  "results": [
    {
      "status": "Ongoing",
      "city": "Miami Lakes",
      "state": "FL",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85397",
      "recalling_firm": "Heartware, Inc.",
      "address_1": "14400 Nw 60th Ave",
      "address_2": "N/A",
      "postal_code": "33014-2807",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution",
      "recall_number": "Z-2056-2020",
      "product_description": "Heartware HVAD Pump Implant Kit, Product (REF) Number 1104",
      "product_quantity": "2225 Units",
      "reason_for_recall": "The Outflow Graft may be subject to tears and the Strain Relief screw may break during the pre-implant pump assembly and attachment to the HVAD Pump.",
      "recall_initiation_date": "20200403",
      "center_classification_date": "20200522",
      "report_date": "20200603",
      "code_info": "GTIN 00888707002028, Serial Numbers:  HW28848, hw29011, HW29154, HW29211, HW29248, HW29249, HW29273, HW29329, HW29426, HW29465, HW29472, HW29522, HW29534, HW29548, HW29549, HW29567, HW29588, HW29593, HW29602, HW29612, HW29667, HW29700, HW29786, HW29815, HW29915, HW29928, HW29949, HW30039, HW30082, HW30083, HW30150, HW30153, HW30208, HW30211, HW30226, HW30320, HW30342, HW30358, HW30360, HW30363, HW30378, HW30380, HW30382, HW30390, HW30398, HW30400, HW30402, HW30406, HW30410, HW30430, HW30475, HW30484, HW30494, HW30498, HW30595, HW30596, HW30597, HW30601, HW30604, HW30651, HW30768, HW30769, HW30770, HW30771, HW30814, HW30821, HW30829, HW30863, HW30865, HW30867, HW30871, HW30953, HW30961, HW30962, HW30966, HW30971, HW30972, HW30974, HW31020, HW31076, HW31101, HW31103, HW31104, HW31107, HW31113, HW31115, HW31117, HW31119, HW31128, HW31132, HW31134, HW31192, HW31219, HW31223, HW31232, HW31235, HW31242, HW31244, HW31249, HW31254, HW31255, HW31260, HW31261, HW31262, HW31267, HW31383, HW31387, HW31388, HW31399, HW31409, HW31410, HW31411, HW31412, HW31414, hw31445, HW31446, hw31447, hw31448, hw31449, hw31451, HW31452, hw31453, hw31454, HW31455, HW31456, HW31457, HW31458, HW31541, HW31543, HW31544, HW31545, HW31547, HW31548, HW31549, HW31550, HW31552, HW31559, HW31564, HW31571, HW31573, HW31578, HW31579, HW31686, HW31694, HW31697, HW31731, HW31749, HW31750, HW31777, HW31778, HW31810, HW31819, HW31830, HW31831, HW31832, HW31834, HW31835, HW31836, HW31838, HW31839, HW31840, HW31842, HW31843, HW31844, HW31845, HW31846, HW31847, HW31849, HW31855, HW31858, HW31876, HW31883, HW31944, HW31947, HW31979, HW31981, HW31986, HW31987, HW31988, HW31989, HW31991, HW32068, HW32116, HW32121, HW32141, HW32143, HW32156, HW32159, HW32161, HW32162, HW32164, HW32167, HW32172, HW32179, HW32189, HW32190, HW32192, HW32193, HW32194, HW32197, HW32199, HW32201, HW32204, HW32206, HW32209, HW32210, HW32214, HW32215, HW32216, HW32218, HW32220, HW32222, HW32223, HW32224, HW32228, HW32229, HW32230, HW32231, HW32248, HW32258, HW32259, HW32260, HW32261, HW32273, HW32279, HW32281, HW32285, HW32286, HW32290, HW32292, HW32315, HW32335, HW32357, HW32386, HW32393, HW32400, HW32419, HW32423, HW32426, HW32427, HW32457, HW32458, HW32459, HW32464, HW32465, HW32468, HW32470, HW32471, HW32472, HW32474, HW32479, HW32499, HW32504, HW32508, HW32509, HW32510