00-EN  6842183 E-MINIC-00-EN  6887223 E-MINIC-00-EN  6888692 E-MINIC-00-EN  6891095 E-MINIC-00-EN  6891108 E-MINIC-00-EN  6892649 E-MINIC-00-EN  6889952 E-MINIC-00-EN  6891045 E-MINIC-00-EN  6860665 E-MINIC-00-EN  6867681 M1013204  6875967 M1013204  6875971 M1013204  6875996 M1013204  6873208 M1013204  6875993 M1013204  6875998 M1013204  6884034 M1013204  6884044 M1013204  6884047 M1013204  6884048 M1013204  6884049 M1013204  6885857 M1013204  6885868 M1013204  6885869 M1013204  6885871 M1013204  6885872 M1013204  6885873 M1013204  6885874 M1013204  6885881 M1013204  6873201 M1013204  6873204 M1013204  6875997 M1013204  6827939 M1013204  6827946 M1013204  6827966 M1013204  6875988 M1013204  6870545 M1013204  6875942 M1013204  6875985 M1013204  6878240 M1013204  6873203 M1013204  6873216 M1013204  6873207 M1013204  6873214 M1013204  6867682 M1013204  6873198 M1013204  6873199 M1013204  6875939 M1013204  6875941 M1013204  6875945 M1013204  6875984 M1013204  6875987 M1013204  6845272 M1013204  6845278 M1013204  6845283 M1013204  6845286 M1013204  6870540 M1013204  6870542 M1013204  6873197 M1013204  6873215 M1013204  6875906 M1013204  6875932 M1013204  6875940 M1013204  6875974 M1013204  6867648 M1013204  6873205 M1013204  6875990 M1013204  6878239 M1013204  6884046 M1013204  6867654 M1013204  6873200 M1013204  6873212 M1013204  6875995 M1013204  6885855 M1013204  6829271 N-FC-00  6839565 N-FC-00  6839571 N-FC-00  6839575 N-FC-00  6839576 N-FC-00  6799192 N-FC-00  6829276 N-FC-00  6829284 N-FC-00  6829292 N-FC-00  6831245 N-FC-00  6831253 N-FC-00  6831254 N-FC-00  6831255 N-FC-00  6831256 N-FC-00  6831261 N-FC-00  6831262 N-FC-00  6831263 N-FC-00  6831264 N-FC-00  6831265 N-FC-00  6831268 N-FC-00  6831274 N-FC-00  6839572 N-FC-00  6839584 N-FC-00  6839592 N-FC-00  6839596 N-FC-00  6849858 N-FC-00  6849884 N-FC-00  6849886 N-FC-00  6849895 N-FC-00  6849897 N-FC-00  6849903 N-FC-00  6849904 N-FC-00  6849905 N-FC-00  6863479 N-FC-00  6863481 N-FC-00  6881148 N-FC-00  6881378 N-FC-00  6884073 N-FC-00  6889541 N-FC-00  6889543 N-FC-00  6881405 N-FC-00  6831252 N-FC-00  6863497 N-FC-00  6863506 N-FC-00  6881404 N-FC-00  6799201 N-FC-00  6829270 N-FC-00  6829278 N-FC-00  6829281 N-FC-00  6829282 N-FC-00  6829288 N-FC-00  6831240 N-FC-00  6831246 N-FC-00  6831275 N-FC-00  6837808 N-FC-00  6837810 N-FC-00  6839555 N-FC-00  6839558 N-FC-00  6839561 N-FC-00  6839573 N-FC-00  6839583 N-FC-00  6839593 N-FC-00  6839594 N-FC-00  6839595 N-FC-00  6849863 N-FC-00  6849896 N-FC-00  6849874 N-FC-00  6849879 N-FC-00  6849881 N-FC-00  6849883 N-FC-00  