  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Thornwood",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80115",
      "recalling_firm": "Zeiss, Carl Inc",
      "address_1": "1 Zeiss Dr",
      "address_2": "N/A",
      "postal_code": "10594-1939",
      "voluntary_mandated": "FDA Mandated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "USA (nationwide)",
      "recall_number": "Z-2058-2018",
      "product_description": "Cell Observer SD, DirectFRAP, Laser TIRF 1, Laser TIRF 2, Laser TIRF3, LightSheet, Elyra, and multiple   Laser Scanning Microscopes.   Laser scanning microscopes are used in cell biology research.",
      "product_quantity": "1070 units to US",
      "reason_for_recall": "In certain eyepiece configurations the laser shutter may not close completely, permitting reflected beams that may be greater than the Class I limit.",
      "recall_initiation_date": "20180508",
      "center_classification_date": "20180711",
      "termination_date": "20200421",
      "report_date": "20180620",
      "code_info": "Model No.  Cell Observer SD, Laser TIRF, and DirectFRAP Laser Scanning Microscopes"
    }
  ]
}