, HW32511, HW32516, HW32534, HW34001, HW34002, HW34025, HW34034, HW34035, HW34036, HW34037, HW34038, HW34040, HW34041, HW34042, HW34043, HW34044, HW34045, HW34046, HW34047, HW34048, HW34051, HW34066, HW34096, HW34100, HW34104, HW34106, HW34109, HW34110, HW34111, HW34113, HW34117, HW34121, HW34126, HW34151, HW34169, HW34179, HW34181, HW34186, HW34187, HW34192, HW34193, HW34196, HW34197, HW34202, HW34204, HW34205, HW34207, HW34209, HW34211, HW34226, HW34317, HW34319, HW34326, HW34328, HW34329, HW34330, HW34331, HW34336, HW34342, HW34344, HW34347, HW34368, HW34369, HW34371, HW34372, HW34373, HW34375, HW34376, HW34377, HW34378, HW34379, HW34425, HW34427, HW34428, HW34429, HW34432, HW34446, HW34495, HW34519, HW34520, HW34521, HW34522, HW34523, HW34546, HW34547, HW34550, HW34551, HW34555, HW34556, HW34588, HW34590, HW34596, HW34603, HW34605, HW34609, HW34610, HW34643, HW34673, HW34752, HW34768, HW34805, HW34806, HW34808, HW34810, HW34824, HW34840, HW34842, HW34845, HW34866, HW34868, HW34877, HW34902, HW34903, HW34904, HW34908, HW34910, HW34913, HW34914, HW34923, HW34924, HW34925, HW34928, HW34933, HW34934, HW34938, HW34939, HW34944, HW34946, HW34947, HW34948, HW34949, HW34957, HW34979, HW34990, HW34997, HW35012, HW35032, HW35099, HW35122, HW35123, HW35124, HW35125, HW35127, HW35128, HW35131, HW35133, HW35134, HW35136, HW35138, HW35141, HW35142, HW35143, HW35161, HW35189, HW35191, HW35204, HW35205, HW35209, HW35211, HW35212, HW35213, HW35214, HW35215, HW35220, HW35284, HW35305, HW35317, HW35319, HW35321, HW35324, HW35325, HW35326, HW35332, HW35334, HW35337, HW35339, HW35340, HW35341, HW35357, HW35360, HW35361, HW35367, HW35385, HW35464, HW35477, HW35478, HW35479, HW35480, HW35481, HW35482, HW35483, HW35487, HW35488, HW35489, HW35492, HW35493, HW35494, HW35495, HW35499, HW35500, HW35501, HW35502, HW35504, HW35506, HW35508, HW35534, HW35535, HW35539, HW35540, HW35541, HW35542, HW35543, HW35545, HW35548, HW35563, HW35577, HW35578, HW35580, HW35582, HW35583, HW35584, HW35586, HW35587, HW35588, HW35624, HW35625, HW35626, HW35627, HW35628, HW35629, HW35630, HW35631, HW35632, HW35633, HW35634, HW35635, HW35636, HW35637, HW35639, HW35641, HW35644, HW35646, HW35647, HW35668, HW35669, HW35670, HW35671, HW35673, HW35674, HW35676, HW35677, HW35678, HW35680, HW35681, HW35682, HW35683, HW35684, HW35686, HW35687, HW35688, HW35689, HW35690, HW35691, HW35692, HW35693, HW35695, HW35696, HW35697, HW35698, HW35701, HW35702, HW35703, HW35704, HW35705, HW35707, HW35708, HW35730, HW35731  GTIN 00888707002387, Serial Numbers:  HW28851, HW29783, HW29793, HW29794, HW29827, HW29870, HW29918, HW30045, HW30046, HW30505, HW30765, HW30842, HW30844, HW30845, HW30846, hw30978, HW31075, HW31077, HW31080, hw31089, hw31091, HW31095, HW31161, HW31342, HW31348, HW31555, HW31556, HW31557, HW31683, HW31860, HW31903, HW31920, HW31924, HW31926, HW32026, HW32027, HW32028, HW32149, HW32153, HW32163, HW32282, HW32513, HW32514, HW32562, HW32563, HW34004, HW34006, HW34091, HW34092, HW34125, HW34272, HW34273, HW34274, HW34349, HW34350, HW34351, HW34352, HW34357, HW34358, HW34359, HW34366, HW34374, HW34436, HW34953, HW34968, HW35080, HW35144, HW35145, HW35148, HW35327, HW35342, HW35344, HW35537, HW35596, HW35603, HW35604, HW35605, HW35606, HW35607, HW35713, HW35714, HW35717, HW35718, HW35721, HW35723  