      "status": "Terminated",
      "city": "Plymouth",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "77120",
      "recalling_firm": "Greatbatch Medical",
      "address_1": "2300 Berkshire Ln N",
      "address_2": "N/A",
      "postal_code": "55441-4575",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Distribution to the state of : TX",
      "recall_number": "Z-2055-2017",
      "product_description": "Linear Straight Broach Handle (Rasp Handle) (Greatbatch Part # 510397; DJO Surgical Inc., # 803-03-159)  and EZ Clean\" Linear Broach Handle Greatbatch Part # T11824, DJO Surgical Inc. # 803-03-059).",
      "product_quantity": "205",
      "reason_for_recall": "Greatbatch Medical has voluntarily initiated a Recall for the Linear Straight Broach Handle (Rasp Handle) and EZ Clean\" Linear Broach Handle. Greatbatch Medical has received reports that that the product is breaking at the welds and that the locating pin at the distal end of the instrument is breaking off during use.",
      "recall_initiation_date": "20170423",
      "center_classification_date": "20170509",
      "termination_date": "20190307",
      "report_date": "20170517",
      "code_info": "Lot numbers :  Linear Straight Broach Handle  (Rasp Handle):  2988946, 3138186, 3185601, 3287315, 3300879, 3333172, 3433392, 3332697.  EZ Clean\" Linear Broach Handle: 3508491."
    }
  ]
}