  "results": [
    {
      "country": "United States",
      "city": "Montvale",
      "address_1": "3 Paragon Dr",
      "reason_for_recall": "Pentax has become aware that operational/cleaning accessories and therapeutic devices (e.g., clips, stents, balloons, etc.) can become lodged in an endoscope's instrument channel.",
      "address_2": "",
      "product_quantity": "4,195 units",
      "code_info": "Model #'s FB-10V, FB015BS, FB-15P, FB-15RBS, FB-15V, FB-18BS, FB-18P, FB-18RBS, FB-18V, FB-19TV, FB-8V, EB--1530T3, EB-1170K, EB01570K, EB-1575K, EB-1970K, EB-1970-TK and EB-1975K",
      "center_classification_date": "20160628",
      "distribution_pattern": "Worldwide Distribution - US (Nation Wide) including Puerto Rico and Internationally to Denmark, Guam",
      "state": "NJ",
      "product_description": "Bronchoscope  Intended to provide optical visualization of and therapeutic access to the Airway and Bronchial Tree. This anatomy includes, but is not restricted to, the organs, tissues and subsystems: Nasal Passage, Trachea and Bronchial Tree. The instrument is introduced via the mouth or the nose, as decided by the physician, when indications consistent with the need for procedure are observed in adult and pediatric patient populations.",
      "report_date": "20160706",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Pentax Medical Company",
      "recall_number": "Z-2065-2016",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "73894",
      "termination_date": "20171221",
      "more_code_info": "",
      "recall_initiation_date": "20160406",
      "postal_code": "07645-1782",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}