6849885 N-FC-00  6849889 N-FC-00  6849892 N-FC-00  6849894 N-FC-00  6863478 N-FC-00  6869027 N-FC-00  6869028 N-FC-00  6869030 N-FC-00  6869032 N-FC-00  6870920 N-FC-00  6870921 N-FC-00  6870922 N-FC-00  6870923 N-FC-00  6870925 N-FC-00  6870928 N-FC-00  6870930 N-FC-00  6870931 N-FC-00  6870932 N-FC-00  6870935 N-FC-00  6870937 N-FC-00  6870938 N-FC-00  6870939 N-FC-00  6870940 N-FC-00  6870941 N-FC-00  6870942 N-FC-00  6870943 N-FC-00  6870944 N-FC-00  6870945 N-FC-00  6870946 N-FC-00  6881384 N-FC-00  6881388 N-FC-00  6881393 N-FC-00  6881395 N-FC-00  6881396 N-FC-00  6881398 N-FC-00  6881403 N-FC-00  6884066 N-FC-00  6884067 N-FC-00  6884075 N-FC-00  6884083 N-FC-00  6885565 N-FC-00  6885566 N-FC-00  6885568 N-FC-00  6885569 N-FC-00  6885570 N-FC-00  6885571 N-FC-00  6885573 N-FC-00  6885574 N-FC-00  6885982 N-FC-00  6889544 N-FC-00  6889557 N-FC-00  6889559 N-FC-00  6889563 N-FC-00  6849893 N-FC-00  6829274 N-FC-00  6831237 N-FC-00  6885572 N-FC-00  6839568 N-FC-00  6839574 N-FC-00  6839597 N-FC-00  6839611 N-FC-00  6870958 N-FC-00  6799191 N-FC-00  6799193 N-FC-00  6799194 N-FC-00  6829286 N-FC-00  6829290 N-FC-00  6831266 N-FC-00  6831267 N-FC-00  6839556 N-FC-00  6839598 N-FC-00  6839599 N-FC-00  6839600 N-FC-00  6839601 N-FC-00  6849859 N-FC-00  6849873 N-FC-00  6849878 N-FC-00  6849880 N-FC-00  6849887 N-FC-00  6863496 N-FC-00  6870947 N-FC-00  6870948 N-FC-00  6870949 N-FC-00  6870950 N-FC-00  6870951 N-FC-00  6870952 N-FC-00  6879585 N-FC-00  6879586 N-FC-00  6879587 N-FC-00  6879588 N-FC-00  6881387 N-FC-00  6881399 N-FC-00  6881400 N-FC-00  6881401 N-FC-00  6885975 N-FC-00  6885990 N-FC-00  6829265 N-FC-00  6839603 N-FC-00  6839604 N-FC-00  6839605 N-FC-00  6839607 N-FC-00  6839608 N-FC-00  6839609 N-FC-00  6839610 N-FC-00  6839612 N-FC-00  6849856 N-FC-00  6849857 N-FC-00  6863505 N-FC-00  6831286 N-FC-00  6829268 N-FC-00  6831250 N-FC-00  6839557 N-FC-00  6839559 N-FC-00  6839570 N-FC-00  6839585 N-FC-00  6839586 N-FC-00  6839590 N-FC-00  6849901 N-FC-00  6863494 N-FC-00  6863500 N-FC-00  6863507 N-FC-00  6881380 N-FC-00  6831238 N-FC-00  6831251 N-FC-00  6831257 N-FC-00  6849867 N-FC-00  6839577 N-FC-00  6839578 N-FC-00  6839579 N-FC-00  6839580 N-FC-00  6839581 N-FC-00  6839582 N-FC-00  6839602 N-FC-00  6849861 N-FC-00  6849864 N-FC-00  6849865 N-FC-00  6849868 N-FC-00  6849869 N-FC-00  6849870 N-FC-00  6849871 N-FC-00  6849877 N-FC-00  6849888 N-FC-00  6849890 N-FC-00  6881147 N-FC-00  6881383 N-FC-00  6881406 N-FC-00  6884069 N-FC-00  6884079 N-FC-00  6884080 N-FC-00  6885973 N-FC-00  6885991 N-FC-00  6849898 