GTIN 00888707001649, Serial Numbers:  HW29132, HW29232, hw29653, HW29682, HW29684, HW29705, HW29864, HW29985, HW29991, HW30096, HW30100, HW30145, HW30146, HW30373, HW30374, HW30421, HW30525, HW30529, HW30726, HW30738, HW30836, HW30840, HW30862, HW30939, hw30949, hw30950, HW31048, HW31071, HW31082, HW31088, hw31133, hw31394, HW31415, HW31417, HW31420, HW31421, HW31423, hw31483, hw31505, hw31506, hw31507, hw31510, HW31511, hw31512, hw31513, hw31515, hw31516, hw31517, HW31546, HW31551, HW31554, HW31566, HW31587, HW31588, HW31589, HW31590, HW31592, HW31632, HW31639, HW31685, HW31795, HW31801, HW31806, HW31808, HW31809, HW31811, HW31812, HW31813, HW31815, HW31891, HW31892, HW31893, HW31919, HW31921, HW31973, HW31982, HW31985, HW32000, HW32001, HW32014, HW32015, HW32018, HW32021, HW32022, HW32024, HW32042, HW32050, HW32055, HW32056, HW32064, HW32081, HW32094, HW32112, HW32123, HW32138, HW32147, HW32186, HW32245, HW32255, HW32264, HW32265, HW32268, HW32270, HW32301, HW32394, HW32410, HW32420, HW32558, HW32559, HW32560, HW32561, HW34029, HW34031, HW34093, HW34094, HW34132, HW34148, HW34180, HW34266, HW34268, HW34269, HW34270, HW34271, HW34296, HW34301, HW34324, HW34327, HW34340, HW34341, HW34343, HW34345, HW34346, HW34348, HW34418, HW34420, HW34431, HW34507, HW34508, HW34641, HW34692, HW34745, HW34764, HW34836, HW34906, HW34932, HW34936, HW34940, HW34941, HW34963, HW34977, HW34992, HW34994, HW35019, HW35021, HW35033, HW35035, HW35038, HW35041, HW35065, HW35068, HW35085, HW35090, HW35094, HW35114, HW35147, HW35157, HW35196, HW35197, HW35210, HW35237, HW35250, HW35290, HW35308, HW35309, HW35310, HW35311, HW35312, HW35313, HW35314, HW35315, HW35318, HW35320, HW35328, HW35329, HW35330, HW35333, HW35338, HW35349, HW35365, HW35366, HW35368, HW35370, HW35371, HW35372, HW35373, HW35375, HW35378, HW35402, HW35404, HW35405, HW35408, HW35409, HW35411, HW35412, HW35418, HW35421, HW35423, HW35465, HW35470, HW35475, HW35476, HW35485, HW35486, HW35546, HW35547, HW35550, HW35551, HW35552, HW35553, HW35554, HW35555, HW35556, HW35557, HW35558, HW35559, HW35560, HW35564, HW35598, HW35599, HW35602, HW35640, HW35657, HW35658, HW35659, HW35662, HW35663, HW35664, HW35665, HW35666, HW35667, HW35672, HW35694, HW35719, HW35750, HW35753, HW36121, HW36154, HW36165, HW36168  GTIN 00888707000581, Serial Numbers:  HW29203, HW29425, HW29427, HW29428, HW29430, HW29439, HW29517, HW29679, HW29680, HW30589, HW30638, HW30706, HW30876, HW31144, HW31187, hw31376, hw31380, HW31442, HW31727, HW31998, HW32023, HW32416, HW34017  GTIN 00888707002417, Serial Numbers:  HW29478, HW30381, HW30831, HW34380, HW34421, HW35348, HW35613, HW35756, HW35761, HW35763   GTIN 00888707002493, Serial Numbers:  HW29483, HW29697, HW29698, hw31178, HW31520, HW31821, HW31823, HW31827, HW32086, HW32242, HW32246, HW32518, HW32519, HW34021, HW34365, HW34367, HW34792, HW34915, HW35155, HW35158, HW35252, HW35343, HW35345, HW35346, HW35350, HW35400, HW35401, HW35529, HW35530, HW35574, HW35591, HW35660, HW35661, HW35785  GTIN 00888707002400, Serial Numbers:  HW29518, HW29693, HW30092, HW32257, HW34012, HW34210, HW35034, HW35154, HW35573, HW35653  