{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Flower Mound",
      "state": "TX",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "67813",
      "recalling_firm": "Stryker Communications",
      "address_1": "1410 Lakeside Pkwy Ste 100",
      "address_2": "N/A",
      "postal_code": "75028-4026",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution.  US nationwide including Puerto Rico, China, Philippines, Mexico, Korea, Japan, Colombia, Chile, Canada, Brazil, and Australia",
      "recall_number": "Z-2055-2014",
      "product_description": "BAM (SHAPE) ARM, LEFT, an articulating, arm bracket that is attached to the service head of boom systems which are used for supporting and positioning equipment in operating rooms and patient care areas.  The product comes as component of kit part number 0682000976, and the Arm part number in the kit is P17087.  The kit, when assembled, supports a flat-panel video monitor that is attached to one of two types of Stryker Communications boom systems, EDS or FLEXiS.",
      "product_quantity": "953 SHAPE Arms of all types",
      "reason_for_recall": "During the use of the SHAPE Arm there is the potential for the monitor mount to come off of the SHAPE Arm if the lock washer and screw are not present.",
      "recall_initiation_date": "20140317",
      "center_classification_date": "20140715",
      "termination_date": "20150515",
      "report_date": "20140723",
      "code_info": "Serial numbers  080924DC001 through 120227DC025     Serial and part numbers are located on underside of the cover of the portion of the Arm that is attached to the boom's service head.  Screws must be removed and the cover detached to view these numbers."
    }
  ]
}