N-FC-00  6849900 N-FC-00  6849902 N-FC-00  6863487 N-FC-00  6863489 N-FC-00  6863490 N-FC-00  6863492 N-FC-00  6863493 N-FC-00  6863495 N-FC-00  6863498 N-FC-00  6863501 N-FC-00  6863502 N-FC-00  6863503 N-FC-00  6881397 N-FC-00  6885567 N-FC-00  6884068 N-FC-00  6885989 N-FC-00  6889536 N-FC-00  6849875 N-FC-00  6849876 N-FC-00  6831272 N-FC-00  6829277 N-FC-00  6829280 N-FC-00  6829283 N-FC-00  6831241 N-FC-00  6831242 N-FC-00  6831243 N-FC-00  6831247 N-FC-00  6831248 N-FC-00  6831249 N-FC-00  6831277 N-FC-00  6831279 N-FC-00  6831280 N-FC-00  6831284 N-FC-00  6837809 N-FC-00  6837811 N-FC-00  6837812 N-FC-00  6837814 N-FC-00  6839553 N-FC-00  6839554 N-FC-00  6839560 N-FC-00  6839562 N-FC-00  6839563 N-FC-00  6839564 N-FC-00  6839566 N-FC-00  6839569 N-FC-00  6839587 N-FC-00  6839588 N-FC-00  6839589 N-FC-00  6849872 N-FC-00  6849891 N-FC-00  6849899 N-FC-00  6863482 N-FC-00  6863483 N-FC-00  6863484 N-FC-00  6863485 N-FC-00  6863486 N-FC-00  6863488 N-FC-00  6863491 N-FC-00  6863499 N-FC-00  6863504 N-FC-00  6869025 N-FC-00  6869026 N-FC-00  6869029 N-FC-00  6869031 N-FC-00  6869033 N-FC-00  6869034 N-FC-00  6870924 N-FC-00  6870926 N-FC-00  6870927 N-FC-00  6870929 N-FC-00  6870933 N-FC-00  6870934 N-FC-00  6870936 N-FC-00  6881385 N-FC-00  6881386 N-FC-00  6881389 N-FC-00  6881390 N-FC-00  6881391 N-FC-00  6881392 N-FC-00  6881394 N-FC-00  6884065 N-FC-00  6884070 N-FC-00  6884071 N-FC-00  6884072 N-FC-00  6884074 N-FC-00  6884081 N-FC-00  6884082 N-FC-00  6884084 N-FC-00  6885972 N-FC-00  6885974 N-FC-00  6885987 N-FC-00  6885988 N-FC-00  6889538 N-FC-00  6889539 N-FC-00  6889540 N-FC-00  6889551 N-FC-00  6889556 N-FC-00  6829269 N-FC-00  6829279 N-FC-00  6831269 N-FC-00  6831270 N-FC-00  6831281 N-FC-00  6837813 N-FC-00  6839567 N-FC-00  6829275 N-FC-00  6831278 N-FC-00  6837807 N-FC-00  6881402 N-FC-00  6829287 N-FC-00  6870959 N-FC-00  6881146 N-FC-00  6881377 N-FC-00  6881379 N-FC-00  6881381 N-FC-00  6839591 N-FC-00-EN  6831244 N-FC-00-EN  6831258 N-FC-00-EN  6831259 N-FC-00-EN  6831260 N-FC-00-EN  6884078 N-FC-00-EN  6849934 N-FCREC-00  6834302 N-FCREC-00  6834308 N-FCREC-00  6834309 N-FCREC-00  6834310 N-FCREC-00  6834311 N-FCREC-00  6834312 N-FCREC-00  6834313 N-FCREC-00  6834314 N-FCREC-00  6834315 N-FCREC-00  6834316 N-FCREC-00  6834317 N-FCREC-00  6834318 N-FCREC-00  6834319 N-FCREC-00  6834320 N-FCREC-00  6839180 N-FCREC-00  6839182 N-FCREC-00  6839183 N-FCREC-00  6839184 N-FCREC-00  6839185 N-FCREC-00  6839186 N-FCREC-00  6839187 N-FCREC-00  6839188 N-FCREC-00  6839189 N-FCREC-00  6839190 N-FCREC-00  6839191 