GTIN 00888707002547, Serial Numbers:  HW29561, HW29601, HW29931, HW30502, HW30593, HW30619, HW30632, HW30783, HW30796, HW30819, HW30826, HW31136, HW31140, HW31141, HW31156, HW31157, HW31158, HW31382, HW31393, HW31406, HW31408, HW31413, HW31416, HW31426, HW31429, HW31553, HW31791, HW31937, HW31994, HW32118, HW32269, HW32309, HW32336, HW32347, HW32373, HW32385, HW32522, HW32523, HW32524, HW32554, HW32557, HW34003, HW34007, HW34008, HW34049, HW34050, HW34053, HW34054, HW34382, HW34383, HW34513, HW34514, HW34516, HW34517, HW34518, HW35251, HW35275  GTIN 00888707000680, Serial Numbers:  HW29571, HW29578, HW29845, HW30635, HW30639, HW30711, HW31150  GTIN 00888707002394, Serial Numbers:  HW30469, HW30506, HW30508, HW30509, HW30510, HW30511, HW30512, HW30655, HW31580, HW31582, HW31583, HW31584, HW31992, HW31993, HW32200, HW32207, HW32217, HW32238, HW32239, HW32240, HW32515, HW32517, HW34022, HW34023, HW34395, HW34461, HW34557, HW34561, HW34582, HW34751, HW34950, HW34955, HW34991, HW34996, HW35016, HW35149, HW35150, HW35186, HW35336, HW35386, HW35397, HW35484, HW35509, HW35510, HW35538, HW35566, HW35609, HW35610, HW35611, HW35648, HW35649, HW35651, HW35679, HW35699, HW35700, HW35706, HW35776, HW35779  GTIN 00888707000239, Serial Numbers:  HW30634, HW30645, HW30877, HW34737  GTIN 00888707002530, Serial Numbers:  HW30797, HW31108, HW31440, HW31828, HW31928, HW32182, HW32253, HW32263, HW34559, HW35000, HW35162, HW35532, HW35533, HW35590   GTIN 00888707002523, Serial Numbers:  HW30859, HW31703, HW31704, HW31859, HW32030, HW32036, HW32040, HW32102, HW32520, HW32521, HW34018, HW34020, HW34027, HW34028, HW34986, HW35040, HW35043, HW35049, HW35216, HW35217, HW35218, HW35490, HW35531, HW35619, HW35786  GTIN 00888707002431, Serial Numbers:  hw30979, hw31123, HW32029, HW32180, HW34024, HW34381, HW34615, HW35119, HW35733  GTIN 00888707002585, Serial Numbers:  hw31079, hw31081, hw31168, hw31170, HW35095, HW35164, HW35351, HW35353, HW35355, HW35496, HW35497, HW35575, HW35576   GTIN 00888707002516, Serial Numbers:  hw31147, hw31162, HW32507, HW34524, HW35474  GTIN 00888707006330, Serial Numbers:  HW31419, HW31699, HW31733, HW32505, HW35393, HW35396, HW35726, HW35749, HW35766, HW35769, HW35784, HW35829, HW35872, HW36005, HW36007, HW36008, HW36009, HW36010, HW36013, HW36014, HW36015, HW36018, HW36019, HW36020, HW36021, HW36067, HW36068, HW36069, HW36070, HW36071, HW36131, HW36132, HW36138, HW36174, HW36175, HW36176, HW36177, HW36178, HW36179, HW36180, HW36182, HW36184, HW36185, HW36186, HW36192, HW36193, HW36194, HW36195, HW36196, HW36197, HW36198, HW36199, HW36200, HW36201, HW36224, HW36225, HW36226, HW36227, HW36228, HW36229, HW36230, HW36231, HW36232, HW36233, HW36248, HW36249, HW36251, HW36252, HW36253, HW36254, HW36255, HW36256, HW36258, HW36259, HW36260, HW36276, HW36298, HW36299, HW36300, HW36303, HW36305, HW36306, HW36307, HW36309, HW36310, HW36313, HW36314, HW36316, HW36317, HW36318, HW36319, HW36320, HW36321, HW36323, HW36324, HW36325, HW36326, HW36328, HW36332, HW36333, HW36334, HW36335, HW36337, HW36354, HW36355, HW36356, HW36357, HW36358, HW36359, HW36360, HW36361, HW36362, HW36363, HW36396, HW36397, HW36398, HW36399, HW36400, HW36401, HW36403, HW36404, HW36405, HW36406, HW36407, HW36408, HW36409, HW36410, HW36411, HW36412, HW36419, HW36420, HW36428, HW36429, HW36430, HW36432, HW36433, HW36434, HW36435, HW36436, HW36437, HW36438, HW36500, HW36501, HW36502, HW36503, HW36504, HW36505, HW36506, HW36507, HW36508, HW36509, HW36510, HW36511, HW36513, HW36514, HW36515, HW36530, HW36553, HW36554, HW36555, HW36556, HW36557, HW36558, HW36559, HW36560, HW36561, HW36562, HW36563, HW36564, HW36595, HW36596, HW36597, HW36598, HW36599, HW36600, HW36601, HW36602, HW36603, HW36624, HW36625, HW36626, HW36628, HW36630, HW36631, HW36632, HW36633, HW36636, HW36637, HW36655, HW36659, HW36660, HW36662, HW36663, HW36664, HW36669, HW36670, HW36671, HW36672, HW36703, HW36704, HW36705, HW36708, HW36709, HW36710, HW36711, HW36712, HW36713, HW36853, HW36854, HW36855, HW36898, HW36899, HW36900, HW36903, HW36919, HW36921  GTIN 00888707002486, Serial Numbers:  HW31444, HW31569, HW31673, HW31814, HW31817, HW31818, HW31820, HW31824, HW31825, HW31826, HW31872, HW31873, HW31875, HW31877, HW31927, HW31932, HW31934, HW31939, HW31940, HW31941, HW31995, HW31996, HW31997, HW31999, HW32002, HW32008, HW32009, HW32010, HW32011, HW32013, HW32032, HW32034, HW32038, HW32046, HW32053, HW32057, HW32062, HW32065, HW32071, HW32085, HW32109, HW32165, HW32166, HW32195, HW32205, HW32211, HW32221, HW32236, HW32243, HW32250, HW32506, HW34015, HW34122, HW34123, HW34124, HW34251, HW34252, HW34275, HW34277, HW34280, HW34354, HW34355, HW34356, HW34361, HW34364, HW34433, HW34434, HW34437, HW34439, HW34440, HW34442, HW34443, HW34444, HW34452, HW34454, HW34619, HW34620, HW34621, HW34622, HW34623, HW34624, HW34629, HW34630, HW34631, HW34636, HW34656, HW34719, HW34738, HW34739, HW34742, HW34748, HW34749, HW34750, HW34753, HW34757, HW34758, HW34761, HW34762, HW34767, HW34769, HW34779, HW34818, HW34853, HW34867, HW34873, HW34879, HW34880, HW34882, HW34885, HW34888, HW34890, HW34897, HW34898, HW34900, HW34905, HW34909, HW34911, HW34912, HW34945, HW34954, HW34960, HW35171, HW35228, HW35230, HW35278, HW35282, HW35283, HW35297, HW35298, HW35300, HW35303, HW35304, HW35306, HW35307, HW35512, HW35513, HW35514, HW35516, HW35517, HW35518, HW35519, HW35520, HW35521, HW35522, HW35523, HW35525, HW35526, HW35527, HW35528, HW35565, HW35567, HW35568, HW35569, HW35570, HW35593, HW35600, HW35612, HW35614, HW35615, HW35616, HW35617, HW35618, HW35645, HW35650, HW35652, HW35685, HW35725, HW35734, HW35737, HW35738, HW35740, HW35742, HW35743, HW35744, HW35745, HW35746, HW35747, HW35748, HW35751, HW35752, HW35760, HW35762, HW35765  GTIN 00888707002462, Serial Numbers:  HW31586, HW31700, HW32035  GTIN 00888707002424, Serial Numbers:  HW31698, HW31851  GTIN 00888707002448, Serial Numbers:  HW31779, HW31780, HW32546, HW34095, HW34617, HW35398, HW35511, HW35589, HW35597   GTIN 00888707002455, Serial Numbers:  HW31852, HW31856, HW32025, HW32198, HW34026, HW34637, HW35544  GTIN 00888707006019, Serial Numbers:  HW31983, HW34720, HW34974, HW35376, HW35377, HW35849, HW35866, HW35869, HW35871, HW35873, HW35877, HW35878, HW35879, HW35880, HW35998, HW36000, HW36056, HW36057, HW36058, HW36059, HW36060, HW36061, HW36062, HW36063, HW36113, HW36114, HW36115, HW36116, HW36117, HW36118, HW36119, HW36120, HW36145, HW36146, HW36155, HW36156, HW36157, HW36158, HW36159, HW36167, HW36221, HW36222, HW36223, HW36350, HW36351, HW36352  