{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Sunnyvale",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "92418",
      "recalling_firm": "Intuitive Surgical, Inc.",
      "address_1": "1266 Kifer Rd Bldg 101",
      "address_2": "N/A",
      "postal_code": "94086-5304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide distribution in the states of PR, FL, MN, PA, AZ, OH, IL, NC, GA, KS, CA, HI, MD, OR, SC, AK, NY, NE, IA, WA, OK, ND, UT, MS, AR, WI, MI, TX, TN, IN, NJ, VA, SD, CO, KY, AL, MO, LA, DE, MA, MT, WV, ID, CT, NV, NH, ME, WY, DC, RI, NM, VT and the countries of Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Cyprus, Czech Republic, Denmark, Dominican Republic, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Luxembourg, Mexico, Monaco, Netherlands, Norway, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, San Marino, Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom.",
      "recall_number": "Z-2059-2023",
      "product_description": "da Vinci Xi/X Tip-Up Fenestrated Grasper, REF: 470347, VER: 12",
      "product_quantity": "46,033",
      "reason_for_recall": "Fenestrated Graspers have a manufacturing issue has the potential to cause breakage on both upper and/or lower grips, which could lead to stainless steel fragments falling into the patient that may be either detected or undetected.",
      "recall_initiation_date": "20230612",
      "center_classification_date": "20230630",
      "report_date": "20230712",
      "code_info": "UDI-DI: 00886874112496, All Lots."
    }
  ]
}