N-FCREC-00  6839192 N-FCREC-00  6839193 N-FCREC-00  6839194 N-FCREC-00  6839195 N-FCREC-00  6839196 N-FCREC-00  6839197 N-FCREC-00  6839198 N-FCREC-00  6839199 N-FCREC-00  6839200 N-FCREC-00  6839201 N-FCREC-00  6839202 N-FCREC-00  6839203 N-FCREC-00  6839204 N-FCREC-00  6839205 N-FCREC-00  6839206 N-FCREC-00  6839207 N-FCREC-00  6839208 N-FCREC-00  6839209 N-FCREC-00  6839210 N-FCREC-00  6839211 N-FCREC-00  6846254 N-FCREC-00  6846256 N-FCREC-00  6846258 N-FCREC-00  6846259 N-FCREC-00  6849935 N-FCREC-00  6849940 N-FCREC-00  6849942 N-FCREC-00  6849946 N-FCREC-00  6849949 N-FCREC-00  6849950 N-FCREC-00  6863457 N-FCREC-00  6863459 N-FCREC-00  6863460 N-FCREC-00  6863462 N-FCREC-00  6863470 N-FCREC-00  6863475 N-FCREC-00  6877795 N-FCREC-00  6877796 N-FCREC-00  6879675 N-FCREC-00  6879678 N-FCREC-00  6879680 N-FCREC-00  6881407 N-FCREC-00  6877797 N-FCREC-00  6879677 N-FCREC-00  6863476 N-FCREC-00  6863477 N-FCREC-00  6876315 N-FCREC-00  6876321 N-FCREC-00  6876322 N-FCREC-00  6876324 N-FCREC-00  6876325 N-FCREC-00  6876326 N-FCREC-00  6876327 N-FCREC-00  6876331 N-FCREC-00  6876333 N-FCREC-00  6876334 N-FCREC-00  6877787 N-FCREC-00  6877788 N-FCREC-00  6877789 N-FCREC-00  6877790 N-FCREC-00  6877791 N-FCREC-00  6877792 N-FCREC-00  6877798 N-FCREC-00  6879589 N-FCREC-00  6879590 N-FCREC-00  6879591 N-FCREC-00  6879592 N-FCREC-00  6879593 N-FCREC-00  6879594 N-FCREC-00  6879595 N-FCREC-00  6879596 N-FCREC-00  6879597 N-FCREC-00  6879598 N-FCREC-00  6839765 N-FCREC-00  6877785 N-FCREC-00  6876323 N-FCREC-00  6881423 N-FCREC-00  6881424 N-FCREC-00  6849939 N-FCREC-00  6839181 N-FCREC-00  6846255 N-FCREC-00  6876332 N-FCREC-00  6877793 N-FCREC-00  6877794 N-FCREC-00  6879599 N-FCREC-00  6879673 N-FCREC-00  6879676 N-FCREC-00  6879679 N-FCREC-00  6879681 N-FCREC-00  6879682 N-FCREC-00  6879683 N-FCREC-00  6881144 N-FCREC-00  6881145 N-FCREC-00  6881409 N-FCREC-00  6881410 N-FCREC-00  6881411 N-FCREC-00  6881412 N-FCREC-00  6881413 N-FCREC-00  6881414 N-FCREC-00  6881415 N-FCREC-00  6881416 N-FCREC-00  6881417 N-FCREC-00  6881419 N-FCREC-00  6892569 N-FCREC-00  6849936 N-FCREC-00  6849937 N-FCREC-00  6849938 N-FCREC-00  6849941 N-FCREC-00  6881408 N-FCREC-00  6881418 N-FCREC-00  6830983 N-FCREC-00-EN  6830984 N-FCREC-00-EN  6830985 N-FCREC-00-EN  6830986 N-FCREC-00-EN  6830987 N-FCREC-00-EN  6830992 N-FCREC-00-EN  6830993 N-FCREC-00-EN  6839212 N-FCREC-00-EN  6839761 N-FCREC-00-EN  6839762 N-FCREC-00-EN  6839763 N-FCREC-00-EN  6839764 N-FCREC-00-EN  6839766 N-FCREC-00-EN  6839767 N-FCREC-00-EN  6839768 N-FCREC-00-EN  6846241 