GTIN 00888707006408, Serial Numbers:  HW34292, HW35932, HW36083, HW36084, HW36087, HW36129, HW36130, HW36171, HW36172, HW36173, HW36212, HW36247, HW36250, HW36261, HW36263, HW36264, HW36265, HW36266, HW36348, HW36349, HW36413, HW36414, HW36415, HW36416, HW36594, HW36638  GTIN 00888707002554, Serial Numbers:  HW34337, HW34362, HW34484, HW34591, HW34594, HW34625, HW34682, HW34693, HW34732, HW34833, HW34838, HW34846, HW34876, HW34884, HW34889, HW35052, HW35057, HW35061, HW35067, HW35316, HW35323, HW35359, HW35362, HW35363, HW35364, HW35458, HW35536, HW35585, HW35620, HW35622, HW35623, HW35709, HW35710, HW35728, HW35732, HW35758, HW35759, HW35764, HW35767, HW35771, HW35772, HW35775, HW35777, HW35780, HW35782  GTIN 00888707004558, Serial Numbers:  HW34759, HW34785, HW35655, HW35656  GTIN 00888707002141, Serial Numbers:  HW34848, HW34849, HW35711  GTIN 00888707006323, Serial Numbers:  HW34982, HW35031, HW35379, HW35380, HW35381, HW35382, HW35383, HW35384, HW35388, HW35389, HW35390, HW35391, HW35392, HW35621, HW35642, HW35654, HW35712, HW35722, HW35727, HW35729, HW35736, HW35755, HW35770, HW35781, HW35783, HW35977, HW36024, HW36025, HW36027, HW36028, HW36029, HW36030, HW36032, HW36033, HW36034, HW36042, HW36043, HW36044, HW36045, HW36046, HW36064, HW36065, HW36072, HW36073, HW36074, HW36075, HW36076, HW36077, HW36078, HW36079, HW36080, HW36081, HW36082, HW36123, HW36124, HW36125, HW36126, HW36127, HW36128, HW36133, HW36134, HW36135, HW36136, HW36137, HW36139, HW36140, HW36141, HW36142, HW36143, HW36144, HW36147, HW36148, HW36149, HW36150, HW36151, HW36152, HW36153, HW36187, HW36188, HW36189, HW36190, HW36191, HW36202, HW36203, HW36204, HW36206, HW36207, HW36208, HW36209, HW36210, HW36234, HW36235, HW36236, HW36237, HW36238, HW36239, HW36242, HW36244, HW36245, HW36267, HW36268, HW36271, HW36272, HW36274, HW36275, HW36284, HW36286, HW36290, HW36291, HW36292, HW36293, HW36295, HW36296, HW36297, HW36339, HW36340, HW36342, HW36343, HW36344, HW36346, HW36364, HW36366, HW36367, HW36375, HW36378, HW36379, HW36380, HW36439, HW36440, HW36441, HW36443, HW36444, HW36445, HW36446, HW36450, HW36451, HW36452, HW36453, HW36454, HW36455, HW36456, HW36520, HW36521, HW36522, HW36524, HW36528, HW36529, HW36532, HW36533, HW36534, HW36535, HW36537, HW36539, HW36548, HW36549, HW36550, HW36551, HW36552, HW36604, HW36605, HW36606, HW36607, HW36608, HW36639, HW36640, HW36641, HW36645, HW36649, HW36651, HW36661, HW36674, HW36675, HW36676, HW36687, HW36714, HW36729, HW36731, HW36732, HW36733, HW36734, HW36735, HW36737, HW36738, HW36739, HW36744, HW36908, HW36909, HW36922, HW36923, HW36925, HW36932  GTIN 00888707005951, Serial Numbers:  