  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Salt Lake City",
      "state": "UT",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85575",
      "recalling_firm": "Innovasis, Inc",
      "address_1": "614 E 3900 S",
      "address_2": "N/A",
      "postal_code": "84107-1902",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution including in the states of CA, AL, LA.",
      "recall_number": "Z-2062-2020",
      "product_description": "Oryx Cervical Screw Caddy, Model Number BTA-172, as part of the Oryx Cervical Plate System - Product Usage: The Cervical Plate System is intended for use in anterior cervical fixation.",
      "product_quantity": "6",
      "reason_for_recall": "Cervical screw caddies have incorrect text markings for the 4.5mm Fixed and Variable rescue screws. The caddy location labeled as \"Fixed\" should be labeled as \"Variable\" and the caddy location labeled as \"Variable\" should be labeled as \"Fixed. If user installs a fixed rescue screw instead of variable, the screw head may not be flush, it could back out, which could interfere with the cover screw.",
      "recall_initiation_date": "20200407",
      "center_classification_date": "20200520",
      "termination_date": "20220126",
      "report_date": "20200527",
      "code_info": "Serial numbers:  2339,  2340,  2346,  2351,  2352,  2354,"
    }
  ]
}