N-FCREC-00-EN  6846242 N-FCREC-00-EN  6846243 N-FCREC-00-EN  6846244 N-FCREC-00-EN  6846245 N-FCREC-00-EN  6846246 N-FCREC-00-EN  6846247 N-FCREC-00-EN  6846248 N-FCREC-00-EN  6846249 N-FCREC-00-EN  6846250 N-FCREC-00-EN  6846251 N-FCREC-00-EN  6846252 N-FCREC-00-EN  6846253 N-FCREC-00-EN  6846257 N-FCREC-00-EN  6846260 N-FCREC-00-EN  6849908 N-FCREC-00-EN  6849909 N-FCREC-00-EN  6849910 N-FCREC-00-EN  6849911 N-FCREC-00-EN  6849912 N-FCREC-00-EN  6849913 N-FCREC-00-EN  6849914 N-FCREC-00-EN  6849915 N-FCREC-00-EN  6849916 N-FCREC-00-EN  6849917 N-FCREC-00-EN  6849918 N-FCREC-00-EN  6849919 N-FCREC-00-EN  6849920 N-FCREC-00-EN  6849921 N-FCREC-00-EN  6849922 N-FCREC-00-EN  6849923 N-FCREC-00-EN  6849924 N-FCREC-00-EN  6849925 N-FCREC-00-EN  6849926 N-FCREC-00-EN  6849927 N-FCREC-00-EN  6849928 N-FCREC-00-EN  6849930 N-FCREC-00-EN  6849931 N-FCREC-00-EN  6849932 N-FCREC-00-EN  6849933 N-FCREC-00-EN  6849943 N-FCREC-00-EN  6849944 N-FCREC-00-EN  6849945 N-FCREC-00-EN  6849947 N-FCREC-00-EN  6849948 N-FCREC-00-EN  6849951 N-FCREC-00-EN  6849952 N-FCREC-00-EN  6849953 N-FCREC-00-EN  6849954 N-FCREC-00-EN  6849955 N-FCREC-00-EN  6863448 N-FCREC-00-EN  6863449 N-FCREC-00-EN  6863450 N-FCREC-00-EN  6863451 N-FCREC-00-EN  6863452 N-FCREC-00-EN  6863453 N-FCREC-00-EN  6863454 N-FCREC-00-EN  6863455 N-FCREC-00-EN  6863456 N-FCREC-00-EN  6863458 N-FCREC-00-EN  6863461 N-FCREC-00-EN  6863463 N-FCREC-00-EN  6863464 N-FCREC-00-EN  6863465 N-FCREC-00-EN  6863466 N-FCREC-00-EN  6863467 N-FCREC-00-EN  6863469 N-FCREC-00-EN  6863471 N-FCREC-00-EN  6863472 N-FCREC-00-EN  6863473 N-FCREC-00-EN  6863474 N-FCREC-00-EN  6876305 N-FCREC-00-EN  6876306 N-FCREC-00-EN  6876307 N-FCREC-00-EN  6876308 N-FCREC-00-EN  6876309 N-FCREC-00-EN  6876310 N-FCREC-00-EN  6876311 N-FCREC-00-EN  6876312 N-FCREC-00-EN  6876313 N-FCREC-00-EN  6876314 N-FCREC-00-EN  6876316 N-FCREC-00-EN  6876317 N-FCREC-00-EN  6876318 N-FCREC-00-EN  6876319 N-FCREC-00-EN  6876320 N-FCREC-00-EN  6876328 N-FCREC-00-EN  6876329 N-FCREC-00-EN  6876330 N-FCREC-00-EN  6877784 N-FCREC-00-EN  6877786 N-FCREC-00-EN  6879600 N-FCREC-00-EN  6879672 N-FCREC-00-EN  6879684 N-FCREC-00-EN  6879685 N-FCREC-00-EN  6881422 N-FCREC-00-EN  6881425 N-FCREC-00-EN  6881426 N-FCREC-00-EN  6881427 N-FCREC-00-EN  6881428 N-FCREC-00-EN  6881429 N-FCREC-00-EN  6881432 N-FCREC-00-EN  6881433 N-FCREC-00-EN  6881434 N-FCREC-00-EN  6881435 N-FCREC-00-EN  6881436 N-FCREC-00-EN  6905190 N-FCREC-00-EN  6905206 N-FCREC-00-EN  6879671 N-FCREC-00-EN  6849929 N-FCREC-00-EN  6863468 N-FCREC-00-EN  6879674 N-FCREC-00-EN"
    }
  ]
}