HW35279, HW35331, HW35503, HW35524, HW35581, HW35675, HW35715, HW35720, HW35739, HW35768, HW35787, HW35788, HW35789, HW35790, HW35791, HW35792, HW35793, HW35794, HW35795, HW35796, HW35797, HW35798, HW35799, HW35821, HW35822, HW35823, HW35824, HW35825, HW35826, HW35827, HW35831, HW35833, HW35834, HW35835, HW35836, HW35837, HW35838, HW35839, HW35840, HW35841, HW35842, HW35843, HW35844, HW35845, HW35846, HW35883, HW35887, HW35888, HW35889, HW35890, HW35891, HW35892, HW35893, HW35894, HW35895, HW35896, HW35897, HW35898, HW35900, HW35905, HW35906, HW35907, HW35921, HW35925, HW35926, HW35928, HW35930, HW35933, HW35934, HW35945, HW35946, HW35947, HW35948, HW35949, HW35950, HW35951, HW35952, HW35953, HW35954, HW35955, HW35956, HW35957, HW35959, HW35960, HW35961  GTIN 00888707002264, Serial Numbers:  HW35741, HW35773  GTIN 00888707006316, Serial Numbers:  HW35754, HW35774, HW35865, HW35868, HW35874, HW35940, HW35958, HW35962, HW35963, HW35964, HW35965  GTIN 00888707005968, Serial Numbers:  HW35801, HW35802, HW35803, HW35804, HW35805, HW35806, HW35807, HW35808, HW35809, HW35810, HW35811, HW35812, HW35813, HW35814, HW35815, HW35816, HW35817, HW35818, HW35819, HW35820, HW35847, HW35848, HW35850, HW35851, HW35852, HW35853, HW35854, HW35855, HW35856, HW35857, HW35858, HW35859, HW35860, HW35861, HW35862, HW35863, HW35864, HW35867, HW35870, HW35875, HW35923, HW35924, HW35927, HW35929, HW35931, HW35935, HW35936, HW35937, HW35938, HW35939, HW35966, HW35967, HW35968, HW35969, HW35970, HW35971, HW35972, HW35973, HW35975, HW35976, HW35978, HW35988, HW35989, HW35990, HW35991, HW35993, HW35994, HW35995, HW35996, HW35997  GTIN 00888707006309, Serial Numbers:  HW35876, HW35882, HW35884, HW35885, HW35886, HW35941, HW35942, HW35999  GTIN 00888707006347, Serial Numbers:  HW35902, HW35903, HW35908, HW35909, HW35910, HW35911, HW35912, HW35913, HW35914, HW35915, HW35918, HW35920, HW36001, HW36002, HW36003, HW36016, HW36035, HW36036, HW36037, HW36038, HW36039, HW36040, HW36048, HW36049, HW36050, HW36051, HW36052, HW36053, HW36054, HW36055, HW36088, HW36089, HW36092, HW36093, HW36094, HW36095, HW36096, HW36097, HW36100, HW36101, HW36102, HW36103, HW36105, HW36106, HW36107, HW36108, HW36109, HW36110, HW36111, HW36112, HW36170, HW36540, HW36541, HW36542, HW36543, HW36544, HW36565, HW36566, HW36568, HW36569, HW36570, HW36571, HW36572, HW36573, HW36574, HW36575, HW36576, HW36577, HW36578, HW36579, HW36581, HW36582, HW36583, HW36584, HW36585, HW36586, HW36611, HW36612, HW36615, HW36616, HW36618, HW36619, HW36620, HW36621, HW36622, HW36623  GTIN 00888707006446, Serial Numbers:  HW35979, HW35980, HW35982, HW35984, HW36160, HW36161, HW36162, HW36163, HW36164, HW36278, HW36287, HW36381, HW36382, HW36383, HW36384, HW36385, HW36386, HW36387, HW36388, HW36460, HW36462, HW36464, HW36465, HW36469, HW36470, HW36471, HW36472, HW36480, HW36481, HW36482, HW36483, HW36484, HW36486, HW36487, HW36489, HW36491, HW36492, HW36494, HW36495  GTIN 00888707006392, Serial Numbers:  HW35986, HW35987, HW36066, HW36085, HW36086, HW36213, HW36214, HW36215, HW36216, HW36217, HW36218, HW36219, HW36220, HW36289, HW36311, HW36417, HW36418, HW36517, HW36587, HW36588, HW36589, HW36590, HW36652, HW36656, HW36860  GTIN 00888707007290, Serial Numbers:  HW36421, HW36422  GTIN 00888707006354, Serial Numbers:  HW36838, HW36880  No GTIN, Serial Numbers:  HW26099, HW26126, HW26330, HW26660, HW26939, HW26940, HW27099, HW27452, HW27678, HW28063, HW28199, HW28373, HW28418, HW28895, HW28973, HW29073, HW29214, HW29298, HW29318, HW29351, HW29353, HW29354, HW29359"
    }
  ]
}